Medline Convenience Kits: 1) KIT UROLOGY SLING, Model Number: DYKMBNDL106; 2) KIT UROLOGY SLING, Model Number: DYKMBNDL106A; 3) KIT UROLOGY PENILE REPAIR, Model Number: DYKMBNDL188; 4) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189; 5) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189A
Research enforcement Z-1416-2026 in seconds
The Research Assistant summarizes enforcement actions and related recalls — and cites every record.
Free to try · every answer cites its records
Enforcement
- Recall Number
- Z-1416-2026
- Event ID
- 98329
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 25, 2026
- Initiation Date
- January 7, 2026
- Classification Date
- February 17, 2026
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Medline Convenience Kits: 1) KIT UROLOGY SLING, Model Number: DYKMBNDL106; 2) KIT UROLOGY SLING, Model Number: DYKMBNDL106A; 3) KIT UROLOGY PENILE REPAIR, Model Number: DYKMBNDL188; 4) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189; 5) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189A
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
1) DYKMBNDL106, UDI-DI: 10193489479294(each), 40193489479295(case), Lot Number: 21BBG085; 2) DYKMBNDL106, UDI-DI: 10193489479294(each), 40193489479295(case), Lot Number: 21ABJ542; 3) DYKMBNDL106A, UDI-DI: 10193489880816(each), 40193489880817(case), Lot Number: 22CLA049; 4) DYKMBNDL106A, UDI-DI: 10193489880816(each), 40193489880817(case), Lot Number: 22BLB059; 5) DYKMBNDL106A, UDI-DI: 10193489880816(each), 40193489880817(case), Lot Number: 22ALA713; 6) DYKMBNDL106A, UDI-DI: 10193489880816(each), 40193489880817(case), Lot Number: 21KLA631; 7) DYKMBNDL188, UDI-DI: 10193489478549(each), 40193489478540(case), Lot Number: 21CBS043; 8) DYKMBNDL188, UDI-DI: 10193489478549(each), 40193489478540(case), Lot Number: 21ABS540; 9) DYKMBNDL188, UDI-DI: 10193489478549(each), 40193489478540(case), Lot Number: 21ABS539; 10) DYKMBNDL189, UDI-DI: 10193489478457(each), 40193489478458(case), Lot Number: 21BBG092; 11) DYKMBNDL189, UDI-DI: 10193489478457(each), 40193489478458(case), Lot Number: 21ABI996; 12) DYKMBNDL189A, UDI-DI: 10193489880823(each), 40193489880824(case), Lot Number: 21JLA521
Worldwide distribution - US Nationwide.
59 units