FDA Enforcement
Class II
Terminated
Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.
Recall: Z-1402-2013
·
Reported June 5, 2013
Enforcement
- Recall Number
- Z-1402-2013
- Event ID
- 65060
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Elekta, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 5, 2013
- Initiation Date
- April 30, 2013
- Classification Date
- May 24, 2013
- Termination Date
- July 26, 2017
- Address
- 400 Perimeter Center Ter Ne, Ste 50, Atlanta, GA, 30346-1227, United States
Description
Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.
Reason
Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0.
Code Info
325134, 325254, 325082, 325159, 325268, 325237, 325331, 325267, 325236, 325314, 325127, 325261, 325286, 325077, 325269, 325129, 325296, 325273, 325139, 325201, 325137, 325184, 325161, 325076, 325086, 325112, 325081, 325083, 325099, 325110, 325271
Distribution
USA Nationwide including the states of CA, FL, IL, MD, MA, NJ, NY, NC, ND, OH, OR, PA, TN, TX, VA, WA, DC and Puerto Rico
Quantity
31