FDA Enforcement
Class II
Ongoing
LifeGlobal Fast Freeze Thawing Kit, Model No. GFT5-055
Recall: Z-1421-2022
·
Reported July 27, 2022
Enforcement
- Recall Number
- Z-1421-2022
- Event ID
- 90469
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- CooperSurgical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- N/A
- Report Date
- July 27, 2022
- Initiation Date
- June 13, 2022
- Classification Date
- July 20, 2022
- Address
- 95 Corporate Dr, N/A, Trumbull, CT, 06611-1350, United States
Description
LifeGlobal Fast Freeze Thawing Kit, Model No. GFT5-055
Reason
There was translation error in the Spanish version of the IFU where the Spanish version instructs the user to leave the blastocyte to air thaw for 5 minutes rather than the correct 5 seconds as indicated in the English version.
Code Info
UDI: (01)00815965020914 (17)220819 (10)GFT5-200218U (01)00815965020914 (17)220819 (10)GFT5-201026U (01)00815965020914 (17)220819 (10)GFT5-21335060 Lots: 21335060, GFT5-200218U, GFT5-201026U
Distribution
Domestic: USA AL, AZ, CA, CT, FL, GA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TX, UT, VT, WA, WI, International: Belgium, Canada, Croatia, Netherlands, Portugal, Spain
Quantity
403 units