FDA Enforcement Class II Terminated

The EOSedge system may acquire two simultaneous orthogonal planar images for a standing or seated patient at low dose. The scan length is defined by the user and may cover the whole body or a specific area (spine, lower limbs, etc.). EOSedge consists of a gantry and an acquisition workstation. ¿ The gantry contains two orthogonal acquisition units, each of which comprises an X-ray source (high frequency generator + X-ray tube + collimator) and an X-ray detection system (detector + electronics). ¿ The acquisition workstation and its software control the generators and detectors. It is also used to enter patient data, via RIS/HIS or manually, to define acquisition parameters, process images and record images in DICOM 3.0 format. The intended use/indications for use, technological characteristics, and principles of operation of the EOSedge system are described below.

Recall: Z-1415-2021 · Reported April 21, 2021

Enforcement

Recall Number
Z-1415-2021
Event ID
87505
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
EOS Imaging
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 21, 2021
Initiation Date
December 22, 2020
Classification Date
April 12, 2021
Termination Date
May 5, 2023
Address
4 Ieme Etage, 10 Rue Mercoeur, Paris, N/A, N/A, France

Description

The EOSedge system may acquire two simultaneous orthogonal planar images for a standing or seated patient at low dose. The scan length is defined by the user and may cover the whole body or a specific area (spine, lower limbs, etc.). EOSedge consists of a gantry and an acquisition workstation. ¿ The gantry contains two orthogonal acquisition units, each of which comprises an X-ray source (high frequency generator + X-ray tube + collimator) and an X-ray detection system (detector + electronics). ¿ The acquisition workstation and its software control the generators and detectors. It is also used to enter patient data, via RIS/HIS or manually, to define acquisition parameters, process images and record images in DICOM 3.0 format. The intended use/indications for use, technological characteristics, and principles of operation of the EOSedge system are described below.

Reason

Inadequate images resizing and 2D measurement errors may occur when biplanar acquisition has been performed with patient orientation different from AP (Antero-Posterior).

Code Info

Product Reference:03663999000108; Serial Numbers: 8.1912.1008, 8.2001.1010, 8.2001.1011, 8.2002.1012, 8.1907.1005, 8.1909.1006, 8.2005.1014, 8.1905.1004, 8.1911.1007, 8.1912.1009, and 8.2006.1016

Distribution

Distributed nationwide to MN, OH and internationally to Australia, Canada, Germany, Spain, and France

Quantity

11 devices