FDA Enforcement Class III Terminated

Soluble Transferrin Receptors (STFR), Catalog Number TF10159 (Research Use Only). IVD

Recall: Z-1422-2022 · Reported July 27, 2022

Enforcement

Recall Number
Z-1422-2022
Event ID
90471
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Randox Laboratories Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 27, 2022
Initiation Date
June 9, 2022
Classification Date
July 20, 2022
Termination Date
February 8, 2024
Address
55 Diamond Road, N/A, Crumlin (North), N/A, N/A, Ireland

Description

Soluble Transferrin Receptors (STFR), Catalog Number TF10159 (Research Use Only). IVD

Reason

A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be identified in the Atellica graphical view of the calibration. In most of the cases when this has occurred, it is picked up by the QC generating a Math Error, however, there are reports of QC shifts after calibration without the Math Error flagged that may be related to this issue. Results reported with STFR Level 1, with a concentration less than 1.7mg/L, using TF10159 and TF10161 can be underestimated by up to 30%. This issue has been reported with reagent lots 575475 and 562691, calibrator lots 537725 and 562696. No other complaints received.

Code Info

UDI-DI (GTIN): 05055273215564 All lots within expiry

Distribution

US Nationwide distribution in the state of IL.

Quantity

12 kits (US)