Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094
Enforcement
- Recall Number
- Z-1411-2021
- Event ID
- 87542
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Covidien
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 21, 2021
- Initiation Date
- March 5, 2021
- Classification Date
- April 12, 2021
- Address
- 201 Sabanetas Industrial Park, Building 911-67, Ponce, PR, 00716, United States
Description
Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094
Investigations related to customer complaints received for particulate matter inside of the Surgiwand" II device tubing components identified the root cause as a damaged Y-Connector component which connects the tubing to the device. This damage was determined to generate the particulate matter observed and is created during the assembly process for connecting the tubing component to the Y-Connector.
UDI-Device Identifier (GTIN/UPN): 10884523000887, 20884523000884 UDI-Device Identifier (GTIN/UPN): 10884523000900, 20884523000907 UDI-Device Identifier (GTIN/UPN): 10884523000917, 20884523000914
Domestic: AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, UT, WA, WI, WV. Foreign: Canada, Japan, Taiwan, Australia, New Zealand, Indonesia, Thailand, Singapore, Malaysia, Philippines, Bahamas, Brazil, Chile, El Salvador, Guatemala, Jamaica, Mexico, Panama, Peru, Puerto Rico, Trinidad And Tobago, Austria, Belgium, Bosnia And Herzegovina, Canary Islands, Cyprus, Denmark, Equatorial Guinea, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Kosovo, Kuwait, Lebanon, Mauritius, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, South Africa, Spain, Swaziland, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.
59,828 devices