HeartWare HVAD System Battery Charger AC Adapter The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.
Enforcement
- Recall Number
- Z-1428-2020
- Event ID
- 84829
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Heartware, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 18, 2020
- Initiation Date
- January 23, 2020
- Classification Date
- March 9, 2020
- Address
- 14400 Nw 60th Ave, Miami Lakes, FL, 33014-2807, United States
Description
HeartWare HVAD System Battery Charger AC Adapter The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.
Potential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port.
Model Number: 1640, all Serial Numbers. Known Model Numbers (Serial Numbers) GTIN Numbers: 1640CA (BAC004576) 00888707001298, 1640GB (BAC004765) 00888707001328, 1640GB (BAC004768) 00888707001328, 1640AR (BAC004772) 00888707001274, 1640IN (BAC004829) 00888707000543, 1640DE (BAC005530) 00888707001304, 1640CA (BAC005842) 00888707001298, 1640CA (BAC005854) 00888707001298, 1640BR (BAC005875) 00888707001281, 1640BR (BAC005876) 00888707001281, 1640GB (BAC005920) 00888707001328, 1640AU (BAC005966) 00888707001267, 1640AU (BAC005980) 00888707001267, 1640IN (BAC005997) 00888707000543, 1640AU (BAC006001) 00888707001267, 1640IL (BAC006027) 00888707001335, 1640IT (BAC006032) 00888707001342, 1640IN (BAC006033) 00888707000543, 1640IN (BAC006034) 00888707000543, 1640IN (BAC006035) 00888707000543, 1640IN (BAC006036) 00888707000543, 1640IL (BAC006037) 00888707001335, 1640IL (BAC006038) 00888707001335, 1640GB (BAC006039) 00888707001328, 1640DE (BAC006040) 00888707001304, 1640DE (BAC006041) 00888707001304, 1640DE (BAC006056) 00888707001304, 1640CA (BAC006087) 00888707001298, 1640IN (BAC006102) 00888707000543, 1640IN (BAC006104) 00888707000543, 1640DE (BAC090005) 00888707002325, 1640CH (BAC090006) 00888707002318, 1640CH (BAC090012) 00888707002318, 1640GB (BAC090079) 00888707002332, 1640IT (BAC090082) 00888707002363, 1640IT (BAC090083) 00888707002363, 1640GB (BAC090084) 00888707002332, 1640CH (BAC090165) 00888707002318, 1640CH (BAC090166) 00888707002318, 1640DE (BAC090167) 00888707002325, 1640DE (BAC090168) 00888707002325, 1640DE (BAC090170) 00888707002325, 1640GB (BAC090171) 00888707002332, 1640GB (BAC090172) 00888707002332, 1640IT (BAC090173) 00888707002363, 1640IT (BAC090174) 00888707002363, 1640DE (BAC090175) 00888707002325, 1640IN (BAC090237) 00888707002356, 1640IN (BAC090238) 00888707002356, 1640DE (BAC090414) 00888707002325, 1640DE (BAC090415) 00888707002325, 1640DE (BAC090416) 00888707002325, 1640DE (BAC090417) 00888707002325, 1640DE (BAC090447) 00888707002325, 1640DE (BAC090448) 00888707002325, 1640IT (BAC090611) 00888707002363, 1640AU (BAC090659) 00888707002288, 1640AU (BAC090660) 00888707002288, 1640CA (bac090661) 00888707002301, 1640CA (BAC090662) 00888707002301, 1640GB (BAC090663) 00888707002332, 1640GB (BAC090664) 00888707002332, 1640IT (BAC090665) 00888707002363, 1640DE (BAC090666) 00888707002325, 1640DE (BAC090667) 00888707002325, 1640AU (BAC090732) 00888707002288, 1640AU (bac090733) 00888707002288, 1640GB (BAC090734) 00888707002332, 1640GB (BAC090735) 00888707002332, 1640DE (BAC090864) 00888707002325, 1640DE (BAC090976) 00888707002325, 1640DE (BAC091324) 00888707003018, 1640DE (BAC091325) 00888707003018, 1640DE (BAC091326) 00888707003018, 1640DE (BAC091327) 00888707003018, 1640DE (BAC091330) 00888707003018, 1640DE (BAC091331) 00888707003018, 1640DE (BAC091332) 00888707003018, 1640DE (BAC091333) 00888707003018, 1640DE (BAC091420) 00888707003018, 1640DE (BAC091421) 00888707003018, 1640CA (BAC091690) 00888707003056, 1640CA (BAC091691) 00888707003056, 1640DE (BAC091695) 00888707003018, 1640DE (BAC091696) 00888707003018, 1640DE (BAC091697) 00888707003018, 1640DE (BAC091698) 00888707003018, 1640DE (BAC091699) 00888707003018, 1640DE (BAC091700) 00888707003018, 1640DE (BAC091701) 00888707003018, 1640DE (BAC091702) 00888707003018, 1640DE (BAC091703) 00888707003018, 1640DE (BAC091704) 00888707003018, 1640GB (BAC091738) 00888707003025, 1640GB (BAC091739) 00888707003025, 1640DE (BAC091741) 00888707003018, 1640GB (BAC091743) 00888707003025, 1640GB (BAC091744) 00888707003025, 1640DE (BAC091758) 00888707003018, 1640DE (BAC091811) 00888707003018, 1640DE (BAC091812) 00888707003018, 1640DE (BAC092205) 00888707003018, 1640DE (BAC092206) 00888707003018, 1640IT (BAC092283) 00888707003032, 1640GB (BAC092289) 00888707003025, 1640JP (BAC092290) 00888707002370, 1640JP (BAC092292) 00888707002370, 1640CA (BAC092294) 00888707003056, 1640DE (BAC092332) 00888707003018, 1640DE (BAC092333) 00888707003018, 1640DE (BAC092334) 00888707003018, 1640DE (BAC092335) 00888707003018, 1640DE (BAC092391) 00888707003018, 1640IN (BAC092396) 00888707003049, 1640JP (BAC092459) 00888707002370, 1640DE (BAC092495) 00888707003018, 1640DE (BAC092506) 00888707003018, 1640DE (BAC092556) 00888707003018, 1640DE (BAC092557) 00888707003018, 1640DE (BAC092558) 00888707003018, 1640DE (BAC092559) 00888707003018, 1640DE (BAC092560) 00888707003018, 1640JP (BAC092595) 00888707002370, 1640JP (BAC092680) 00888707002370, 1640JP (BAC092681) 00888707002370, 1640JP (BAC092682) 00888707002370, 1640JP (BAC092683) 00888707002370, 1640JP (BAC092684) 00888707002370, 1640JP (BAC092685) 00888707002370, 1640JP (BAC092687) 00888707002370, 1640JP (BAC092689) 00888707002370, 1640JP (BAC092690) 00888707002370, 1640GB (BAC092748) 00888707003025, 1640GB (BAC092752) 00888707003025, 1640DE (BAC092760) 00888707003018, 1640IN (BAC100001) 00888707005913, 1640IN (BAC100002) 00888707005913, 1640GB (BAC100003) 00888707005890, 1640GB (BAC100004) 00888707005890, 1640DE (BAC100005) 00888707005883, 1640CA (BAC100006) 00888707005869, 1640IT (BAC100009) 00888707005920, 1640IT (BAC100010) 00888707005920, 1640CA (BAC100015) 00888707005869, 1640DE (BAC100030) 00888707005883, 1640DE (BAC100031) 00888707005883, 1640CA (BAC100256) 00888707005869, 1640CA (BAC100349) 00888707005869, 1640DE (BAC100385) 00888707005883, 1640CA (BAC100391) 00888707005869, 1640GB (BAC100402) 00888707005890, 1640IT (BAC100412) 00888707005920, 1640GB (BAC100413) 00888707005890, 1640GB (BAC100414) 00888707005890, 1640GB (BAC100415) 00888707005890, 1640DE (BAC100451) 00888707005883, 1640DE (BAC100452) 00888707005883, 1640GB (BAC100525) 00888707005890, 1640CH (BAC100597) 00888707005876, 1640IL (BAC100599) 00888707005906, 1640CH (BAC100600) 00888707005876, 1640DE (BAC100725) 00888707005883, 1640GB (BAC100742) 00888707005890, 1640GB (BAC100743) 00888707005890, 1640GB (BAC100866) 00888707005890, 1640GB (BAC100867) 00888707005890, 1640GB (BAC100868) 00888707005890, 1640DE (BAC000048) 00888707001304, 1640DE (BAC000122) 00888707001304, 1640DE (BAC000123) 00888707001304, 1640DE (BAC000125) 00888707001304, 1640CA (BAC000544) 00888707001298, 1640DE (BAC002505) 00888707001304, 1640DE (BAC091485) 00888707003018, 1640DE (BAC091486) 00888707003018, 1640DE (BAC091487) 00888707003018, 1640DE (BAC091488) 00888707003018, 1640DE (BAC091489) 00888707003018, 1640DE (BAC091494) 00888707003018, 1640DE (BAC091495) 00888707003018, 1640DE (BAC091496) 00888707003018, 1640DE (BAC091497) 00888707003018, 1640DE (BAC091498) 00888707003018, 1640DE (BAC091499) 00888707003018, 1640DE (BAC091500) 00888707003018, 1640DE (BAC091501) 00888707003018, 1640DE (BAC091502) 00888707003018, 1640DE (BAC091503) 00888707003018, 1640DE (BAC091504) 00888707003018, 1640DE (BAC091505) 00888707003018, 1640DE (BAC091506) 00888707003018, 1640DE (BAC091507) 00888707003018, 1640DE (BAC091539) 00888707003018, 1640IN (BAC091710) 00888707003049, 1640IN (BAC091711) 00888707003049, 1640IT (BAC091715) 00888707003032, 1640CA (BAC091716) 00888707003056, 1640CA (BAC091717) 00888707003056, 1640CA (BAC091718) 00888707003056, 1640CA (BAC091719) 00888707003056, 1640IT (BAC091735) 00888707003032, 1640IT (BAC091736) 00888707003032, 1640IT (BAC091737) 00888707003032, 1640GB (BAC091742) 00888707003025, 1640GB (BAC091805) 00888707003025, 1640GB (BAC091806) 00888707003025, 1640GB (BAC091807) 00888707003025, 1640GB (BAC091808) 00888707003025, 1640DE (BAC091813) 00888707003018, 1640DE (BAC091814) 00888707003018, 1640JP (BAC091961) 00888707002370, 1640AU (BAC000247) 00888707001267, 1640AU (BAC000679) 00888707001267, 1640AU (BAC001066) 00888707001267, 1640AU (BAC001076) 00888707001267, 1640DE (BAC001196) 00888707001304, 1640CH (BAC001384) 00888707001311, 1640IL (BAC001707) 00888707001335, 1640CH (BAC001811) 00888707001311, 1640CH (BAC001812) 00888707001311, 1640IT (BAC001815) 00888707001342, 1640CH (BAC001818) 00888707001311, 1640IT (BAC001828) 00888707001342, 1640AR (BAC001852) 00888707001274, 1640BR (BAC001980) 00888707001281, 1640IL (BAC001982) 00888707001335, 1640IN (BAC002000) 00888707000543, 1640AR (BAC002157) 00888707001274, 1640AU (BAC002192) 00888707001267, 1640AU (BAC002333) 00888707001267, 1640JP (BAC002547) 00888707001359, 1640JP (BAC002548) 00888707001359, 1640CA (BAC002777) 00888707001298, 1640CA (BAC002778) 00888707001298, 1640CA (BAC002779) 00888707001298, 1640CA (BAC002780) 00888707001298, 1640CA (BAC002781) 00888707001298, 1640CA-CLIN (BAC003225) 00888707001571, 1640CA-CLIN (BAC003226) 00888707001571, 1640CA-CLIN (BAC003227) 00888707001571, 1640CA-CLIN (BAC003228) 00888707001571, 1640DE (BAC003345) 00888707001304, 1640DE (BAC003346) 00888707001304, 1640DE (BAC003349) 00888707001304, 1640GB (BAC003354) 00888707001328, 1640GB (BAC003357) 00888707001328, 1640GB (BAC003358) 00888707001328, 1640GB (BAC003359) 00888707001328, 1640GB (BAC003462) 00888707001328, 1640DE (BAC004553) 00888707001304, 1640CA (BAC004575) 00888707001298, 1640CH (BAC004665) 00888707001311, 1640IL (BAC004763) 00888707001335, 1640IL (BAC004766) 00888707001335, 1640GB (BAC004769) 00888707001328, 1640AR (BAC004771) 00888707001274, 1640AU (BAC004773) 00888707001267, 1640IT (BAC004774) 00888707001342, 1640GB (BAC004775) 00888707001328, 1640IT (BAC004777) 00888707001342, 1640IN (BAC004830) 00888707000543, 1640DE (BAC005320) 00888707001304, 1640IN (BAC005967) 00888707000543, 1640IN (BAC005968) 00888707000543, 1640IN (BAC005969) 00888707000543, 1640IN (BAC005970) 00888707000543, 1640AU (BAC005973) 00888707001267, 1640AU (BAC005981) 00888707001267, 1640IL (BAC006018) 00888707001335, 1640IL (BAC006019) 00888707001335, 1640IL (BAC006020) 00888707001335, 1640IL (BAC006021) 00888707001335, 1640IT (BAC006022) 00888707001342, 1640IT (BAC006023) 00888707001342, 1640IT (BAC006024) 00888707001342, 1640IT (BAC006025) 00888707001342, 1640IN (BAC006103) 00888707000543, 1640IN (BAC006105) 00888707000543, 1640IL (BAC090222) 00888707002349, 1640IL (BAC090226) 00888707002349, 1640GB (BAC090437) 00888707002332, 1640IT (BAC092284) 00888707003032, 1640IN (BAC092285) 00888707003049, 1640IN (BAC092287) 00888707003049, 1640GB (BAC092288) 00888707003025, 1640IN (BAC092397) 00888707003049, 1640IN (BAC092460) 00888707003049, 1640IN (BAC092465) 00888707003049, 1640JP (BAC092592) 00888707002370, 1640JP (BAC092593) 00888707002370, 1640DE (BAC006047) 00888707001304, 1640DE (BAC006002) 00888707001304, 1640GB (BAC006107) 00888707001328, 1640GB (BAC006108) 00888707001328, 1640AU (BAC005907) 00888707001267, 1640AU (BAC001853) 00888707001267, 1640AU (BAC001856) 00888707001267, 1640DE (BAC003922) 00888707001304, 1640DE (BAC004552) 00888707001304, 1640DE (BAC004347) 00888707001304, 1640IT (BAC004173) 00888707001342, 1640GB (BAC004836) 00888707001328, 1640DE (BAC003839) 00888707001304, 1640IN (BAC004701) 00888707000543, 1640CH (BAC004116) 00888707001311, 1640DE (BAC004350) 00888707001304, 1640DE (BAC004176) 00888707001304, 1640DE (BAC004354) 00888707001304, 1640DE (BAC004348) 00888707001304, 1640DE (BAC004351) 00888707001304, 1640DE (BAC003112) 00888707001304, 1640DE (BAC003347) 00888707001304, 1640GB (BAC003451) 00888707001328, 1640GB (BAC001199) 00888707001328, 1640GB (BAC001820) 00888707001328, 1640GB (BAC001823) 00888707001328, 1640DE (BAC002494) 00888707001304, 1640DE (BAC002506) 00888707001304, 1640GB (BAC000809) 00888707001328, 1640CA (BAC002260) 00888707001298, 1640CA (BAC002271) 00888707001298, 1640CA (BAC002335) 00888707001298, 1640JP (BAC002631) 00888707001359, 1640DE (BAC001834) 00888707001304, 1640DE (BAC001829) 00888707001304, 1640DE (BAC000030) 00888707001304, 1640DE (BAC001833) 00888707001304, 1640DE (BAC000026) 00888707001304, 1640CA (BAC001804) 00888707001298, 1640CA (BAC001808) 00888707001298, 1640CA (BAC002334) 00888707001298, 1640DE (BAC001830) 00888707001304, 1640DE (BAC000065) 00888707001304, 1640DE (BAC000069) 00888707001304, 1640DE (BAC000070) 00888707001304, 1640DE (BAC000071) 00888707001304
No affected devices distributed within the US. International distribution to Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, EGYPT, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kazakhstan, Korea, Korea, Republic of, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Netherlands, New Zealand, Norway, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Taiwan, Province of China, Turkey, United Arab Emirates, United Kingdom, Vietnam.
5,489 devices total