FDA Enforcement Class II Ongoing

Beckman Coulter Access Free T3, Triiodothyronine, Free, Catalog #A13422, each box contains 2 reagent packs.

Recall: Z-1419-2023 · Reported April 26, 2023

Enforcement

Recall Number
Z-1419-2023
Event ID
91916
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beckman Coulter, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 26, 2023
Initiation Date
March 15, 2023
Classification Date
April 20, 2023
Address
1000 Lake Hazeltine Dr, N/A, Chaska, MN, 55318-1037, United States

Description

Beckman Coulter Access Free T3, Triiodothyronine, Free, Catalog #A13422, each box contains 2 reagent packs.

Reason

The reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers nor Free T3 results generated on other platforms. This introduces a risk of the positive bias not being detected by QC with the potential that clinical correlation with other thyroid function tests will be inconsistent requiring additional testing; however the clinical assessment and treatment decisions are not made in isolation on a single test result especially, in the setting of thyroid function testing. False positive test results could lead to a euthyroid patient being interpreted clinically as having elevated thyroid function (hyperthyroid) or a hypothyroid patient being within the normal range. Such misdiagnoses may lead to inappropriate treatment of a euthyroid patient with anti-thyroid medications or hypothyroid patient not getting sufficient treatment since their free T3 test result demonstrates them to be in the euthyroid range.

Code Info

Lot #233968, exp. 2023-06-30, UDI (01)15099590201661(17)230630(11)220630(10)233968.

Distribution

Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Bahrain, Brazil, Canada, China, Costa Rica, El Salvador, Georgia, Hong Kong, Malaysia, Mexico, Myanmar, New Zealand, Peru, Philippines, Qatar, Singapore, Taiwan, Thailand, Uruguay, and Viet Nam.

Quantity

31,076 units