FDA Enforcement Class II Ongoing

Exactech Equinoxe REVERSE SHOULDER,38mm Constrained Humeral Liner: a) +0mm, Item Number 320-38-10, b) +2.5mm, Item Number 320-38-13, Shoulder Arthroplasty

Recall: Z-1419-2024 · Reported April 10, 2024

Enforcement

Recall Number
Z-1419-2024
Event ID
94092
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Exactech, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 10, 2024
Initiation Date
March 6, 2024
Classification Date
March 29, 2024
Address
2320 Nw 66th Ct, Gainesville, FL, 32653-1630, United States

Description

Exactech Equinoxe REVERSE SHOULDER,38mm Constrained Humeral Liner: a) +0mm, Item Number 320-38-10, b) +2.5mm, Item Number 320-38-13, Shoulder Arthroplasty

Reason

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Code Info

a) Item Number 320-38-10, UDI/DI 10885862086679, Serial Numbers: 0982088, 0982095, 1035622, 0982086, 1035885, 0982078, 1035878, 1035894, 1036369, 1036373, 1013872, 1013887, 1035623, 1035875, 1035880, 1035881, 1035883, 1119110, 1119136, 1119124, 1119130, 1119144, 1119109, 1119104, 1119123, 1119128, 1119129, 1119146, 1036368, 0982100, 1013871, 1013874, 1074102, 1035619, 1013884, 1035629, 1074121, 1035889, 1035876, 1036374, 1074138, 1035893, 1013886, 1035624, 1074108, 1074122, 1074127, 1074128, 1074139, 1035882, 1036366, 0982084, 1374085, 1119094, 1119102, 0982089, 0982081, 1074131, 0982101, 1036359, 1036364, 1074120, 0982087, 1074147, 1074109, 1442856, 1416801, 1416805, 1374050, 1545547, 1374052, 1416803, 1545546, 1374079, 1374342, 1374339, 1374347, 1416802, 1374077, 1374355, 1416811, 1505058, 1505060, 1442866, 1505054, 1516382, 1516384, 1374351, 1505050, 1516381, 1374043, 1516390, 1119121, 1119127, 1119119, 1119145, 1119098, 1119143, 1074119, 0982090, 1074137, 1013878, 1036365, 1074146, 1013881, 1035620, 1074115, 1035877, 1074136, 1036360, 1036361, 1035890, 1442853, 1374045, 1516375, 1416793, 1374056, 1374082, 1505061, 1442867, 1442857, 1442845, 1416809, 1505053, 1545537, 1416799, 1505041, 1505059, 1545548, 1545542, 1545536, 1374337, 1442858, 1545540, 1416810, 1374357, 1374080, 1516394, 1516372, 1545538, 1442861, 1766544, 1690524, 1690533, 1690523, 1766547, 1816484, 1690531, 1766552, 1816473, 1816486, 1816489, 1816491, 1766946, 1833749, 1816470, 1816477, 1833760, 1766553, 1833757, 1119103, 1119097, 1119120, 1119096, 1119113, 1119133, 1013885, 1074125, 1074107, 1074104, 1074103, 1074112, 1035628, 1374054, 1374338, 1516388, 1545535, 1416812, 1545531, 1505040, 1374344, 1442846, 1545530, 1416806, 1545533, 1416795, 1516391, 1374348, 1374091, 1505049, 1416816, 1374058, 1505063, 1833756, 1766955, 1766539, 1766541, 1816479, 1766970, 1983364, 1983371, 2063115, 2063130, 2063132, 1766951, 1766545, 1983357, 1983354, 1983360, 1983373, 1983374, 1983368, 1833754, 1690532, 1833755, 1766956, 1833762, 1690542, 1690537, 1766972, 2063121, 2135920, 2135917, 2135916, 1119132, 1119112, 1119126, 1074149, 1074129, 1074134, 1374346, 1505048, 1374049, 1374053, 1374047, 1374084, 1505057, 1545525, 1545532, 1545529, 1545539, 1505056, 1505045, 1416819, 1983375, 1983359, 1983376, 1833743, 1766950, 1816474, 2063122, 1983377, 2063134, 1983366, 2063125, 1766949, 1816464, 1766965, 1766554, 1766947, 1983367, 2063120, 2063128, 1833758, 1983353, 1766973, 1816482, 1816488, 1833765, 1766537, 1690534, 1816475, 2063124, 1833752, 1833748, 1816483, 1766534, 1766548, 2293485, 2488822, 2376481, 2274075, 2152443, 2488823, 2239335, 2427322, 2376474, 2152432, 2451622, 2291369, 2293483, 2274069, 2291366, 2239333, 2376472, 2400180, 2451619, 2293500, 2293507, 2152429, 2239322, 2239313, 2239314, 2239317, 2239327, 2239328, 2239332, 2239336, 2239341, 2152449, 2274078, 2195075, 2239331, 2239316, 2239320, 2239324, 2291372, 2293466, 2293469, 2293472, 2293493, 2274077, 2291367, 2291368, 2293487, 2293491, 2239340, 2274076, 2274081, 2293492, 2376463, 2400182, 2400191, 2400192, 2400194, 2400201, 2400188, 2400190, 2400196, 2400199, 2427332, 2427338, 2427333, 2451625, 2451633, 2427334, 2427342, 2488792, 2488798, 2451623, 2488802, 2512975, 2152437, 2152438, 2376458, 2427335, 2488819, 2488811, 2532031, 2451621, 2427339, 2195078, 2239339, 2293490, 2239337, 2293495, 2376455, 2376480, 2400189, 2400187, 2488805, 2451624, 2152434, 2376457, 2135924, 2239319, 2152430, 2135925, 2293496, 1545543, 1516377, 1374086, 1374048, 1374349, 1442850, 1416813, 1516387, 1505043, 1516389, 1416800, 1545545, 1545541, 1442864, 1516386, 1690527, 1690529, 1766974, 2063137, 1766964, 1766538, 1833764, 1833744, 1833746, 1833751, 1690525, 1690541, 1766540, 1816467, 1766952, 1983352, 2063138, 1816471, 2063129, 1766969, 1766963, 1766959, 1766975, 2512984, 2532023, 2532004, 2609759, 2376478, 2376468, 2609732, 2532024, 2609735, 2609719, 2532017, 2274067, 2488814, 2488808, 2427319, 2293478, 2400183, 2532003, 2609761, 2609751, 2293508, 2400179, 2531998, 2451640, 2488815, 2400200, 2427325, 2427330, 2274068, 2488826, 2512978, 2532020, 2195065, 2488806, 2609763, 2609739, 2427320, 2376453, 2293484, 2376456, 2239338, 2376462, 2532030, 2609754, 2609755, 2293464, 2274058, 2488807, 2488789, 2488813, 2531995, 2531996, 2531997, 2532002, 2609709, 2609712, 2609721, 2609722, 2609724, 2609729, 2531993, 2609713, 2609740, 2609746, 2609766, 2152447, 2195084, 2512979, 2152442, 2609723, 2152452, 2195083, 2195071, 2293470, 2195068, 2195072, 2239321, 2488812, 2488790, 2293467, 2152451, 2451639, 2823707, 2823727, 2564801, 2788522, 2564812, 2823698, 2788521, 2748093, 2843341, 2843342, 2564791, 2788531, 2748111, 2788541, 2788526, 2564811, 2823703, 2823697, 2564805, 2748113, 2788540, 2843312, 2748098, 2564814, 2564830, 2564839, 2564825, 2564829, 2564832, 2843314, 2564836, 2647179, 2647183, 2647184, 2647186, 2647188, 2647191, 2647194, 2647195, 2647196, 2564802, 2564806, 2564834, 2564837, 2564804, 2564809, 2564816, 2564793, 2564794, 2564808, 2748096, 2748110, 2748090, 2748095, 2748103, 2788518, 2788519, 2788532, 2823692, 2823708, 2823693, 2843304, 2843317, 2843338, 2564798, 2564803, 2564810, 2564813, 2647180, 2647182, 2564797, 2748104, 2748105, 2748107, 2748092, 2564790, 2564796, 2564826, 2564835, 2843303, 2843316, 2647190, 2647185, 2647193, 2647189, 2748108, 2788536, 1416808, 1545528, 1505047, 1505046, 1516383, 1766967, 1816465, 1690543, 1766953, 1983363, 1983372, 1983361, 2063114, 1766542, 1766948, 1816468, 2063126, 2063127, 1690538, 1816466, 2531999, 2532005, 2609756, 2293498, 2451631, 2609727, 2609748, 2451636, 2532012, 2239330, 2512981, 2609749, 2451632, 2376466, 2293503, 2532025, 2532006, 2532034, 2293506, 2293477, 2609762, 2531983, 2195082, 2488820, 2532015, 2531986, 2532008, 2427329, 2427340, 2488821, 2274074, 2532033, 2195077, 2239315, 2531988, 2376476, 2152444, 2609742, 2274082, 2609731, 2451638, 2291370, 2531987, 2488810, 2532000, 2274062, 2293489, 2239329, 2488803, 2609745, 2609716, 2152445, 2195079, 2376471, 2609714, 2609728, 2488788, 2609715, 2843325, 2973206, 2973244, 2843307, 3741419, 2748091, 3662548, 2973243, 2748109, 3623216, 2973197, 2919582, 2973245, 2973181, 2973240, 3623210, 3662540, 3662513, 2919577, 3623183, 3571630, 2823702, 2919578, 3623177, 3623197, 2823728, 2973241, 2973203, 2919559, 3623199, 3695757, 3695761, 2823722, 2919590, 2919564, 2919571, 2919588, 3662504, 2843336, 2973252, 2919566, 2973251, 2973200, 2788524, 2843349, 2843337, 2919569, 2919592, 2973246, 2919567, 2973237, 2919594, 3623182, 3662512, 2843323, 3591105, 2973201, 3571626, 3741443, 2973185, 3741440, 3571636, 3591104, 2973192, 2919570, 3571617, 2919580, 3623176, 3662511, 2843308, 2919579, 2843344, 2843350, 2843332, 2823720, 2843329, 2843306, 2919581, 2919589, 2919563, 2973236, 3571613, 2973186, 2973191, 2973195, 2973257, 2973210, 3662526, 2748106, 3695752, 3591108, 3695767, 2843326, 2973184, 3623225, 2788528, 3571622, 2973198, 3695788, 2823687, 3662531, 2973189, 2823688, 3695773, 3662536, 2564838, 2823711, 2823719, 2564828, 3623193, 2823705, 3623192, 3623194, 3571623, 3571605, 2973260, 3695779, 2823699, 3623217, 2973196, 3571637, 3571602, 3623221, 3571611, 3623220, 2823721, 3662537, 3695765, 3571639, 3662510, 2823696, 3571638, 2564842, 2973253, 3571633, 3571621, 2843334, 3571599, 3623198, 3571620, 2823701, 3571625, 3623180, 3623189, 2748112, 2788525, 2788538, 2823704, 2823710, 2823714, 2748097, 2748099, 2748100, 2748114, 2919562, 2919583, 2973205, 3571614, 3623196, 3662506, 3662520, 3571635, 3623195, 3662518, 3695769, 3695771, 3571593, 3571594, 3571629, 2788530, 2823706, 3571612, 3623187, 2973247, 3741410, 3695785, 2919584, 2919558, 2919585, 2919573, 2919574, 2973183, 2973204, 2973242, 2973207, 2973256, 2973209, 2973190, 3571618, 3571634, 3623211, 3695775, 3741431, 3571628, 3571600, 3571607, 3741412, 2823686, 2843305, 2919593, 2843339, 2647187, 2843313, 2843346, 2647178, 2647192, 2564800, 3662503, 2564807, 3623179, 3662502, 3695777, 2788520, 2919572, 2919565, 2919560, 3695763, 1766546, 1983369, 1766960, 1766954, 1983355, 2488787, 2376479, 2152450, 2293476, 2274060, 2531991, 2609726, 2274080, 2531994, 2488816, 2531981, 2152435, 2532022, 2274070, 2427326, 2488817, 2451617, 2427328, 2531984, 2609708, 2531982, 2532014, 2512982, 2532032, 2488800, 2293502, 2512977, 2376477, 2512980, 2451616, 2532009, 2195066, 2532010, 2451620, 2135923, 2376467, 2376475, 2609733, 2609725, 2293473, 2239325, 2152440, 3741427, 3741424, 3741411, 3774870, 3695800, 3741403, 3571627, 3741415, 3623213, 3695787, 3741447, 3741444, 3623212, 3741433, 3741428, 3741407, 3741413, 3695801, 3774871, 3695755, 3741406, 3623208, 3662544, 3695799, 3774852, 2973254, 3695766, 3741442, 3774839, 3741423, 3774829, 3774868, 3741404, 3774843, 3774841, 3774830, 3741441, 3695798, 3741422, 3695754, 3741446, 3741450, 3741437, 3774837, 3623191, 3741409, 3774827, 3774826, 3623214, 3695797, 3774867, 3774849, 3662524, 3662550, 3774831, 3571609, 3774832, 3623229, 3774834, 3695770, 3695758, 3623190, 2564827, 3774858, 3774861, 3623223, 3571598, 2973238, 3571610, 3695764, 2823726, 2843345, 2973187, 2843352, 3662515, 3774855, 3695776, 3662547, 3571632, 2823690, 3662551, 3662509, 3774846, 3571603, 2823731, 3695782, 3623215, 3695774, 3662530, 2843333, 3662538, 3774840, 3571597, 3623222, 3774864, 3662539, 3623202, 3695793, 3741430, 3774856, 3571595, 3695789, 3774838, 3774863, 3571601, 3741414, 3741416, 3741402, 3741426, 3741438, 2973194, 3741448, 3623188, 3741445, 3741420, 3774875, 3774873, 3774848, 3774833, 3774859, 3774860, 3774872, 3741429, 3695796, 3774850, 3741439, 3571640, 3741418, 2919575, 3695794, 3571616, 3695759, 3774865, 2919568, 3571624, 3741435, 3741425, 2748089, 3774857, 2919586, 3741449, 3695753, 3774842, 2919591, 3623204, 3774874, 3774847, 2973250, 3843861, 3936248, 3998793, 3936234, 3874466, 3874442, 3843875, 3962184, 3962178, 3936238, 3874474, 3998779, 4050423, 3874477, 3936241, 3962157, 3962152, 3936228, 4034711, 3936231, 3892039, 4050407, 3962191, 3843869, 4105124, 3998755, 4050416, 3843880, 3874467, 3874454, 3936242, 3962151, 3998760, 4050390, 4050409, 3843886, 3957547, 3962187, 3957545, 3957543, 3892036, 3874451, 3874453, 3874460, 3998758, 3892038, 3936259, 4094925, 3962150, 3843898, 3874472, 3936249, 3962193, 3892032, 3957544, 4050402, 3874458, 4094950, 3874452, 4094954, 4094934, 3998754, 3843852, 3998781, 3843857, 3843891, 4105098, 3843855, 3843889, 3874464, 3892019, 4105115, 3936270, 3843868, 4050425, 3936247, 3998766, 3998759, 3843867, 3936230, 4034712, 4094955, 4094910, 4105154, 4105096, 4105086, 3843854, 3936223, 3936264, 3962164, 3936225, 3998790, 3962199, 4050396, 4105179, 3843896, 3962176, 3874479, 4050388, 3936257, 4105099, 3843874, 3892021, 4105100, 3998769, 3936252, 3998752, 3843899, 4050417, 3843856, 3892024, 3892037, 3892040, 3936229, 3874443, 3936265, 4105108, 3843873, 4050393, 3843883, 3874436, 3998772, 3936243, 3998767, 4050420, 3962146, 3936226, 3962147, 3962154, 4050403, 4050418, 4050439, 4105082, 3962167, 3892034, 3874459, 3962182, 4050406, 4050424, 3874480, 3843894, 3874470, 3936227, 3936266, 3998785, 3962156, 3998780, 3998762, 3936255, 3957546, 3874455, 3874456, 3874463, 3874469, 3874473, 3874476, 3892026, 3936221, 3962161, 3998777, 3936258, 3874462, 3843890, 3843853, 3936232, 3962190, 3998763, 4094919, 2293494, 2609711, 2532007, 2609717, 2274079, 2488804, 2293501, 2609718, 2293499, 2488799, 2293468, 2291371, 2488794, 2376473, 2293480, 2293471, 2488829, 2451614, 2451637, 2532026, 2451628, 2488793, 2512983, 2400181, 2488827, 2609757, 2293497, 2274063, 2843324, 2843330, 3662533, 3741421, 3774845, 3662546, 2843335, 3662529, 2843343, 2843318, 2973199, 3571592, 2564831, 3662525, 2788537, 3774854, 2823732, 3662543, 3623203, 2823723, 3571606, 3695762, 3695786, 2823691, 2564840, 2564792, 2564795, 3774869, 3741401, 3774866, 3741405, 2919561, 2788534, 2823685, 2843319, 3662545, 3571631, 3623178, 3774835, 2843310, 3741408, 3695760, 2919556, 3662507, 3695790, 2973248, 3695768, 3662508, 3662549, 2788523, 4198739, 4300276, 4050408, 4105141, 4105094, 4198758, 4222170, 3998776, 4346519, 4300286, 4300270, 4198751, 4198716, 4094920, 4300262, 4439677, 4492222, 3892028, 4105166, 4198730, 3874482, 4198752, 3998789, 4365645, 4198737, 4478153, 3874457, 4105118, 4105110, 4105134, 3998749, 4222169, 4300301, 4492235, 4500729, 4198735, 4365673, 3962194, 3962155, 3998765, 4346486, 4222178, 4198738, 4300263, 4346523, 4198713, 3998786, 4222136, 4105167, 4346507, 4439676, 4346474, 4105138, 4198712, 4300282, 3962158, 4365658, 4365650, 3936244, 4478144, 4050419, 4492233, 4105119, 4198719, 4439642, 4365661, 4300304, 4105084, 4198746, 4365644, 4346481, 4439675, 4365641, 4050437, 4105132, 4105164, 4105135, 3998745, 3962180, 4222171, 4222165, 3998788, 4492216, 4222142, 4105169, 3962168, 4346483, 4365665, 4094924, 4105103, 4105137, 4105145, 3998784, 4198743, 3998782, 4222149, 4222143, 4300264, 4222174, 4222167, 4439637, 4365649, 4439678, 4105165, 4105105, 4346522, 4222157, 4346478, 4439650, 4478142, 3962148, 4222146, 4198736, 4439648, 3843897, 3962149, 3843870, 4198718, 4105136, 4105143, 4198744, 4222130, 4222155, 4439638, 4365629, 4492225, 3998787, 4222158, 4222184, 4222168, 4300302, 4346487, 4050400, 4105150, 4094937, 4105093, 4105156, 3936222, 4222153, 4050399, 3998771, 4439662, 3962163, 4105178, 4094927, 4094939, 4492217, 3843885, 4478147, 4439664, 4500727, 4492219, 4198760, 4105126, 4365647, 4365648, 3962188, 4222180, 4222144, 3843881, 4105159, 4300278, 4300266, 4094963, 4346501, 4105128, 4050435, 4105085, 4198732, 4346489, 4346510, 4346493, 4365672, 4346499, 4439679, 4439667, 4346480, 4500732, 4492231, 4198717, 4346505, 4050430, 4346504, 4222156, 4222162, 4300307, 4050410, 3998756, 4094918, 4492220, 4050421, 4439661, 3936239, 4105121, 4094931, 4105113, 3843866, 4105106, 4105088, 4105091, 4222139, 4365660, 4094923, 4439659, 3936240, 4300296, 4346502, 3998753, 3998768, 3874441, 4094911, 3998757, 3998773, 3874445, 4222183, 3962162, 4198740, 3874471, 4105147, 4222175, 3843879, 4050392, 4094941, 3998791, 4105127, 4346482, 4346515, 4365635, 4478149, 4198725, 4198748, 4094933, 4222138, 4300275, 4365654, 4222132, 4094936, 4222163, 4222134, 4222137, 4222141, 4105104, 4198763, 4198741, 4222147, 4222154, 4222164, 4222135, 4222145, 4222161, 4365653, 4365668, 4439643, 4439657, 4365627, 4492210, 4105161, 4105112, 4105181, 4198757, 3998764, 4050426, 4300281, 4198766, 4439630, 4050415, 4365656, 4105101, 3892033, 4105122, 4346516, 4222140, 4500739, 4105171, 4198747, 4365651, 4346517, 4365646, 3936237, 4439633, 4439632, 4492232, 4492218, 4222150, 4222152, 4222177, 4198723, 4198734, 4198755, 4222160, 4346513, 4346484, 4346494, 4346497, 4346503, 4365643, 4365625, 4365638, 4365655, 4365663, 4365670, 4105172, 4105144, 4050436, 3892027, 4365657, 4105083, 4222148, 4105097, 3843863, 4050431, 3936261, 4105149, 3962183, 3962159, 4222182, 4222181, 4198762, 4198721, 4198759, 4300285, 4300305, 4365636, 4365632, 4365628, 4365662, 4198756, 2376459, 2427331, 2451634, 2609736, 2376460, 2293474, 2512976, 2609706, 2400178, 2512985, 2488824, 2609734, 2293505, 2843347, 2823730, 3774844, 2823700, 2973255, 2823715, 2843348, 3662517, 2843351, 2823709, 3695784, 3695780, 2564824, 2788529, 2823716, 2823733, 2823718, 2843321, 3662527, 2843315, 3695756, 3623206, 2973202, 2647181, 2973182, 2564789, 2919587, 3591107, 3741434, 3741436, 2788535, 3662522, 3695795, 2919576, 4492215, 4300303, 4500731, 4439631, 4105162, 4439635, 4500749, 4492223, 4500748, 4346509, 4439654, 4492214, 4439634, 4365667, 4346518, 4198727, 4050432, 4198726, 4365669, 4500737, 4439644, 4365674, 4500736, 3962175, 4439647, 4346485, 4478139, 4198745, 4478156, 3962192, 4346477, 4198731, 4094932, 4050414, 4500743, 4346495, 4500745, 4094945, 4300271, 3892035, 4478155, 4094917, 4300268, 4222159, 3874440, 4492228, 3874439, 4492211, 4478140, 4105089, 3998750, 3998761, 4050440, 4478150, 4094949, 4094951, 4105125, 3936267, 4346506, 4439670, 4500738, 3892030, 4198729, 4198749, 4300272, 3998778, 3936224, 4492212, 3874438, 3962196, 4094946, 4439652, 4198742, 4050405, 4300274, 4300279, 4300273, 4478138, 4094957, 4365639, 3892023, 4105168, 4365659, 3892031, 4050387, 4439674, 4105176, 4050413, 4105092, 4346491, 4500728, 4346508, 4346498, 4500741, 4492209, 4492230, 4105180, 4094959, 4198754, 4500753, 4492221, 3843865, 4105152, 3962172, 3936253, 4365630, 4500734, 4439636, 3843888, 3843872, 3962179, 4365640, 3843876, 4105155, 4050394, 4105140, 3936236, 4105151, 4852048, 4896966, 4896975, 4707164, 4896979, 4896958, 4728253, 4852034, 4823903, 4823884, 4896964, 4925884, 4896956, 4852064, 4852036, 4707173, 4794321, 4754827, 4810032, 4847689, 4852035, 5064236, 4728250, 4728247, 5064230, 4975687, 4975670, 4925871, 4728249, 4810047, 5064264, 5064256, 4823904, 4788507, 5018934, 4852050, 5066343, 4810038, 4823885, 4896968, 4975692, 4810041, 4823886, 4896963, 4975689, 4852061, 4967787, 4975682, 5074969, 4707155, 4823881, 4847680, 4925869, 4810051, 4967783, 5064232, 5064212, 4707175, 4707168, 4794305, 4810039, 4810033, 4852047, 4925876, 4925878, 4975693, 4975681, 4852042, 4852060, 4852038, 5066330, 5066344, 5066337, 5134733, 5066345, 5064223, 4847683, 5066346, 4794311, 4728264, 4967797, 4754826, 4754843, 5066350, 4896953, 4707162, 4847668, 4788516, 4967794, 5074965, 5165655, 5066327, 4788511, 4823891, 4872253, 5154802, 4810049, 4707157, 4847686, 4728274, 4896981, 4896974, 4847673, 4847695, 4925865, 4896972, 4754850, 4975688, 4788506, 4794320, 4788496, 5066353, 5064233, 5066352, 4823902, 4823898, 4794303, 4728271, 4896980, 4896957, 4754842, 4823895, 4823896, 4967792, 4728254, 4872235, 4967799, 4794317, 4975668, 4975683, 4810056, 4810040, 4707163, 5134714, 4794308, 4788518, 4810055, 4847674, 4788513, 4823889, 4872238, 5066356, 4810045, 4896971, 5064209, 5152400, 4925863, 4754852, 4852057, 4967795, 5018929, 4975695, 4872240, 5064244, 4925867, 4967782, 4823906, 4707178, 4975674, 5134685, 4754847, 4847681, 5066317, 5066316, 4754834, 5066318, 4794314, 4707180, 4810053, 4810036, 4810052, 4788500, 4788509, 4810046, 4896962, 4788517, 4823888, 4852046, 4872229, 4925862, 4872233, 4872257, 4925870, 4925881, 4925889, 5064211, 5064263, 4847691, 4925890, 4925891, 5064242, 5064216, 5064240, 5064241, 5152396, 5152394, 5134731, 4754844, 4728272, 4810050, 4925872, 4794307, 4852049, 4810044, 4847667, 4852031, 4728257, 4728259, 4847684, 4896967, 4872242, 4925864, 5154799, 4728251, 4754839, 4788498, 4872244, 5152420, 4788505, 4852032, 4852056, 5152410, 5152419, 5154801, 5152395, 5152416, 5165660, 5165670, 4728265, 4728269, 4754846, 4754836, 4788501, 5152406, 5154807, 4852045, 4852063, 5066311, 5074972, 5066331, 5134713, 5066351, 5066354, 5064224, 5064257, 5066309, 4872237, 4925875, 4788491, 5064225, 4707174, 5064215, 4872248, 4707159, 4925882, 4707161, 4794304, 4728252, 4847671, 4754831, 4707170, 4810054, 4754841, 4707165, 4794318, 4794299, 4728255, 4754837, 4754838, 4823908, 4847692, 4852043, 5064218, 4728273, 2609760, 2609730, 3774853, 2823712, 3623224, 2973239, 3571615, 2973259, 3623207, 3662523, 3623228, 3695781, 2973258, 4365626, 4198715, 4346476, 4346475, 3843893, 4439645, 4050395, 3998744, 4300298, 4439668, 4094916, 4094962, 4198765, 4439671, 4050401, 4198761, 4050422, 4105117, 4034715, 4439656, 4365671, 3936262, 4478154, 4105160, 4094942, 4198733, 3936246, 4492234, 4478146, 3874449, 4492226, 3874435, 3843877, 4500733, 4365631, 4300300, 3874444, 4439673, 4300290, 3998770, 4300291, 4439640, 4478157, 3936260, 4365664, 4346492, 4105123, 4300299, 4050428, 4105130, 3874433, 4500751, 4050412, 4105175, 4222176, 4105109, 3936235, 4198764, 4105142, 4105146, 4300269, 4050434, 3998775, 4050427, 4105095, 4105170, 3874461, 3874478, 4754824, 4872254, 5154823, 4872258, 5232595, 5220677, 5018938, 5165664, 5064221, 5152408, 5191847, 5239234, 4852058, 5165668, 5165646, 5066322, 5239248, 5260318, 5229850, 5232610, 5229843, 5066362, 5220687, 5220658, 5137413, 5239255, 5134678, 5260330, 4810034, 5220685, 5066364, 5064259, 5239243, 5239238, 5220675, 5064238, 4810035, 5239253, 5152402, 5229825, 5232586, 5260337, 5018937, 5152401, 5074963, 4925877, 4847694, 4872243, 5260343, 5220670, 5137397, 4975691, 4967796, 5152398, 5260338, 5229851, 5232583, 5066363, 5165665, 5239233, 4872234, 5260313, 5191844, 5260315, 4896959, 5018930, 5229840, 5137412, 4896960, 5066338, 5191849, 5229833, 4810030, 4847677, 5066307, 5229834, 4728258, 5018940, 5152403, 5137398, 5134686, 5064208, 5165659, 5239247, 5260327, 5239240, 5229823, 5220672, 5220671, 4925883, 4847688, 5137395, 5018924, 5066326, 5191848, 5239235, 5134728, 4975685, 4975675, 4823894, 5260339, 5232596, 4823880, 5165657, 5066347, 5154816, 5220664, 5165654, 5220659, 5018933, 4896954, 5154798, 4788502, 5165647, 5232598, 5260314, 5134703, 5134701, 4852053, 5232591, 4975690, 5165658, 5152415, 4925887, 5165643, 5154806, 5134680, 5074967, 5260333, 5134694, 5229845, 5232606, 4728246, 5064229, 5064254, 4823905, 5220678, 5239258, 5134679, 5134688, 5239251, 5064251, 4788494, 5134732, 5232608, 5064253, 5064249, 5152412, 4788515, 5066332, 5018931, 5134693, 4788499, 5018945, 5191840, 5066349, 5239237, 5134684, 5066340, 5152399, 5191829, 5154813, 5066365, 4975684, 5137402, 4975676, 5066355, 5154811, 5018932, 5229842, 5260331, 4847693, 5154819, 5134681, 5018943, 4967785, 5239239, 5260334, 4794295, 5154814, 5220669, 4810037, 5134709, 4852041, 4975678, 5064246, 5066321, 5232593, 5165661, 5220661, 5260335, 4967793, 5152414, 4788510, 5064237, 5064220, 5066335, 5229824, 4728268, 5191830, 4788493, 5165645, 5191838, 4810028, 5134696, 5229852, 4925873, 5134711, 5134675, 4975672, 4872251, 5064262, 5134718, 4975669, 5165652, 5134683, 4788492, 4707182, 4823887, 4794312, 5220681, 5191832, 4794297, 4967784, 5229828, 5066329, 5239257, 5154808, 5154815, 5154821, 5232611, 5229829, 4728261, 4754833, 4852040, 5137394, 5220676, 5229827, 5074966, 5064239, 5165653, 4852033, 5018926, 5229844, 5239232, 5232592, 5232604, 5137407, 5137403, 5152417, 5220674, 5220683, 5232582, 5232585, 5260323, 5260326, 5260332, 5229849, 5134699, 5260342, 5260341, 5018939, 5152418, 5152393, 4925885, 4847682, 5152397, 5064226, 5066312, 4847672, 4872231, 5165662, 4975680, 5134698, 5134723, 5229839, 4872255, 5165672, 5165667, 5074971, 5191843, 4872247, 4925868, 5191833, 5066325, 5154809, 4707167, 4810027, 5134724, 4754853, 5191845, 5191831, 5152413, 5220663, 5165649, 5239256, 3662532, 3662516, 3662519, 3741432, 4105139, 4105129, 3936254, 4500747, 4105133, 4105157, 4365652, 4492227, 4300294, 3962195, 4198728, 4492237, 4094921, 4439641, 4439663, 4300287, 4094956, 4198750, 3892025, 4094926, 4198753, 4500726, 4492224, 3962170, 4222173, 3843878, 4094938, 4198724, 3874475, 5018927, 5191852, 5239254, 5154803, 5239245, 5154817, 5066315, 5239260, 5232597, 5066328, 5232601, 5232603, 4728248, 4852059, 4975673, 5232605, 4810043, 4823907, 5066339, 5220668, 4975686, 5260311, 5064261, 4872241, 4707156, 5134729, 5191851, 4823892, 4728270, 5018942, 5134700, 5220682, 4728266, 4967789, 5229832, 5137410, 4754848, 4975667, 5066361, 4810048, 4794323, 4823883, 4794316, 5229846, 5260329, 5152411, 5260321, 5152405, 4794300, 4925866, 5137408, 5260328, 5232600, 5165644, 5232588, 5229826, 5066334, 5260317, 5191834, 5134674, 4788514, 4794294, 5191835, 4810029, 4707184, 5137404, 5260340, 5064227, 5134715, 4754845, 4852037, 5134705, 5134710, 5134704, 5260316, 5239241, 4852052, 4728275, 5165669, 4925879, 5191841, 5064219, 5066323, 5064207, 5232599, 5064213, 4728256, 4896978, 5074968, 4794296, 5152404, 5064260, 4754832, 4794302, 4754835, 4852051, 4794306, 4896976, 5975360, 6112517, 5870269, 5910994, 6021941, 5869767, 5869778, 5911018, 5975358, 6021946, 6023855, 6073216, 6178778, 6112484, 6215537, 6178755, 6197116, 5868551, 5872091, 5975347, 6178780, 5869786, 6178757, 6073211, 5910968, 6079839, 5910992, 6073254, 6178754, 6112537, 5903036, 6073208, 6178760, 6112553, 6178776, 6178779, 5910967, 5869775, 5910977, 6023824, 6112523, 6112528, 6021953, 5903042, 5910979, 5910740, 6112521, 6112532, 6197119, 5870244, 6023842, 5911020, 6023848, 5868528, 5903050, 6023826, 6207384, 5870264, 5975339, 5869789, 5910996, 5975366, 5872088, 6023836, 5975309, 6215533, 6112548, 6112491, 6112533, 5868539, 5868546, 5911021, 6023835, 5910976, 6021933, 6072724, 6073215, 6112552, 6112506, 5868548, 5868529, 5870247, 5910750, 5911024, 5910985, 5975320, 6072736, 6112549, 6178756, 6112489, 6112526, 6178731, 6207397, 6023827, 5910984, 5868536, 5870259, 5870255, 6178763, 5872092, 5975315, 6072737, 6178743, 5975340, 6178770, 5869765, 5872087, 5869760, 5910997, 5910973, 5975348, 6073210, 6072735, 5869769, 6023860, 5975334, 6197118, 6023854, 6079833, 5911002, 6197134, 5910981, 6023813, 5910975, 5975319, 5911011, 5975351, 5975364, 6021952, 6072746, 6021955, 6021956, 6073256, 6023818, 6023832, 5910728, 6112487, 6197115, 6112536, 6073255, 6072725, 5910725, 6073248, 5870253, 5975359, 6178766, 5868547, 5872100, 5872085, 5870260, 5870263, 5903037, 5903044, 5903049, 5903040, 5903046, 5903047, 5975342, 6023834, 6073241, 6023809, 6023819, 6023803, 6023804, 6023805, 6023820, 6023807, 6023839, 5975357, 6079842, 6079845, 6112540, 6178758, 6178785, 6023857, 6178786, 6207407, 6207390, 6207401, 6207405, 6207418, 6073253, 5868527, 5868544, 5868549, 5868554, 5868531, 5868545, 5868538, 5870256, 5872095, 5868533, 5872081, 5870261, 5910735, 5975310, 5975312, 6178730, 5869780, 5910998, 5911000, 5911009, 5911015, 5910969, 6021951, 6021932, 6021935, 6021945, 6021948, 5911022, 5910731, 6023825, 6023829, 6023814, 6023840, 6023858, 6072726, 6072729, 6072731, 6072741, 6072745, 6112543, 6207398, 6215549, 5872096, 5910991, 5911004, 6021938, 6021944, 6178732, 6178775, 6178735, 6178769, 6215544, 5974714, 5974694, 5910989, 6178777, 5870267, 5910987, 5910993, 6112480, 5975343, 5974697, 5910738, 5870241, 6112530, 5974712, 5975338, 5910746, 4094913, 4300261, 4300267, 4300288, 4198720, 4346511, 4105131, 4439672, 4478151, 4492229, 4492238, 4105177, 5220660, 5137409, 5154804, 4896977, 5066359, 4847666, 4872246, 4754825, 5239244, 4975666, 4975677, 5239246, 4872249, 5064265, 5134722, 5064235, 5220662, 5064247, 4852054, 4967790, 5137406, 5232602, 4896982, 5154820, 5064243, 5137401, 4754829, 5137399, 5152407, 5232584, 5066324, 5066336, 5191837, 5191846, 4872236, 5260336, 4925888, 5154810, 5074970, 4754830, 4975671, 6079843, 6325767, 6325772, 6112550, 6289836, 6197120, 6498685, 5910726, 6394667, 6215543, 6325777, 6315802, 6325756, 6112494, 6394677, 6394705, 6512574, 6178761, 6265475, 6315797, 6215536, 6315807, 6207409, 6394670, 6289846, 6498682, 6498653, 6498677, 5910990, 6512578, 6178774, 6178740, 6197117, 6112493, 6315808, 6265463, 6315793, 6325780, 6498686, 6498654, 6197124, 6215559, 6315800, 5868525, 5869784, 6178738, 6265469, 6178759, 6265467, 6394690, 6498618, 6394697, 6512580, 6265483, 6325789, 6073205, 5975316, 6023844, 6112524, 6325764, 6330591, 6325757, 5911007, 6512583, 6498664, 6112544, 5910999, 6498615, 6325795, 6215558, 6178748, 6215570, 6315796, 6315805, 6197140, 6498678, 6112509, 6112522, 6215572, 6512554, 6112518, 6394685, 5870250, 6178767, 6512588, 6215573, 6394706, 6073207, 6178742, 6315795, 6325790, 6512596, 6512571, 6215587, 6207400, 6197112, 6112534, 6112496, 6394715, 6330583, 6498675, 6315789, 6178781, 6330580, 6394682, 6073252, 6215538, 6394708, 6178784, 5975356, 6112519, 6197138, 6498635, 6073259, 6112490, 6289849, 6498688, 6498626, 6498683, 6394679, 6197136, 6178782, 6197125, 6315790, 6178765, 5869776, 6325786, 6197133, 6498689, 6512592, 6112512, 6112500, 6512609, 6512598, 6394663, 5910988, 6073239, 6215561, 6178752, 6330600, 6394674, 6498631, 5910751, 6289833, 6073258, 5869788, 6498642, 6394666, 6512565, 5868542, 5872082, 6394671, 5975362, 5910753, 6207416, 6112503, 6512555, 6072738, 6289835, 6325775, 6265461, 6330584, 6289839, 6215542, 6289825, 6498655, 6215552, 6112482, 6207392, 5903043, 5975346, 6394668, 6215548, 5910741, 6215554, 6289840, 6207387, 6394692, 6512581, 6512582, 6023861, 6178753, 5868550, 6023823, 5975349, 5975361, 6207391, 6197139, 6289826, 6197130, 6394710, 6498647, 6197127, 6498671, 6023812, 5872090, 5974721, 5974706, 6215588, 6207412, 6112486, 6512557, 5975331, 6112488, 6498650, 6265474, 6498660, 6498638, 6073250, 6512576, 6325787, 6512551, 6498636, 5910729, 6289852, 6023849, 6330574, 5869781, 6215560, 6265459, 6330587, 6021934, 6289837, 6289838, 6325776, 6325779, 6325782, 6325791, 5911003, 6330586, 6330597, 6512600, 6215575, 6215582, 6265471, 6207388, 6207415, 6289844, 6325774, 6325753, 6498625, 6394681, 6394684, 6394688, 6394696, 6498670, 6498681, 6498690, 6498612, 6498613, 6498622, 6498628, 6498639, 6498644, 6498648, 6498649, 6498657, 6512613, 6315781, 6325751, 6325788, 6512606, 6330596, 6215562, 6215585, 6112495, 6315782, 6207413, 6112531, 6207386, 5903045, 6021940, 5911006, 6315784, 6112525, 6207396, 6325763, 6394680, 6289831, 6178733, 6498669, 6112542, 6112545, 6512607, 6330588, 5975308, 6207389, 6178771, 6197128, 6215531, 6178746, 6072730, 6073203, 6021936, 6325792, 6325765, 6197114, 6265458, 6112527, 6315804, 6072727, 6207399, 6215534, 6498672, 5869783, 6178729, 6498676, 6265464, 6112515, 6178772, 6265473, 6265477, 5911017, 5870262, 6394704, 6567930, 6567971, 6566916, 6567972, 6566913, 6566903, 6567981, 6567938, 6567948, 6566888, 6575646, 6473764, 6566937, 6575607, 6734562, 6473789, 6567922, 6575650, 6584157, 6584171, 6473770, 6567949, 6567976, 6566919, 6575614, 6567963, 6567924, 6566886, 6567978, 6702865, 6566921, 6566938, 6575645, 6567921, 6567951, 6567980, 6566908, 6567955, 6567968, 6473767, 6566929, 6473768, 6566933, 6566935, 6575600, 6575597, 6575593, 6567960, 6566901, 6567920, 6567928, 6567954, 6584165, 6473769, 6566891, 6575638, 6734563, 6566928, 6566910, 6473777, 6473778, 6567937, 6473779, 6473788, 6473776, 6566893, 6567926, 6566885, 6566926, 6566934, 6566902, 6575588, 6575591, 6575609, 6575615, 6575622, 6575627, 6575639, 6575642, 6575643, 6575644, 6575594, 6575605, 6702869, 6567931, 6567961, 6566914, 6566932, 6566940, 6702863, 6702868, 6734540, 6734549, 6702864, 6473765, 6566892, 6566899, 4439660, 4478141, 5018936, 5191839, 4967786, 5066314, 5066348, 5137405, 5134695, 5191842, 5018944, 5239249, 5066360, 4728260, 5134719, 4788519, 4852039, 4794315, 5066319, 6512559, 6079852, 5974716, 6079848, 6512593, 6265482, 6079851, 6394702, 6512585, 5910980, 6512591, 6498652, 6112529, 6325784, 6265460, 6079846, 5910974, 5974705, 6207411, 6023845, 6498667, 6215541, 6315799, 5910749, 5869770, 6215586, 6325794, 6112535, 6330592, 5903041, 5975335, 6023852, 6498630, 6512573, 6072732, 6498620, 6215590, 6073213, 6073214, 6325752, 6325760, 6072733, 6512575, 6325761, 6073198, 6394691, 6073206, 6394687, 6079834, 6079837, 6325754, 6512594, 6394712, 6207395, 6197122, 6215550, 5868526, 6023816, 5870268, 6498617, 6330594, 5975345, 5975327, 6498656, 6079838, 6512589, 6197131, 6112510, 6315792, 6207403, 5974695, 6498627, 5870265, 6215571, 6315798, 6197132, 6498662, 6512568, 6315785, 5911001, 6178741, 6325758, 6512595, 6178739, 5903039, 6325778, 5975329, 5870245, 6498661, 6498641, 6079841, 5910744, 5910972, 5910971, 6215530, 6512558, 6112547, 6315787, 5870242, 6178737, 6023859, 5975313, 5870251, 6567965, 6734551, 6825762, 6575632, 6825731, 6897997, 6575620, 6898009, 6898007, 6936212, 7011471, 6991104, 7011442, 6734052, 6898003, 6897988, 6897994, 6886463, 6566915, 7011473, 6567950, 6936222, 6734000, 6584151, 6734039, 6734055, 6825767, 6734037, 6986890, 6825770, 6991111, 6575637, 6936213, 6897974, 6734526, 6575628, 6566883, 6584172, 6734021, 6734047, 6734048, 6734032, 6825758, 6870216, 6897987, 6825745, 6897991, 6897998, 6999294, 6936219, 7011457, 7011447, 6575596, 6991108, 6897960, 6575589, 6870214, 6734564, 6567977, 6567925, 6734016, 6734023, 6734018, 6886484, 6897981, 6575587, 6897959, 6567985, 6897970, 7011444, 6575608, 6733997, 6897983, 6898002, 6999318, 6999310, 6999286, 6566931, 6886502, 6886452, 6575606, 6734015, 6825733, 6886460, 6886464, 6825779, 6886473, 6897996, 6986886, 6986923, 6886459, 6991120, 6999297, 6575603, 6897955, 6897953, 6734034, 6936209, 6825771, 6567958, 6734043, 6734003, 6734014, 6575648, 6936223, 6584177, 6584169, 6786661, 6986883, 6734005, 6566924, 6567947, 6575592, 6567929, 7011451, 6986925, 6999309, 6897984, 6575641, 6575599, 6734040, 6991106, 6999274, 6897958, 6584164, 6936211, 6898008, 6986911, 6986906, 6886467, 6575612, 6734046, 6734011, 6986895, 6999270, 7011462, 6886503, 6999273, 6999293, 6566896, 6786652, 6734022, 6897982, 7011459, 7011461, 6734051, 6825768, 6870221, 7011458, 6825732, 6897980, 6575625, 6734054, 6825747, 6786658, 6734566, 6897969, 6991103, 6991117, 7011464, 6575618, 6734514, 6566897, 6986914, 6991109, 6734567, 6897962, 6575613, 6734058, 6986931, 6936226, 6584153, 6734519, 6734007, 6734557, 6786665, 6825739, 6897979, 6999298, 6584160, 6986884, 6566890,

More Code Info

6566894, 6734053, 6734045, 6936220, 6986924, 6870213, 7011455, 6575619, 6734025, 6886465, 6886491, 6897992, 6584162, 6733998, 6734024, 6473782, 6734013, 6734516, 6825773, 6734511, 6566911, 6567959, 6734027, 6567984, 6999296, 6825783, 6825782, 6734523, 6734017, 6734030, 6734031, 6575610, 6897986, 6897995, 6898001, 6999322, 7011450, 6734049, 6986907, 6986929, 6986930, 6936218, 6786664, 6897961, 6897966, 6897967, 6986899, 6986877, 6734033, 6734028, 6734035, 6734061, 6825754, 6870217, 6870222, 6870225, 6870226, 6870227, 6886450, 7011446, 7011466, 7011475, 6734560, 7011469, 6999292, 6999301, 6734533, 6825735, 6825755, 6825772, 6825730, 6825778, 6897972, 6999271, 6999276, 6999308, 6999319, 6999321, 7011456, 7011460, 6786666, 6786667, 6886451, 6825753, 6886457, 6567941, 6986875, 6473766, 6566941, 6999289, 6566925, 6886445, 6567923, 6584150, 6567953, 6870218, 6566927, 6936214, 6734546, 6567940, 6734535, 6584149, 6566884, 6567933, 6734547, 6898006, 6567973, 6936217, 6567919, 6567939, 6567935, 6566905, 6786656, 6734530, 6886487, 6734537, 6734524, 6567946, 6786654, 6786671, 6734553, 5239231, 5064214, 5260309, 5239252, 5239259, 5134697, 5064252, 5134702, 5152409, 5066310, 5064255, 5066308, 5134676, 5229837, 6215581, 6289851, 5974717, 6079840, 6215580, 6021949, 6023851, 5868537, 6289850, 6394713, 6215568, 6197113, 5974698, 6330595, 6394664, 5974699, 6394669, 5911019, 6023810, 6315786, 6394707, 6512605, 5869782, 6315783, 6215546, 6023841, 6112554, 5872093, 6394701, 6289829, 6112504, 5868553, 5910986, 6197126, 6315806, 6072743, 6197141, 6207393, 5975307, 5869764, 6178768, 6079836, 6498614, 6325766, 5868532, 6079854, 6325793, 6112501, 6112485, 6215583, 6215576, 6073242, 5975332, 6325768, 6330590, 6394693, 6021942, 6289847, 6289830, 5870266, 6207414, 5975363, 5974711, 6498673, 6079835, 6289842, 5910982, 5868534, 6289824, 5868535, 5975311, 5870257, 6265476, 6512602, 6498619, 6999288, 6575647, 6567952, 6999306, 6936221, 6999302, 6999305, 6999303, 7011449, 6999311, 6897964, 6886461, 6999325, 6999282, 6999299, 6999285, 7011478, 6734042, 6999316, 6999324, 7011452, 7011477, 6886499, 6473771, 6567956, 6897973, 6999315, 6575602, 6936210, 7011470, 6986928, 6786653, 6825728, 6999317, 6999281, 6999320, 6991110, 6886448, 6584158, 6734010, 6473787, 6986921, 6825741, 6734521, 6825737, 6566907, 6825759, 6986912, 6584152, 6567975, 6473791, 6575601, 6897978, 6734509, 6734026, 6886479, 6825750, 6897993, 6886458, 6986882, 6986889, 6898000, 6566939, 6566904, 6886466, 6936216, 6986892, 6936215, 6986916, 6575611, 6991115, 6734041, 6999278, 6897957, 6734548, 6986887, 6584176, 6575616, 6566930, 6886469, 6886468, 6734550, 6886447, 6870210, 6734532, 6734555, 6567969, 7011443, 6986926, 6986908, 6870219, 6330573, 6330602, 6289848, 6215574, 6178734, 6112499, 6197123, 6112502, 6394673, 6498679, 5869779, 5910733, 6289845, 6498629, 6207408, 6289853, 5974701, 6498674, 5870249, 6512561, 6325762, 6073244, 6289834, 6207404, 6330585, 6073251, 6112481, 6289828, 6498687, 6394711, 6512604, 6215540, 6023833, 6112551, 6512550, 6178751, 5975355, 6215551, 5870258, 5870246, 6112539, 6112541, 6512570, 5910732, 6315779, 5868530, 6325759, 6734520, 6936225, 6575633, 6870215, 6734536, 6999291, 6825744, 6991121, 6886486, 6473774, 6986918, 6566920, 6825777, 6575590, 6584163, 6825757, 6999314, 7011467, 6567936, 6733999, 6886476, 6734517, 6473775, 6566936, 6986898, 6584155, 7011480, 6825769, 6897985, 6986896, 6999287, 6825736, 6886498, 6734044, 6786655, 6825774, 6986878, 6986933, 6870211, 6734038, 6575626, 6986894, 6999295, 6825740, 6734538, 6986919, 6886504, 6786657, 6886483, 6886446, 6986910, 6567967, 5975322, 6473783, 6734002, 6897956, 7011468, 6870220, 6584159, 6734534, 6567957, 6986902, 6734508, 6786660; b) Item Number 320-38-13, UDI/DI 10885862086686, Serial Numbers: 1032660, 0982202, 0982217, 0982225, 1032662, 0982234, 0982205, 0982206, 1032653, 1095410, 1095714, 1095745, 1032644, 1032680, 1095403, 1095412, 1095419, 1095421, 1095422, 1032676, 1095423, 1095742, 1032654, 1095427, 1032674, 0982230, 0982212, 1095430, 0982203, 1032651, 1032672, 1095382, 1095752, 1095727, 1095386, 0982216, 1095415, 1095416, 1032678, 1095393, 1095395, 1095400, 1095408, 1095409, 1095718, 1095741, 1095751, 1095760, 1095381, 0982208, 0982211, 1349310, 1349311, 1349329, 1032665, 0982219, 1095732, 1095730, 0982232, 1032667, 0982224, 1095739, 1032647, 1032671, 1095417, 1095401, 1095749, 0982207, 0982218, 0982228, 1095396, 0982215, 1095755, 1032646, 1032648, 1095398, 1095756, 1095424, 1445354, 1437789, 1349321, 1349305, 1445350, 1349317, 1349306, 1349327, 1408444, 1330599, 1437784, 1445351, 1532929, 1482628, 1532935, 1482625, 1330605, 1482621, 1532920, 1532921, 1408432, 1408425, 1330603, 1349322, 1330607, 0982223, 1095428, 1095429, 1095413, 1095750, 1032652, 1095715, 1095388, 1032679, 1032668, 0982213, 1095728, 1095399, 1437795, 1445363, 1437781, 1408430, 1482627, 1512859, 1482615, 1330596, 1437787, 1437780, 1445345, 1437778, 1512851, 1545421, 1545420, 1437782, 1437776, 1482633, 1349324, 1532936, 1545407, 1545409, 1512858, 1330606, 1512848, 1330602, 1445346, 1349302, 1688491, 1784422, 1763407, 1688476, 1912606, 1688493, 1912601, 1912603, 1784449, 1912605, 1922793, 1922794, 1784430, 1784440, 1688479, 1763424, 1779334, 1688492, 1784445, 1688488, 1763427, 1763442, 1912608, 1688471, 1688490, 1688497, 1688473, 1779319, 1095713, 1032656, 1095731, 1095414, 0982226, 1095378, 1032670, 1095392, 1095725, 1095753, 1032661, 0982235, 1095379, 1095406, 1408446, 1408448, 1349332, 1545414, 1545408, 1437793, 1349318, 1437791, 1512853, 1445361, 1482616, 1545411, 1532917, 1330595, 1330601, 1482614, 1532931, 1330604, 1512854, 1349309, 1532927, 1482630, 1512845, 1482632, 1482629, 1512856, 1482620, 1763439, 1779316, 2049214, 1779322, 1784438, 1779327, 1912615, 1912619, 1688483, 2049234, 1779320, 1688482, 1763415, 1763437, 1779314, 1779315, 1779332, 1779333, 1779336, 1784429, 2010965, 2010966, 2010967, 2010968, 1688481, 1763405, 1779324, 1779325, 1779328, 1779330, 1779331, 1779337, 1784428, 1784435, 1784437, 1784446, 1922803, 2049195, 2049196, 2049235, 2049207, 2049223, 2049224, 2049191, 2049193, 2049201, 2049205, 2049211, 2049228, 1922796, 1784450, 1922800, 1784444, 1784423, 1922799, 1922792, 1784426, 1763433, 2049189, 1763410, 2117600, 2117604, 2108025, 2108035, 2108036, 2117596, 2117598, 2117599, 2117601, 2117607, 2117597, 2117626, 2168170, 1095746, 1095426, 1095404, 0982229, 1349308, 1408445, 1349300, 1408442, 1349330, 1512846, 1349301, 1545404, 1445362, 1330597, 1545415, 1349315, 1512855, 1408440, 1545416, 1349312, 1349319, 1482626, 1445360, 1545413, 1532930, 1545417, 1532926, 1512843, 1408438, 1779317, 1784424, 1688485, 1688474, 2010964, 1763423, 1763420, 1912612, 1784443, 1922795, 1912604, 1784421, 1688475, 1922787, 1912618, 1912616, 1763417, 1784425, 1922786, 2049209, 1763418, 2010963, 2049194, 2049188, 2049202, 2049230, 2049203, 2431351, 2369628, 2195686, 2168173, 2108027, 2195695, 2108037, 2168171, 2117627, 2168188, 2108022, 2108026, 2117620, 2369633, 2195673, 2168190, 2117602, 2409561, 2278452, 2320265, 2369608, 2320274, 2168191, 2369631, 2117625, 2195672, 2320257, 2108033, 2195685, 2168179, 2168180, 2168192, 2195680, 2195691, 2168172, 2278445, 2278463, 2117629, 2278453, 2278456, 2278457, 2108024, 2108032, 2108034, 2369617, 2409551, 2369605, 2369621, 2409542, 2409554, 2409553, 2117623, 2369599, 2369627, 2409549, 2409555, 2409557, 2369607, 2369613, 2409546, 2431344, 2431353, 2488847, 2431178, 2108028, 2320280, 2431357, 2488852, 2431198, 2278449, 2369604, 2369614, 2531969, 2369630, 2409559, 2431340, 2488837, 2117619, 2117603, 2369606, 2117612, 2278450, 2531963, 2195678, 2488830, 1349320, 1408427, 1408433, 1532923, 1482623, 1408428, 1330598, 1445356, 1482624, 1545403, 1482619, 1545405, 1532918, 1512840, 1445359, 1545406, 1512844, 1349314, 1482622, 1532933, 1784433, 1922789, 1688495, 1763411, 1922801, 1763426, 2049210, 2049186, 1688470, 1784431, 1784448, 1784432, 1779335, 1922797, 1912617, 2049221, 2049231, 2049199, 1912607, 1763434, 1912610, 1763430, 1922798, 1912602, 2049233, 1779329, 2488856, 2320278, 2108029, 2431338, 2195696, 2278469, 2488838, 2278460, 2431356, 2531960, 2409556, 2531958, 2195679, 2609769, 2488844, 2195677, 2278467, 2531961, 2431345, 2278464, 2369619, 2531973, 2488855, 2488833, 2609771, 2431187, 2431194, 2409560, 2609784, 2168184, 2369620, 2369612, 2369601, 2609887, 2369625, 2531966, 2431196, 2609780, 2609894, 2609876, 2278466, 2369632, 2609777, 2278454, 2117628, 2117621, 2431188, 2431197, 2609789, 2609791, 2609793, 2609796, 2609795, 2609891, 2609896, 2609897, 2609898, 2609875, 2609774, 2609889, 2609778, 2609893, 2609797, 2609790, 2609788, 2195683, 2195689, 2195671, 2168187, 2117614, 2195690, 2369618, 2488832, 2488846, 2369616, 2616302, 2771926, 2647210, 2641869, 2577485, 2641852, 2577517, 2641865, 2647214, 2616291, 2577508, 2577482, 2823740, 2843266, 2577493, 2647207, 2647198, 2577474, 2577515, 2792712, 2647209, 2616303, 2577498, 2577494, 2577478, 2577509, 2843302, 2577469, 2577476, 2577504, 2577518, 2641860, 2641863, 2641864, 2577477, 2577505, 2577500, 2616305, 2616290, 2616310, 2616287, 2616289, 2616293, 2616299, 2616307, 2616308, 2647201, 2792703, 2792704, 2792705, 2792706, 2792707, 2792708, 2823758, 2823770, 2577470, 2647212, 2647202, 2641849, 2641858, 2843296, 2641868, 2641856, 2577520, 2641853, 2771925, 2577519, 2616282, 1512852, 1445364, 1445349, 1512850, 1763431, 1784441, 1688487, 1688486, 1922805, 2049213, 2049216, 1912622, 1763406, 1912613, 1784442, 1784436, 1912600, 2010962, 1763441, 2049237, 2320267, 2431175, 2108019, 2431190, 2431185, 2320271, 2369603, 2609888, 2431182, 2488853, 2320273, 2409550, 2431192, 2431176, 2195682, 2431199, 2609890, 2431358, 2117616, 2320269, 2431177, 2168175, 2108020, 2278468, 2488836, 2168177, 2431350, 2609884, 2531979, 2609882, 2609775, 2369600, 2168169, 2431184, 2278465, 2278461, 2609776, 2117615, 2488841, 2488857, 2320268, 2843297, 2928393, 3622957, 2843270, 2843280, 2928422, 2928425, 2928399, 2928398, 2843259, 3534797, 2928390, 2843284, 3534812, 2928408, 3534803, 3707525, 3534823, 2771933, 2843290, 2928429, 3534843, 3622987, 3534828, 2843287, 2641850, 2843293, 2843300, 2647200, 2928402, 2928382, 3534806, 2928421, 2641861, 2928405, 2928407, 2792715, 2928416, 3646955, 2771935, 2928419, 2928423, 2928409, 2641862, 2641851, 2843275, 3683600, 2843268, 2928403, 2616286, 2771945, 2823759, 2577507, 3534802, 2647211, 2823782, 3646933, 3683601, 3534815, 2771936, 2928384, 3534835, 2928395, 2577481, 3622961, 2647197, 2771938, 2792701, 3534804, 2928406, 2823765, 2647215, 3534808, 2823781, 2928412, 2928424, 2771921, 2823761, 2616304, 3534837, 3622982, 3646970, 3707537, 3683596, 3622956, 2577487, 2843263, 3646932, 3646934, 2577479, 2577496, 2577499, 3622943, 2577483, 2771927, 2823742, 3646944, 2823752, 3646946, 3622971, 2771928, 2823764, 3646960, 2823778, 2928396, 3683607, 3534831, 2928388, 2843254, 2823771, 2577511, 2577490, 2928401, 3707529, 3755637, 3755644, 2843258, 2843265, 2843271, 2843278, 2843279, 2843285, 2843286, 2843253, 2843264, 2843281, 2823769, 3534794, 2616288, 2641848, 2641854, 3646948, 3646964, 2641859, 2641866, 3534821, 2843274, 2843288, 2843294, 2928411, 2928383, 3707539, 3622990, 2843289, 2928414, 2928417, 2928410, 3534829, 3622952, 3622966, 3646940, 3646965, 3646980, 3622981, 3622985, 3683595, 3683612, 3707530, 3534796, 2843269, 2843292, 2641855, 2577484, 2928400, 2823768, 2577486, 2843273, 2771930, 3622968, 2823784, 3622973, 2049217, 2049190, 2049227, 2049222, 2049197, 2049229, 2488831, 2488843, 2320266, 2320282, 2431346, 2117618, 2278455, 2409544, 2431336, 2117605, 2108031, 2278451, 2278458, 2278462, 2488845, 2409543, 2531972, 2117595, 2369629, 2609786, 2609770, 2278446, 2609879, 2609885, 2431355, 2278447, 2108021, 2609892, 2168174, 2320275, 2531967, 2488851, 2117617, 2609773, 3646973, 2792716, 3755654, 2928420, 3683605, 3707520, 3683590, 3738012, 3755649, 3683589, 3683604, 3775070, 3683594, 3738010, 3622994, 3755656, 3755657, 3755659, 3775062, 3707524, 3775079, 3622954, 3683599, 2928392, 3707519, 3534824, 3755638, 3622976, 3775058, 3775074, 3683593, 3622980, 3755653, 3775065, 3622983, 3534819, 3775064, 3622975, 3534832, 3775080, 3683597, 3755651, 3755647, 3534816, 2823748, 2843283, 3646968, 3622955, 2771918, 3775060, 2823767, 3622945, 3534839, 2616294, 3707536, 3534840, 2823783, 2616284, 2771919, 3622951, 2823772, 3646976, 3622940, 2843262, 3646953, 2771920, 2771939, 3775078, 3646939, 3646947, 3646951, 2823760, 2843291, 2647208, 2823766, 2577488, 2823741, 2616306, 3646938, 3707527, 2928428, 2577495, 3646937, 3534814, 2928413, 3646958, 2577480, 2577501, 2823751, 3534817, 3755636, 3646975, 3646979, 2771931, 2616311, 2792709, 2577492, 3534805, 3646977, 3707517, 3707521, 2928391, 2928397, 2928386, 2928394, 2928426, 3534826, 3646957, 3646967, 3755641, 3755635, 3707528, 3775059, 3775082, 3775061, 3775071, 3707522, 3755639, 2843276, 3738011, 3707538, 3755643, 3622977, 3738009, 3622967, 3622991, 3755648, 3775068, 3707526, 3683591, 2843295, 3707534, 3707531, 3755646, 3707515, 3937070, 3963200, 3937077, 3937045, 3963201, 3880834, 3880841, 3937036, 3937082, 3937044, 3963212, 3937041, 3880853, 3937081, 3937060, 4076654, 3880812, 4028500, 3937065, 4028474, 3880807, 3817335, 3817358, 3862336, 3937057, 3880810, 4028484, 4028492, 3937055, 4028496, 3817351, 4028486, 3862346, 3862341, 3862332, 3963193, 3817345, 3963232, 4028452, 3862345, 3862325, 3862339, 4028482, 3817355, 3880842, 4076678, 4028473, 3880859, 3817350, 3963228, 3880825, 3862342, 4136347, 4028489, 4136365, 4028487, 3880822, 3937049, 3963213, 4076662, 3937067, 3880833, 3880835, 3880847, 3880850, 3880852, 3880855, 3937043, 3937064, 3937054, 4076650, 3937038, 4076663, 3880860, 3880848, 3880808, 3880844, 3880851, 2049200, 2010970, 2195667, 2488839, 2168182, 2117609, 2609878, 2609880, 2488842, 2431348, 2609779, 2117611, 2531977, 2431352, 2609787, 2195675, 2409552, 2320260, 2320263, 2431180, 2117624, 2609772, 2609874, 2195688, 2278459, 2278448, 3775063, 2792717, 3646971, 3775075, 3775072, 3534838, 2616295, 3683598, 2577516, 2823755, 2843256, 3775081, 2771941, 3534798, 3534830, 2928389, 2577489, 2616296, 2823762, 3622969, 3622941, 2647204, 2843257, 2771932, 2647218, 2616301, 2823779, 3755640, 3622984, 2823753, 2792713, 3646966, 3775069, 3534799, 3707535, 2616300, 3622944, 2823737, 2771943, 3683611, 3755658, 3775067, 2823736, 3707516, 3622993, 3622974, 2843277, 3622986, 3534810, 3646974, 4214350, 4199405, 3880856, 4199406, 4230811, 4306526, 3880837, 3937072, 4306513, 4346540, 4439013, 3880849, 3963203, 4214370, 3880840, 4230822, 3963236, 4199389, 4230824, 3963196, 4365295, 4497749, 4214380, 4076652, 3963199, 4306505, 4365282, 4365321, 4439026, 3963211, 4346557, 3880836, 4439005, 3880858, 4214367, 3937047, 4199407, 4346533, 4365309, 4365294, 4365297, 3963221, 4028481, 4294489, 3963235, 4199398, 4214349, 4076653, 3937051, 4365305, 4365306, 4230823, 4214373, 4076667, 3937048, 3963226, 4214346, 4076656, 4136352, 4214388, 4028464, 3880866, 4346561, 4294475, 4294466, 4076683, 3880830, 3880864, 4136362, 4294485, 4365285, 4497741, 4199393, 4136364, 4028475, 4497743, 3862326, 4306522, 4076679, 4346526, 4136349, 4365320, 4076686, 4136360, 4497750, 4028451, 4199387, 3963216, 4136346, 3817344, 4294488, 4214376, 4136358, 4346569, 4365284, 4365308, 4365311, 4028462, 4214394, 4076697, 4365300, 4214379, 4306521, 4028467, 4497757, 3937085, 3862343, 3963229, 4076665, 4346564, 3880815, 4497738, 4076659, 3937083, 4365301, 4214345, 3880843, 4028478, 3880818, 4439012, 3880819, 4214393, 4214348, 4076698, 4214354, 3937068, 4230818, 4230814, 4214385, 4294468, 4028483, 4439014, 4306500, 3817357, 4076684, 4199391, 4346555, 4346556, 4214371, 4214374, 4439015, 4028466, 4365283, 4365328, 4214366, 3880857, 4028488, 3963191, 4136357, 4076675, 4294483, 3880861, 4076671, 4346566, 4136355, 3880865, 4294480, 4439019, 3963197, 4076695, 4076692, 4076664, 4199384, 4199394, 3937076, 4306502, 4306503, 4306518, 4230810, 4306504, 4199388, 4199392, 4199397, 4306506, 4306519, 4439008, 4497745, 4497736, 3963231, 4214353, 4294476, 4306510, 4365296, 4230820, 3963227, 4306497, 4365303, 4076674, 4076694, 4214351, 4439009, 3880839, 4439032, 4497739, 4214355, 4365310, 4439021, 4294465, 4294482, 4294486, 4306515, 4306516, 4306520, 4214372, 3817356, 4214383, 4306499, 3817352, 3817341, 3880817, 3880828, 3817346, 4076672, 3880826, 4294481, 4076676, 3937039, 4365289, 4214361, 4199390, 4346545, 4076681, 4076669, 4199401, 3963192, 3963224, 4136359, 3937058, 3880854, 4346534, 3880832, 4439020, 4439052, 4439017, 2609792, 2431360, 2369611, 2488834, 2531959, 2531976, 2531964, 2431342, 2431343, 2431349, 2431354, 2431179, 2369615, 2369622, 2609881, 2531957, 2369626, 3646959, 2843298, 3775066, 3622972, 3646963, 2616309, 3755642, 2843272, 3622949, 3646942, 2823739, 3622979, 2928385, 3646941, 2823776, 3646969, 3534818, 3622946, 3646954, 3755650, 3622958, 2641870, 3755655, 2577468, 3534820, 3534827, 4346565, 4214359, 4214363, 4230813, 4306507, 4214358, 3963195, 4346536, 4365333, 4365313, 4214386, 3963189, 4439035, 4365288, 4214389, 4439022, 4439039, 4076691, 3963210, 4076685, 4365302, 4346573, 4439027, 4199395, 4365323, 4230807, 4076670, 4199400, 4076651, 4439004, 4346574, 4076690, 4365307, 4076657, 4306508, 4497735, 4439010, 4439002, 4439018, 4028485, 4028453, 4346552, 4497759, 4346548, 4365319, 4076682, 4230826, 3862333, 4346530, 4028470, 3937079, 3880827, 4346537, 4214387, 4306517, 4028477, 3963230, 4365332, 4497744, 4136343, 3937075, 4214356, 4439050, 4497756, 4439043, 4365334, 4136353, 3817343, 4199385, 3817340, 4230808, 4346575, 4439023, 4346559, 4346546, 4346542, 4199383, 3963209, 3862338, 4365317, 4346560, 4346547, 4346525, 4346529, 4346544, 4076661, 4028468, 4346531, 4214382, 4230806, 4439034, 3963202, 4497753, 4306524, 4365324, 4439030, 4365292, 4028456, 4076666, 4439051, 3963206, 4365291, 4497758, 4439006, 4199399, 4439029, 4294484, 3963214, 4727445, 4834163, 4769376, 4788481, 4903768, 4973336, 4769386, 4810099, 5092539, 4973339, 4857865, 4769367, 4788461, 4881847, 4810085, 4769359, 4834168, 4973340, 4727437, 4788463, 4727429, 4834175, 4769368, 4834155, 5007429, 5160019, 4834166, 4727424, 4769366, 5024080, 4727422, 5007405, 4788469, 4810086, 4973318, 4727423, 4676464, 4769387, 4903748, 5075933, 4857849, 4788488, 4727449, 4788477, 4769374, 5076539, 4973341, 4676441, 4788465, 4834170, 4727430, 4810095, 5076545, 5007418, 4788464, 4881838, 4769378, 4903759, 5075947, 4903765, 4973324, 4881856, 5116041, 4810087, 4769373, 4676461, 5007423, 4834177, 4810097, 4973335, 4788486, 4903752, 4769362, 4834176, 4727443, 4769364, 4788475, 4810080, 4810103, 4881844, 5024085, 5092547, 4903750, 5075949, 4769384, 4788462, 4903767, 4810077, 4810104, 4676439, 4676442, 4676463, 4676466, 4676467, 5007427, 5007415, 5007428, 4857864, 5007432, 5007410, 5152260, 5152262, 4676458, 4788473, 4727440, 4769380, 4769375, 4676468, 5024096, 5092533, 5092538, 4881850, 5152255, 5152261, 4788484, 4834178, 5152253, 5152263, 4676455, 4676460, 4676462, 5024092, 4857873, 5007422, 4676459, 4903751, 5007426, 4727448, 4727444, 4769363, 4834173, 4788483, 3646962, 2823750, 3534822, 3683608, 3683610, 2823773, 2928380, 3775073, 3622970, 3707533, 3622947, 2823756, 3775076, 2823774, 3646961, 2843267, 3622989, 2823777, 3622978, 3755645, 2771934, 4214369, 4497746, 4438999, 4214377, 4497740, 4439045, 4497751, 4346532, 4076660, 4497752, 4439037, 4365290, 4028463, 4136366, 4365318, 4346571, 3937063, 4439047, 4076693, 4076680, 4028469, 4230819, 3862344, 3817354, 3880814, 3963223, 4439048, 3937069, 4076673, 4199386, 4497737, 4439040, 4214364, 3862331, 4439001, 4497731, 4497754, 3937071, 4028457, 3862348, 4136348, 4365298, 4365335, 4076655, 3862324, 4136367, 4076688, 4136356, 4230825, 3963215, 4028455, 4076687, 3862334, 4230812, 4214347, 3880829, 4294487, 3937053, 4306498, 3880824, 4439042, 4214360, 4306523, 4214368, 4294469, 4439038, 4199404, 3937059, 4076696, 3937046, 3937037, 4788468, 5092554, 5075948, 5024090, 5255147, 4973327, 5024082, 4903743, 5236172, 5207456, 5024075, 5236170, 5236137, 4788485, 4881861, 5160029, 5207455, 4881843, 4973333, 4727432, 5207467, 5152259, 5092531, 5160044, 4788472, 4903758, 5255144, 5075943, 4834158, 5207469, 4834159, 4973320, 4881855, 4676456, 4881863, 5160031, 5236140, 4857850, 4903764, 4834172, 5255141, 5236138, 5007407, 4727431, 5076541, 4857866, 4973321, 5255135, 5255161, 4857871, 4881839, 4903766, 5207448, 4727447, 4788487, 4857863, 4881860, 4834162, 4973326, 4676457, 5076554, 5116040, 5236143, 4810083, 5255163, 4857874, 5207468, 5075950, 4881866, 5160033, 4834154, 5024067, 5076531, 4903747, 4903754, 5255165, 5255160, 5092555, 4769385, 5092543, 5024089, 4727425, 4973325, 5007434, 4788459, 5207452, 4903744, 5076540, 5255125, 4881836, 4834179, 5207451, 4834153, 5236159, 4788470, 5255154, 4676446, 5255167, 4676447, 4727439, 5160025, 4903746, 5236155, 5236167, 5092553, 4903771, 4881852, 5075935, 5255149, 4769372, 5024070, 5116027, 5024076, 5160017, 5116032, 5207460, 5152258, 5007420, 5160038, 5152265, 4973314, 5236148, 5236133, 4973322, 4881859, 4881835, 5160022, 5116045, 4857862, 5207450, 5207462, 5160035, 5092550, 5116023, 5255136, 5255138, 5236149, 5255143, 5255166, 5236176, 4769377, 4788474, 4857867, 4769383, 5076546, 5255137, 4881840, 5236139, 5076530, 5236166, 5207458, 5255158, 5255146, 5152266, 5236132, 4727436, 4834165, 5255134, 4810106, 5152254, 5092540, 5236160, 5236162, 5024079, 5236169, 4857875, 5255123, 4727426, 5207453, 5152251, 5076526, 4769371, 5255126, 4810098, 4769388, 5255145, 5152264, 5207449, 4769370, 4676465, 4973337, 3646978, 3646931, 3622953, 3646935, 3534836, 4306511, 4306514, 3963187, 4365326, 4028459, 3937084, 4214381, 3880813, 4230821, 4136351, 4439025, 4346524, 3963198, 3817336, 4346551, 4346535, 4230809, 4028458, 4136350, 3880831, 3963225, 3963233, 4076677, 3862330, 4439033, 4214384, 3963217, 4497755, 4439028, 3937040, 4294477, 4306525, 4365286, 4214365, 4365330, 4214378, 4365327, 4788467, 5075938, 5152250, 4810102, 4810079, 5255156, 5236146, 4810100, 4810094, 4810101, 4727433, 4903760, 5007409, 4973315, 4769360, 5007406, 4857856, 5076537, 4676454, 5236171, 4834180, 5236141, 5236152, 5076528, 5255152, 4810105, 4973334, 4973338, 5092530, 5207454, 5092545, 5076535, 4857848, 5076534, 4834171, 5092551, 5236163, 5116022, 4727438, 4834164, 4857868, 5092557, 5076548, 5152257, 4788471, 5160026, 5236135, 5075951, 4857857, 5007417, 5236158, 5255151, 4676450, 4788460, 4903757, 5076536, 5116043, 5092552, 5092556, 4788478, 5236145, 5255127, 5255142, 5255150, 5075936, 5116033, 4810084, 4810082, 5092529, 4810081, 5076533, 5024068, 4857870, 5255122, 5160036, 5092532, 4727427, 4973317, 4881864, 5007408, 5874363, 5910675, 6084826, 5899070, 6084841, 5993633, 5910680, 5899071, 6044151, 6080562, 6084825, 6094877, 6188132, 6080565, 5910946, 6084847, 5936818, 6188128, 6044137, 5993637, 5936804, 6094867, 6112589, 6112597, 6112607, 6188137, 5936809, 6044162, 6112581, 6084830, 6079291, 6093625, 5910666, 6094857, 5899089, 5874239, 6084859, 5874229, 5899088, 6162023, 6094872, 5899066, 5899080, 5910959, 6084855, 6112575, 5910682, 5874222, 6080558, 6162024, 6215424, 5910659, 5937537, 6084831, 6094861, 5874357, 5910660, 5937535, 5993627, 5910942, 6084840, 6079292, 6079281, 5937522, 5899073, 6080566, 6084834, 5910676, 5874359, 5874217, 5936816, 6084851, 6188145, 5993620, 6094859, 5874212, 6112566, 5910657, 5899069, 6084849, 6093641, 5993628, 5874143, 6079286, 6079290, 5899095, 5936797, 5937515, 5874144, 5874151, 5910961, 5910965, 5899074, 5899077, 5899081, 5899093, 5874378, 5910655, 5910663, 5910670, 5910672, 5910678, 5910656, 5910669, 5910677, 6079287, 6079289, 6084848, 6084850, 6084853, 6093622, 6093631, 6112586, 6112587, 6112591, 6112595, 6112596, 6112611, 5937533, 6162038, 6162039, 6215508, 6215513, 6215521, 6215522, 5874350, 5874351, 5874137, 5874362, 5874211, 5874140, 5874141, 5899091, 5874138, 5910674, 5874372, 6079294, 5874375, 5936811, 6044142, 6044149, 6044158, 6094875, 5899083, 5993619, 5993635, 5993640, 6079285, 5910679, 5937510, 5937512, 5937530, 5937532, 5937534, 5910941, 6093620, 6093621, 6093623, 6093626, 6093628, 6093629, 6093637, 6093638, 6188129, 6188152, 5899085, 5937516, 5937527, 5910950, 5936791, 5937538, 5993616, 6094866, 5936808, 5874356, 5936792, 5936805, 5936796, 5874367, 5874136, 5937521, 5937520, 5910944, 5936806, 5993728, 5993730, 4439049, 4076668, 4439016, 4214352, 4497748, 4497742, 4028454, 4199396, 4214392, 4346553, 4306509, 4214390, 4306501, 4136361, 4346527, 4136354, 3862329, 3880811, 4199403, 4973316, 5236174, 4881832, 4903772, 4881851, 5255148, 5075942, 4834157, 4810090, 5152256, 4973330, 5236164, 5092534, 4857861, 5116037, 5076550, 5207464, 5076538, 5160043, 5236154, 5007413, 5007412, 5075945, 4881857, 5092537, 5116030, 4769361, 5236131, 5024077, 5236150, 4857852, 4903745, 4727428, 4769381, 5024095, 4881841, 5255128, 6112599, 5874240, 6321604, 6488254, 6397974, 6162026, 6112585, 6488268, 6397927, 6512432, 6512440, 6044143, 6507046, 6512442, 6397915, 6215520, 6215420, 6044138, 6397904, 6488249, 6507099, 6512451, 6112579, 6084845, 6287305, 6215500, 6080554, 6215489, 6316418, 6112601, 6112598, 6397902, 5936795, 6397983, 6094865, 6112571, 5936794, 6287290, 6215493, 6215519, 6349127, 6507077, 6507039, 6512487, 6316416, 6287287, 6215528, 6215416, 6488238, 6215429, 6507064, 6094860, 6321601, 6093639, 6215506, 6112592, 6488255, 6397908, 6093645, 6112576, 6349125, 6215514, 6507091, 5874234, 6316414, 6084828, 6080569, 6162021, 6162028, 6215498, 6397920, 6321609, 6215415, 6112600, 6215409, 6084861, 6397986, 6084860, 6488248, 6507062, 6321594, 6507033, 5910686, 6044147, 6397921, 6188155, 6094879, 6397991, 6507040, 6512458, 5910948, 5910951, 6507048, 6112608, 6188156, 6321608, 6321620, 6162037, 6316432, 6349131, 6397990, 6162035, 5937540, 6215511, 6316420, 6316405, 6080560, 6162017, 6079293, 6512484, 6488250, 5874232, 5899090, 6321617, 6084835, 6094856, 6507094, 6215414, 6084858, 5910668, 6112558, 6094870, 6215432, 6215441, 6321599, 6321602, 6215502, 6316422, 6397928, 6507074, 6094869, 6397989, 6488252, 6397919, 6397922, 6316410, 6397975, 6112602, 6044155, 6094874, 6287279, 6507060, 5910935, 6397979, 6188135, 6188151, 6215438, 6349134, 6215403, 6215504, 6215524, 6215427, 6215436, 6215437, 6215447, 6321598, 6112614, 6094858, 6215430, 6287297, 6287303, 6287307, 6512478, 6321618, 6512436, 6512438, 6512452, 6512473, 6507101, 6507043, 6287278, 6488246, 6215515, 6215517, 6507047, 6507067, 6507068, 6188130, 6349130, 6188131, 6287293, 6507085, 6162041, 6112569, 5899082, 6188153, 6507061, 6488239, 6507071, 6188139, 6079279, 6215523, 6488247, 6397906, 5910952, 6349148, 6215446, 6316408, 6162027, 6094862, 5874354, 6349143, 6507086, 6468788, 6468795, 6468792, 6584836, 6584868, 6584800, 6468803, 6468812, 6584832, 6584874, 6584853, 6468817, 6468798, 6584835, 6600303, 6584846, 6584867, 6584809, 6584837, 6584869, 6584818, 6584812, 6604334, 6584876, 6468790, 6468793, 6468794, 6468796, 6468799, 6468811, 6468800, 6468802, 5993740, 5993735, 5993723, 5993731, 5007431, 4788480, 4973312, 4857853, 5024078, 4676452, 5024093, 5076527, 4810078, 5160034, 5160042, 5092542, 4881848, 4727421, 5236165, 5075934, 5255153, 5076544, 5092544, 5160021, 4810092, 5024088, 4973319, 5207470, 5076542, 5007430, 5236136, 5092535, 6044144, 6512443, 6112567, 5874146, 6079288, 5937514, 6080567, 6397923, 6215423, 5874214, 6397914, 6507042, 6507056, 6507097, 6162016, 6397918, 6397911, 6507035, 6093642, 6507096, 6215503, 5874216, 6044160, 6079280, 6512483, 6488267, 6321613, 6112613, 6084829, 6512449, 6397981, 6215516, 6512453, 6512485, 6112594, 5910947, 6215412, 6215410, 6507050, 5874219, 6488237, 6397988, 5874215, 6397916, 6321596, 6488261, 6079278, 6316423, 6321611, 6215408, 5899076, 6215421, 6512486, 6215501, 6044150, 6079284, 6094864, 6397976, 6321605, 6321622, 6512479, 6507089, 6080561, 5874213, 6321606, 6215435, 6084832, 6507038, 6044140, 6215494, 6349128, 6162032, 5993632, 6112574, 6112573, 6079283, 6512489, 6044163, 6287295, 5874233, 6507087, 5874226, 6215497, 6397971, 6112580, 6507037, 6316411, 5874139, 5993639, 6080559, 5910937, 6287285, 6349135, 6188134, 6215529, 6349137, 6093630, 6215434, 6507036, 5993634, 6188148, 6507070, 5993613, 6507088, 6507093, 5874148, 6112577, 5937529, 6188140, 6287281, 6512462, 6349124, 6188143, 6349126, 6488240, 6507090, 5910936, 6112559, 6468804, 6600292, 6604332, 6623215, 6906544, 6584795, 6844064, 6822423, 6822444, 6584848, 6623222, 6822415, 6584793, 6936522, 6797351, 6797338, 6906536, 6604331, 6622416, 6797344, 6822425, 6844043, 6584863, 6797340, 6622421, 6797339, 6822413, 6822427, 6906554, 6844013, 6584865, 6844063, 6906537, 6822435, 6604336, 6906564, 6936525, 6584801, 6970131, 6822426, 6600291, 6906553, 6844034, 6844055, 6822447, 6622419, 6584815, 6604333, 6468813, 6972587, 6584817, 6584813, 6972638, 6584833, 6844024, 6965484, 6972601, 6970175, 6970160, 6622420, 6822440, 6468808, 6936481, 6906566, 6622428, 6584810, 6622405, 6906562, 6972604, 6970169, 6584829, 6844047, 6822432, 6970172, 6906543, 6584825, 6584850, 6468801, 6844059, 6622417, 6584831, 6844008, 6965482, 6970157, 6972636, 6468807, 6924538, 6972641, 6622425, 6600300, 6970164, 6972633, 6844033, 6584870, 6604329, 6600307, 6584852, 6822418, 6584803, 6906573, 6906552, 6584843, 6924540, 6622402, 6623218, 6468815, 6844015, 6623202, 6623207, 6623221, 6600304, 6622408, 6622423, 6924534, 6924536, 6924537, 6924546, 6924547, 6970166, 6970177, 6970185, 6822430, 6906551, 6844036, 6906565, 6936478, 6936499, 6965474, 6936477, 6936501, 6822419, 6970132, 6844035, 6584856, 6965466, 6584879, 6584806, 6822450, 6844048, 6797336, 6797335, 6584834, 6584826, 6623192, 6936528, 6970136, 6584860, 6822451, 6936519, 6600309, 6468805, 6600317, 6600316, 6822436, 6600310, 6822452, 5993739, 4973329, 5236134, 4881833, 4857858, 5207459, 5160020, 5024086, 5160024, 5075940, 4903761, 5024083, 5024074, 5160028, 6316424, 6316415, 6397924, 6080564, 6507072, 6093624, 5993630, 6287284, 6397926, 6215428, 6349147, 6397900, 6512445, 6397910, 6112604, 5936793, 6397982, 6112612, 6084833, 6112560, 6215413, 6084827, 5874221, 6512437, 6349138, 6162031, 6397968, 6397966, 6162014, 5937536, 6512454, 6512456, 6488265, 6112563, 6512450, 6397909, 5874150, 6512448, 6512463, 5937513, 6112582, 6044148, 6397987, 6215404, 6093635, 5874223, 6044152, 6094871, 6349144, 5910953, 6488262, 5993618, 6084843, 6215507, 5993617, 5874145, 6507053, 5937528, 6215426, 5874352, 6507045, 6162034, 5993629, 5993623, 6094880, 6215402, 5993612, 5874231, 5899092, 6488256, 6397960, 6507044, 6162019, 6488263, 6349129, 5874365, 6397929, 6084854, 5910949, 5937524, 6287294, 6112565, 6906540, 6972592, 6844051, 6604330, 6970150, 6584830, 6965495, 6906547, 6972565, 6972626, 6970142, 6970170, 6936521, 6970155, 6468789, 6972578, 6965468, 6972625, 6936486, 6936512, 6972594, 6972607, 6604335, 6936520, 6797350, 6972616, 6972630, 6972598, 6970135, 6604327, 6936514, 6936515, 6965487, 6972622, 6584864, 6965476, 6936493, 6972614, 6936517, 6970147, 6972570, 6584816, 6965496, 6972624, 6965481, 6970173, 6972562, 6584845, 6972629, 6972613, 6972621, 6844020, 6623205, 6623212, 6970159, 6584840, 6844040, 6584819, 6972581, 6970151, 6468806, 6600308, 6936504, 6965478, 6970181, 6623223, 6584866, 6970152, 6468797, 6936523, 6844038, 6822449, 6970129, 6972573, 6965493, 6584851, 6965473, 6970161, 6600314, 6844041, 6622406, 6972571, 6972583, 6972585, 6584861, 6584802, 6972567, 6936495, 6844053, 6797349, 6844031, 6844014, 6844060, 6936492, 6936490, 6844067, 6972635, 6965472, 6936498, 6965469, 6906546, 6584811, 6965471, 6600295, 6970162, 6970158, 6970178, 6972620, 6972568, 6797334, 6844062, 6972631, 6600311, 6844037, 6972589, 6600290, 6600312, 6965488, 6972615, 6972564, 6965465, 5993738, 6397925, 6507069, 6488251, 6316425, 6507052, 6397970, 6507055, 5874227, 5993615, 6188133, 6162033, 5899067, 6084836, 6162025, 6316419, 6512472, 6349146, 6215487, 5874360, 6397980, 6507075, 6084852, 6287296, 6215405, 6287306, 6094881, 6512482, 6397972, 5936799, 5937517, 6188144, 6215505, 6397912, 6316431, 5874236, 5874228, 5936807, 5874376, 6397967, 6316421, 6488241, 6215417, 6287304, 6972610, 6972606, 6906555, 6936533, 6970183, 6906558, 6972588, 6822429, 6972593, 6965480, 6936536, 6972586, 6822416, 6970186, 6936491, 6584859, 6844030, 6822439, 6965492, 6970156, 6972572, 6972603, 6622415, 6468791, 6623198, 6797353, 6844029, 6972619, 6584844, 6822445, 6972623, 6844016, 6965483, 6584847, 6936508, 6797347, 6584794, 6972600, 6822446, 6600301, 6844049, 6844026, 6844066, 6924545, 6936509, 6584841, 6822422, 5993737, 6094863, 6044161, 6512475, 6972579, 6972627

Distribution

Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand

Quantity

7213 units