FDA Enforcement Class II Terminated

Canon Full Auto Ref-Keratometer RK-F2, Catalog Number 6937B002AA. Product Usage: This device is intended to be used to measure the refractive power and the radius of corneal curvature of the human eye.

Recall: Z-1404-2013 · Reported June 5, 2013

Enforcement

Recall Number
Z-1404-2013
Event ID
65082
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Canon Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 5, 2013
Initiation Date
December 5, 2012
Classification Date
May 25, 2013
Termination Date
September 20, 2013
Address
30-2, Shimomaruko 3-chome,, Ohta-ku, Tokyo, N/A, N/A, Japan

Description

Canon Full Auto Ref-Keratometer RK-F2, Catalog Number 6937B002AA. Product Usage: This device is intended to be used to measure the refractive power and the radius of corneal curvature of the human eye.

Reason

Canon Inc. had received 17 claims from our foreign sales companies between October to November 2012 that measurement value of RK-F2s refractive power (SPH value) sometimes indicated abnormal value which occurs approximately 5% in probability.

Code Info

Catalog Number 6937B002AA. All serial numbers from the first unit onward.

Distribution

Worldwide Distribution - USA Nationwide and countries of: Australia, Japan, Korea, Singapore and United Kingdom.

Quantity

384 units