FDA Enforcement
Class II
Terminated
NobelReplace CC PMC RP 4.3x10mm, Article No. 37291.
Recall: Z-1438-2015
·
Reported April 22, 2015
Enforcement
- Recall Number
- Z-1438-2015
- Event ID
- 70821
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Nobel Biocare Usa Llc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 22, 2015
- Initiation Date
- March 9, 2015
- Classification Date
- April 13, 2015
- Termination Date
- December 30, 2015
- Address
- 22715/22725 Savi Ranch Pkwy, N/A, Yorba Linda, CA, 92887, United States
Description
NobelReplace CC PMC RP 4.3x10mm, Article No. 37291.
Reason
A material used in the packaging of the dental implants may impart a sticky substance on the top surfaces of the implants and in some cases on the first threads.
Code Info
Batch No. 13011148
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Austria, Australia, Belgium, Bulgaria, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Greece, Croatia, Hungary, Ireland, Italy, Lithuania, Luxembourg, Mexico, Netherlands, Norway, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Sweden, Turkey, and United Arab Emirates.
Quantity
5 units