FDA Enforcement
Class II
Ongoing
VERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
Recall: Z-1398-2025
·
Reported March 26, 2025
Enforcement
- Recall Number
- Z-1398-2025
- Event ID
- 96344
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Merge Healthcare, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- March 26, 2025
- Initiation Date
- February 19, 2025
- Classification Date
- March 20, 2025
- Address
- 900 Walnut Ridge Dr, N/A, Hartland, WI, 53029-8347, United States
Description
VERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
Reason
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
Code Info
The following versions were introduced prior to the implementation of UDI. Version Numbers: 9.0.6, 9.0.8, Part Numbers: 88-00061-00, 88-00062-00, 88-00063-00, 88-00054-00, 88-00064-00,
Distribution
US
Quantity
2