FDA Enforcement Class II Terminated

RayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0; CLASSIFICATION NAME: System, Planning, Radiation Therapy Treatment. RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Recall: Z-1419-2013 · Reported June 5, 2013

Enforcement

Recall Number
Z-1419-2013
Event ID
65136
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
RAYSEARCH LABORATORIES AB
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
June 5, 2013
Initiation Date
February 15, 2013
Classification Date
May 28, 2013
Termination Date
February 17, 2017
Address
Sveavaegen 9, N/A, Stockholm, N/A, N/A, Sweden

Description

RayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0; CLASSIFICATION NAME: System, Planning, Radiation Therapy Treatment. RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Reason

Under some circumstances, imported CT, MR and PET images are offset from their true positions by one pixel in one or two directions. Users must be aware of this information to avoid incorrect dose calculations during treatment planning.

Code Info

Software version and build numbers 2.0.0.15, 2.5.1.89, and 3.0.0.251 (to determine if the version used is affected, bring up the About box in the RayStation application and check the build number reported0.

Distribution

Worldwide Distribution - USA including Connecticut, Florida, Massachusetts, Missouri, New York, Ohio, and foreign distribution to(Canada, Italy, Australia, Germany, Netherlands, and Switzerland.

Quantity

Domestic: 70 units; Foreign: 27 units