Medline procedure kits labeled as: 1) AV FISTULA PACK, Pack Number DYNJ80088; 2) AV FISTULA PACK-LF, Pack Number DYNJ0131782J; 3) CABG, Pack Number DYNJ906462B; 4) CABG ACCESSORY PACK-LF, Pack Number DYNJ0261346U; 5) CARDIAC CATH PACEMAKER-LF, Pack Number DYNJ42997C; 6) CARDIAC CATH PACK, Pack Number DYNJ61256A; 7) CARDIAC CATH PACK-LF, Pack Number DYNJ61527; 8) CARDIAC CATH PACK-LF, Pack Number DYNJ53642C; 9) CARDIOVASCULAR PACK LJ-LF, Pack Number DYNJ49635F; 10) CAROTID PACK, Pack Number DYNJ0810062AI; 11) CASTRO CABG-COMPONENT PACK, Pack Number DYNJ44857T; 12) CATH PACK, Pack Number DYNJ59830A; 13) CHOG CARDIAC CATH PACK, Pack Number SYNJ10304B; 14) CICC, Pack Number DYNJ53854A; 15) CMHP PACEMAKER PACK, Pack Number DYNJ30217G; 16) DR NACKMAN VEIN PACK, Pack Number DYNJ56235; 17) ENDOVASCULAR AAA PACK-LF, Pack Number DYNJ55482F; 18) EP DEVICE, Pack Number DYNJ65916B; 19) EP STUDY PACK-LF, Pack Number DYNJ38059A; 20) FHZ PACEMAKER PACK, Pack Number DYNJ36931D; 21) GV OPEN HEART PACK, Pack Number DYNJ60618C; 22) HEART CABG CDS, Pack Number CDS983376Q; 23) HEART DRAPE & SUPPLY PACK, Pack Number DYNJ65226; 24) HH ARTERIOGRAM PACK, Pack Number DYNJ59466A; 25) LEX OPEN HEART, Pack Number DYNJ900968I; 26) MAJOR VASCULAR PACK SAFETY, Pack Number DYNJ54738C; 27) NIR PACK, Pack Number DYNJ63892A; 28) OPEN HEART, Pack Number DYNJ904261B; 29) OPEN HEART, Pack Number DYNJ904261B; 30) OPEN HEART, Pack Number DYNJ905021J; 31) OPEN HEART A PACK, Pack Number DYNJ04478R; 32) OPEN HEART ACC, Pack Number DYNJ23533D; 33) OPEN HEART CDS, Pack Number CDS840171AB; 34) OPEN HEART DRAPE PACK-LF, Pack Number DYNJ0382048N; 35) OPEN HEART PACK, Pack Number DYNJ45122C; 36) OPEN HEART PACK, Pack Number DYNJ55648; 37) OPEN HEART PACK, Pack Number DYNJ04381J; 38) OPEN HEART PACK & BASIN SET-LF, Pack Number DYNJ24300M; 39) OPEN HEART PACK-LF, Pack Number DYNJ0369758N; 40) OPEN HEART PART B PACK, Pack Number DYNJ43038N; 41) OPEN HEART SUPPLY, Pack Number DYNJ65641; 42) OPEN PACK, Pack Number DYNJ67442B; 43) PACEMAKER, Pack Number DYNJ59399I; 44) PACEMAKER IMPLANT PACK-LF, Pack Number DYNJ0373237M; 45) PACEMAKER PACK, Pack Number DYNJ66165; 46) PACEMAKER PACK, Pack Number DYNJ41134B; 47) PACEMAKER PACK, Pack Number DYNJ61260A; 48) PACEMAKER PACK, Pack Number DYNJ60635; 49) PACEMAKER PACK, Pack Number DYNJ42602C; 50) PACEMAKER PACK 0121271-LF, Pack Number DYNJ32897J; 51) PACEMAKER PACK CV-LF, Pack Number DYNJ39912B; 52) PACEMAKER PACK-LF, Pack Number DYNJ53112B; 53) PACEMAKER PACK-LF, Pack Number DYNJ51844B; 54) PACEMAKER PACK-LF, Pack Number DYNJ0854874R; 55) PEDI EP PACK, Pack Number DYNJ54750G; 56) PERMANENT PACING PACK-LF, Pack Number DYNJ22036F; 57) PK, OPEN HEART, Pack Number DYNJ44015B; 58) PK, OPEN HEART-CABG, Pack Number DYNJ45048C; 59) RR-ADULT PUMP PACK-LF, Pack Number DYNJ0394805N; 60) RRMC CAROTID PACK-2, Pack Number DYNJ33544L; 61) RR-PEDIATRIC PUMP PACK-LF, Pack Number DYNJ0394730P; 62) TAVR PACK, Pack Number DYNJ58386A; 63) THORACIC PACK, Pack Number DYNJ80309; 64) VASCULAR, Pack Number DYNJ45376G; 65) VASCULAR PACK, Pack Number DYNJ48288C; 66) VASCULAR PACK, Pack Number DYNJ54979; 67) VASCULAR PACK, Pack Number DYNJ64883C; 68) VASCULAR PACK, Pack Number DYNJ43945I; 69) VASCULAR PACK, Pack Number DYNJ43945I; 70) VASCULAR PACK, Pack Number DYNJ67426A; 71) VASCULAR PACK WRO-LF, Pack Number DYNJ41863C; 72) VASCULAR SUB, Pack Number DYNJ60681A; 73) VASCULAR SUB, Pack Number DYNJ60681A
Enforcement
- Recall Number
- Z-1438-2025
- Event ID
- 96374
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- MEDLINE INDUSTRIES, LP - Northfield
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- April 2, 2025
- Initiation Date
- February 11, 2025
- Classification Date
- March 25, 2025
- Address
- 3 Lakes Dr, N/A, Northfield, IL, 60093-2753, United States
Description
Medline procedure kits labeled as: 1) AV FISTULA PACK, Pack Number DYNJ80088; 2) AV FISTULA PACK-LF, Pack Number DYNJ0131782J; 3) CABG, Pack Number DYNJ906462B; 4) CABG ACCESSORY PACK-LF, Pack Number DYNJ0261346U; 5) CARDIAC CATH PACEMAKER-LF, Pack Number DYNJ42997C; 6) CARDIAC CATH PACK, Pack Number DYNJ61256A; 7) CARDIAC CATH PACK-LF, Pack Number DYNJ61527; 8) CARDIAC CATH PACK-LF, Pack Number DYNJ53642C; 9) CARDIOVASCULAR PACK LJ-LF, Pack Number DYNJ49635F; 10) CAROTID PACK, Pack Number DYNJ0810062AI; 11) CASTRO CABG-COMPONENT PACK, Pack Number DYNJ44857T; 12) CATH PACK, Pack Number DYNJ59830A; 13) CHOG CARDIAC CATH PACK, Pack Number SYNJ10304B; 14) CICC, Pack Number DYNJ53854A; 15) CMHP PACEMAKER PACK, Pack Number DYNJ30217G; 16) DR NACKMAN VEIN PACK, Pack Number DYNJ56235; 17) ENDOVASCULAR AAA PACK-LF, Pack Number DYNJ55482F; 18) EP DEVICE, Pack Number DYNJ65916B; 19) EP STUDY PACK-LF, Pack Number DYNJ38059A; 20) FHZ PACEMAKER PACK, Pack Number DYNJ36931D; 21) GV OPEN HEART PACK, Pack Number DYNJ60618C; 22) HEART CABG CDS, Pack Number CDS983376Q; 23) HEART DRAPE & SUPPLY PACK, Pack Number DYNJ65226; 24) HH ARTERIOGRAM PACK, Pack Number DYNJ59466A; 25) LEX OPEN HEART, Pack Number DYNJ900968I; 26) MAJOR VASCULAR PACK SAFETY, Pack Number DYNJ54738C; 27) NIR PACK, Pack Number DYNJ63892A; 28) OPEN HEART, Pack Number DYNJ904261B; 29) OPEN HEART, Pack Number DYNJ904261B; 30) OPEN HEART, Pack Number DYNJ905021J; 31) OPEN HEART A PACK, Pack Number DYNJ04478R; 32) OPEN HEART ACC, Pack Number DYNJ23533D; 33) OPEN HEART CDS, Pack Number CDS840171AB; 34) OPEN HEART DRAPE PACK-LF, Pack Number DYNJ0382048N; 35) OPEN HEART PACK, Pack Number DYNJ45122C; 36) OPEN HEART PACK, Pack Number DYNJ55648; 37) OPEN HEART PACK, Pack Number DYNJ04381J; 38) OPEN HEART PACK & BASIN SET-LF, Pack Number DYNJ24300M; 39) OPEN HEART PACK-LF, Pack Number DYNJ0369758N; 40) OPEN HEART PART B PACK, Pack Number DYNJ43038N; 41) OPEN HEART SUPPLY, Pack Number DYNJ65641; 42) OPEN PACK, Pack Number DYNJ67442B; 43) PACEMAKER, Pack Number DYNJ59399I; 44) PACEMAKER IMPLANT PACK-LF, Pack Number DYNJ0373237M; 45) PACEMAKER PACK, Pack Number DYNJ66165; 46) PACEMAKER PACK, Pack Number DYNJ41134B; 47) PACEMAKER PACK, Pack Number DYNJ61260A; 48) PACEMAKER PACK, Pack Number DYNJ60635; 49) PACEMAKER PACK, Pack Number DYNJ42602C; 50) PACEMAKER PACK 0121271-LF, Pack Number DYNJ32897J; 51) PACEMAKER PACK CV-LF, Pack Number DYNJ39912B; 52) PACEMAKER PACK-LF, Pack Number DYNJ53112B; 53) PACEMAKER PACK-LF, Pack Number DYNJ51844B; 54) PACEMAKER PACK-LF, Pack Number DYNJ0854874R; 55) PEDI EP PACK, Pack Number DYNJ54750G; 56) PERMANENT PACING PACK-LF, Pack Number DYNJ22036F; 57) PK, OPEN HEART, Pack Number DYNJ44015B; 58) PK, OPEN HEART-CABG, Pack Number DYNJ45048C; 59) RR-ADULT PUMP PACK-LF, Pack Number DYNJ0394805N; 60) RRMC CAROTID PACK-2, Pack Number DYNJ33544L; 61) RR-PEDIATRIC PUMP PACK-LF, Pack Number DYNJ0394730P; 62) TAVR PACK, Pack Number DYNJ58386A; 63) THORACIC PACK, Pack Number DYNJ80309; 64) VASCULAR, Pack Number DYNJ45376G; 65) VASCULAR PACK, Pack Number DYNJ48288C; 66) VASCULAR PACK, Pack Number DYNJ54979; 67) VASCULAR PACK, Pack Number DYNJ64883C; 68) VASCULAR PACK, Pack Number DYNJ43945I; 69) VASCULAR PACK, Pack Number DYNJ43945I; 70) VASCULAR PACK, Pack Number DYNJ67426A; 71) VASCULAR PACK WRO-LF, Pack Number DYNJ41863C; 72) VASCULAR SUB, Pack Number DYNJ60681A; 73) VASCULAR SUB, Pack Number DYNJ60681A
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
1) Pack Number DYNJ80088: UDI/DI 10195327010522 (each) 40195327010523 (case), Lot Number 23DBG477A; 2) Pack Number DYNJ0131782J: UDI/DI 10195327467548 (each) 40195327467549 (case), Lot Number 23LMD975A; 3) Pack Number DYNJ906462B: UDI/DI 10195327057886 (each) 40195327057887 (case), Lot Number 22GBV822Z; 4) Pack Number DYNJ0261346U: UDI/DI 10195327055967 (each) 40195327055968 (case), Lot Number 22GMB113A; 5) Pack Number DYNJ42997C: UDI/DI 10193489988420 (each) 40193489988421 (case), Lot Number 21JMG551A; 6) Pack Number DYNJ61256A: UDI/DI 10193489694918 (each) 40193489694919 (case), Lot Number 23JBH403A; 7) Pack Number DYNJ61527: UDI/DI 10193489578881 (each) 40193489578882 (case), Lot Number 21LBJ158A; 8) Pack Number DYNJ53642C: UDI/DI 10193489721607 (each) 40193489721608 (case), Lot Number 24AMF039A; 9) Pack Number DYNJ49635F: UDI/DI 10195327191405 (each) 40195327191406 (case), Lot Number 23LMF122A; 10) Pack Number DYNJ0810062AI: UDI/DI 10193489866049 (each) 40193489866040 (case), Lot Number 22CMB582A; 11) Pack Number DYNJ44857T: UDI/DI 10195327221959 (each) 40195327221950 (case), Lot Number 22JBD481Z; 12) Pack Number DYNJ59830A: UDI/DI 10193489570236 (each) 40193489570237 (case), Lot Number 21EMF434A; 13) Pack Number SYNJ10304B: UDI/DI 10193489786965 (each) 40193489786966 (case), Lot Number 21LBD584A; 14) Pack Number DYNJ53854A: UDI/DI 10193489631517 (each) 40193489631518 (case), Lot Number 21GMB561A; 15) Pack Number DYNJ30217G: UDI/DI 10193489922141 (each) 40193489922142 (case), Lot Number 23CBN679A; 16) Pack Number DYNJ56235: UDI/DI 10889942621552 (each) 40889942621553 (case), Lot Number 23EBD843A; 17) Pack Number DYNJ55482F: UDI/DI 10195327332273 (each) 40195327332274 (case), Lot Number 23BBS877A; 18) Pack Number DYNJ65916B: UDI/DI 10193489902044 (each) 40193489902045 (case), Lot Number 23JBT190A; 19) Pack Number DYNJ38059A: UDI/DI 10193489401110 (each) 40193489401111 (case), Lot Number 20LMF803A; 20) Pack Number DYNJ36931D: UDI/DI 10193489927559 (each) 40193489927550 (case), Lot Number 23BBS364Z; 21) Pack Number DYNJ60618C: UDI/DI 10193489475630 (each) 40193489475631 (case), Lot Number 22GMD148Z; 22) Pack Number CDS983376Q: UDI/DI 10195327379797 (each) 40195327379798 (case), Lot Number 23IBG370Z; 23) Pack Number DYNJ65226: UDI/DI 10193489253672 (each) 40193489253673 (case), Lot Number 20BDA909A; 24) Pack Number DYNJ59466A: UDI/DI 10193489562613 (each) 40193489562614 (case), Lot Number 21BBF996A; 25) Pack Number DYNJ900968I: UDI/DI 10193489453188 (each) 40193489453189 (case), Lot Number 22IBM300N; 26) Pack Number DYNJ54738C: UDI/DI 10193489778755 (each) 40193489778756 (case), Lot Number 21EBF317A; 27) Pack Number DYNJ63892A: UDI/DI 10195327176518 (each) 40195327176519 (case), Lot Number 23LME320A; 28) Pack Number DYNJ904261B: UDI/DI 10193489479539 (each) 40193489479530 (case), Lot Number 22IBN057Z; 29) Pack Number DYNJ904261B: UDI/DI 10193489479539 (each) 40193489479530 (case), Lot Number 22JBJ452Z; 30) Pack Number DYNJ905021J: UDI/DI 10195327281625 (each) 40195327281626 (case), Lot Number 23DBI088Z; 31) Pack Number DYNJ04478R: UDI/DI 10888277748835 (each) 40888277748836 (case), Lot Number 22GMA554A; 32) Pack Number DYNJ23533D: UDI/DI 10193489470659 (each) 40193489470650 (case), Lot Number 22GMC009A; 33) Pack Number CDS840171AB: UDI/DI 10195327167042 (each) 40195327167043 (case), Lot Number 22GMG011Z; 34) Pack Number DYNJ0382048N: UDI/DI 10195327537364 (each) 40195327537365 (case), Lot Number 23JBV598A; 35) Pack Number DYNJ45122C: UDI/DI 10193489496369 (each) 40193489496360 (case), Lot Number 21EME856A; 36) Pack Number DYNJ55648: UDI/DI 10889942581900 (each) 40889942581901 (case), Lot Number 21LBN735A; 37) Pack Number DYNJ04381J: UDI/DI 10888277745155 (each) 40888277745156 (case), Lot Number 23GMB245A; 38) Pack Number DYNJ24300M: UDI/DI 10193489823608 (each) 40193489823609 (case), Lot Number 23FBL428A; 39) Pack Number DYNJ0369758N: UDI/DI 10193489501247 (each) 40193489501248 (case), Lot Number 22GBO541Z; 40) Pack Number DYNJ43038N: UDI/DI 10195327263911 (each) 40195327263912 (case), Lot Number 23EMD464A; 41) Pack Number DYNJ65641: UDI/DI 10193489282986 (each) 40193489282987 (case), Lot Number 21BBF660A; 42) Pack Number DYNJ67442B: UDI/DI 10195327133542 (each) 40195327133543 (case), Lot Number 23DBR515A; 43) Pack Number DYNJ59399I: UDI/DI 10195327442071 (each) 40195327442072 (case), Lot Number 23HMB191A; 44) Pack Number DYNJ0373237M: UDI/DI 10193489369014 (each) 40193489369015 (case), Lot Number 23LBU493Z; 45) Pack Number DYNJ66165: UDI/DI 10193489319729 (each) 40193489319720 (case), Lot Number 20DBP548A; 46) Pack Number DYNJ41134B: UDI/DI 10889942011643 (each) 40889942011644 (case), Lot Number 21BBI489A; 47) Pack Number DYNJ61260A: UDI/DI 10193489857481 (each) 40193489857482 (case), Lot Number 21HMF542A; 48) Pack Number DYNJ60635: UDI/DI 10193489518153 (each) 40193489518154 (case), Lot Number 21IMG997A; 49) Pack Number DYNJ42602C: UDI/DI 10195327019198 (each) 40195327019199 (case), Lot Number 23BBR329A; 50) Pack Number DYNJ32897J: UDI/DI 10193489648102 (each) 40193489648103 (case), Lot Number 20AKB914A; 51) Pack Number DYNJ39912B: UDI/DI 10889942705078 (each) 40889942705079 (case), Lot Number 21GMA579A; 52) Pack Number DYNJ53112B: UDI/DI 10193489650617 (each) 40193489650618 (case), Lot Number 21IMF023A; 53) Pack Number DYNJ51844B: UDI/DI 10889942615773 (each) 40889942615774 (case), Lot Number 21LME446A; 54) Pack Number DYNJ0854874R: UDI/DI 10195327094966 (each) 40195327094967 (case), Lot Number 22CMB704A; 55) Pack Number DYNJ54750G: UDI/DI 10195327021351 (each) 40195327021352 (case), Lot Number 22GBH035A; 56) Pack Number DYNJ22036F: UDI/DI 10889942528899 (each) 40889942528890 (case), Lot Number 20FBH674A; 57) Pack Number DYNJ44015B: UDI/DI 10193489215939 (each) 40193489215930 (case), Lot Number 23EMD927A; 58) Pack Number DYNJ45048C: UDI/DI 10193489808889 (each) 40193489808880 (case), Lot Number 21IMD394A; 59) Pack Number DYNJ0394805N: UDI/DI 10193489245240 (each) 40193489245241 (case), Lot Number 23LME396A; 60) Pack Number DYNJ33544L: UDI/DI 10195327279318 (each) 40195327279319 (case), Lot Number 23LMF755A; 61) Pack Number DYNJ0394730P: UDI/DI 10193489430943 (each) 40193489430944 (case), Lot Number 21LMG091A; 62) Pack Number DYNJ58386A: UDI/DI 10193489397468 (each) 40193489397469 (case), Lot Number 21VBB871A; 63) Pack Number DYNJ80309: UDI/DI 10195327023218 (each) 40195327023219 (case), Lot Number 23LMF645A; 64) Pack Number DYNJ45376G: UDI/DI 10195327209353 (each) 40195327209354 (case), Lot Number 23KMD075A; 65) Pack Number DYNJ48288C: UDI/DI 10888277745599 (each) 40888277745590 (case), Lot Number 21IMH225A; 66) Pack Number DYNJ54979: UDI/DI 10889942519910 (each) 40889942519911 (case), Lot Number 21LBL535Z; 67) Pack Number DYNJ64883C: UDI/DI 10193489369670 (each) 40193489369671 (case), Lot Number 23BBR373A; 68) Pack Number DYNJ43945I: UDI/DI 10193489704242 (each) 40193489704243 (case), Lot Number 23GME703A; 69) Pack Number DYNJ43945I: UDI/DI 10193489704242 (each) 40193489704243 (case), Lot Number 23HMC425A; 70) Pack Number DYNJ67426A: UDI/DI 10195327049317 (each) 40195327049318 (case), Lot Number 23HMC554A; 71) Pack Number DYNJ41863C: UDI/DI 10195327463632 (each) 40195327463633 (case), Lot Number 23HBW577A; 72) Pack Number DYNJ60681A: UDI/DI 10193489780819 (each) 40193489780810 (case), Lot Number 20EBI994A; 73) Pack Number DYNJ60681A: UDI/DI 10193489780819 (each) 40193489780810 (case), Lot Number 21BBI579A
Worldwide distribution - US Nationwide and the country of Canada.
3860 units