FDA Enforcement Class II Ongoing

Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117C; 3) KIT PLASTIC INSERT TISSUE EXPA, Model Number: DYKMBNDL158; 4) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170; 5) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170A; 6) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170B; 7) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182; 8) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182A; 9) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182B; 10) KIT PLASTIC ROTATIONAL FLAP, Model Number: DYKMBNDL183; 11) KIT PLASTIC ROTATIONAL FLAP, Model Number: DYKMBNDL183A; 12) KIT PLASTIC ROTATIONAL FLAP, Model Number: DYKMBNDL183B; 13) KIT STC FACIAL FRACTURE, Model Number: DYKMBNDL34A; 14) KIT SURG ONC MASTECTOMY A, Model Number: DYKMBNDL39; 15) KIT SURG ONC MASTECTOMY A, Model Number: DYKMBNDL39A; 16) KIT SURG ONC MASTECTOMY, Model Number: DYKMBNDL39C; 17) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL46; 18) KIT STC PLASTIC FREE FLAP, Model Number: DYKMBNDL54; 19) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL69; 20) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL69A; 21) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL69B; 22) KIT PLASTICS BREAST AUGMENTATI, Model Number: DYKMBNDL78; 23) KIT PLASTICS BREAST AUGMENTATI, Model Number: DYKMBNDL78A; 24) KIT PLASTICS BREAST AUGMENTATI, Model Number: DYKMBNDL78B; 25) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84; 26) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84C; 27) MJR PLASTICS PK MV SURG CT-LF, Model Number: DYNJ0785310G; 28) ASC ENT PLASTIC PACK-LF, Model Number: DYNJ0854650J; 29) BREAST PACK ASC-LF, Model Number: DYNJ36469I; 30) ENT TYMPANOPLASTY PACK-LF, Model Number: DYNJ38988J; 31) DR WILDE BRSTMASS/PARTMAST-XM, Model Number: DYNJ39831F; 32) BREAST RECONSTRUCTION-HOULE, Model Number: DYNJ44852K; 33) BREAST RECONSTRUCTION-HOULE, Model Number: DYNJ44852M; 34) BREAST PACK, Model Number: DYNJ50957G; 35) PLASTICS BREAST PACK, Model Number: DYNJ55358G; 36) BHN IMPLANT PACK, Model Number: DYNJ56737A; 37) BREAST PACK, Model Number: DYNJ60210A; 38) PLASTICS PACK 2, Model Number: DYNJ64927F; 39) SPECIALTY FACIAL PLASTICS PACK, Model Number: DYNJ68844A; 40) BREAST IMPLANT PACK, Model Number: DYNJ69351; 41) BREAST/TUMMY PACK, Model Number: DYNJ83788; 42) HMSL PLASTICS, Model Number: DYNJ83884B; 43) CHEST/BREAST-LF, Model Number: DYNJ83950A; 44) CHEST/BREAST-LF, Model Number: DYNJ83950B; 45) FREE FLAP A DIRTY, Model Number: DYNJ86165; 46) MAJOR SURGERY PACK, Model Number: DYNJ86816; 47) BREAST REDUCTION PACK, Model Number: DYNJ87444; 48) OMF FREE FLAP, Model Number: DYNJ902005I; 49) BREAST CDS, Model Number: DYNJ902558G; 50) TRUNCAL CONTOURING, Model Number: DYNJ903555D; 51) TRUNCAL CONTOURING, Model Number: DYNJ903555F; 52) NEW TISSUE RETRIEVAL MH, Model Number: DYNJ903849C; 53) LARGE PROCEDURE-LF, Model Number: DYNJ907847A; 54) (34) STC FACIAL FRACTURE, Model Number: DYNJ907874A; 55) FACELIFT, Model Number: DYNJ910462; 56) MAJOR PLASTIC BREAST PACK-LF, Model Number: DYNJT3283; 57) MAJOR BREAST PACK-LF, Model Number: DYNJT3417; 58) PLASTICS PACK-LF, Model Number: PHS806973G

Recall: Z-1397-2026 · Reported February 25, 2026

Enforcement

Recall Number
Z-1397-2026
Event ID
98329
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medline Industries, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 25, 2026
Initiation Date
January 7, 2026
Classification Date
February 17, 2026
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117C; 3) KIT PLASTIC INSERT TISSUE EXPA, Model Number: DYKMBNDL158; 4) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170; 5) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170A; 6) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170B; 7) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182; 8) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182A; 9) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182B; 10) KIT PLASTIC ROTATIONAL FLAP, Model Number: DYKMBNDL183; 11) KIT PLASTIC ROTATIONAL FLAP, Model Number: DYKMBNDL183A; 12) KIT PLASTIC ROTATIONAL FLAP, Model Number: DYKMBNDL183B; 13) KIT STC FACIAL FRACTURE, Model Number: DYKMBNDL34A; 14) KIT SURG ONC MASTECTOMY A, Model Number: DYKMBNDL39; 15) KIT SURG ONC MASTECTOMY A, Model Number: DYKMBNDL39A; 16) KIT SURG ONC MASTECTOMY, Model Number: DYKMBNDL39C; 17) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL46; 18) KIT STC PLASTIC FREE FLAP, Model Number: DYKMBNDL54; 19) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL69; 20) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL69A; 21) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL69B; 22) KIT PLASTICS BREAST AUGMENTATI, Model Number: DYKMBNDL78; 23) KIT PLASTICS BREAST AUGMENTATI, Model Number: DYKMBNDL78A; 24) KIT PLASTICS BREAST AUGMENTATI, Model Number: DYKMBNDL78B; 25) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84; 26) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84C; 27) MJR PLASTICS PK MV SURG CT-LF, Model Number: DYNJ0785310G; 28) ASC ENT PLASTIC PACK-LF, Model Number: DYNJ0854650J; 29) BREAST PACK ASC-LF, Model Number: DYNJ36469I; 30) ENT TYMPANOPLASTY PACK-LF, Model Number: DYNJ38988J; 31) DR WILDE BRSTMASS/PARTMAST-XM, Model Number: DYNJ39831F; 32) BREAST RECONSTRUCTION-HOULE, Model Number: DYNJ44852K; 33) BREAST RECONSTRUCTION-HOULE, Model Number: DYNJ44852M; 34) BREAST PACK, Model Number: DYNJ50957G; 35) PLASTICS BREAST PACK, Model Number: DYNJ55358G; 36) BHN IMPLANT PACK, Model Number: DYNJ56737A; 37) BREAST PACK, Model Number: DYNJ60210A; 38) PLASTICS PACK 2, Model Number: DYNJ64927F; 39) SPECIALTY FACIAL PLASTICS PACK, Model Number: DYNJ68844A; 40) BREAST IMPLANT PACK, Model Number: DYNJ69351; 41) BREAST/TUMMY PACK, Model Number: DYNJ83788; 42) HMSL PLASTICS, Model Number: DYNJ83884B; 43) CHEST/BREAST-LF, Model Number: DYNJ83950A; 44) CHEST/BREAST-LF, Model Number: DYNJ83950B; 45) FREE FLAP A DIRTY, Model Number: DYNJ86165; 46) MAJOR SURGERY PACK, Model Number: DYNJ86816; 47) BREAST REDUCTION PACK, Model Number: DYNJ87444; 48) OMF FREE FLAP, Model Number: DYNJ902005I; 49) BREAST CDS, Model Number: DYNJ902558G; 50) TRUNCAL CONTOURING, Model Number: DYNJ903555D; 51) TRUNCAL CONTOURING, Model Number: DYNJ903555F; 52) NEW TISSUE RETRIEVAL MH, Model Number: DYNJ903849C; 53) LARGE PROCEDURE-LF, Model Number: DYNJ907847A; 54) (34) STC FACIAL FRACTURE, Model Number: DYNJ907874A; 55) FACELIFT, Model Number: DYNJ910462; 56) MAJOR PLASTIC BREAST PACK-LF, Model Number: DYNJT3283; 57) MAJOR BREAST PACK-LF, Model Number: DYNJT3417; 58) PLASTICS PACK-LF, Model Number: PHS806973G

Reason

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Code Info

1) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 25BDB702; 2) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 25BDA228; 3) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 25ADB705; 4) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 24JDA650; 5) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 24IDA403; 6) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 23LDC086; 7) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 23IDB182; 8) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 23HDB786; 9) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21JDB530; 10) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21HDB434; 11) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21GDB740; 12) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21FDA888; 13) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21DDA661; 14) DYKMBNDL117C, UDI-DI: 10195327371104(each), 40195327371105(case), Lot Number: 23DLA834; 15) DYKMBNDL158, UDI-DI: 10193489478266(each), 40193489478267(case), Lot Number: 21CBE548; 16) DYKMBNDL158, UDI-DI: 10193489478266(each), 40193489478267(case), Lot Number: 21ABI978; 17) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 22IBC515; 18) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 22HBN456; 19) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 22EBA143; 20) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 21IBV640; 21) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 21DBN049; 22) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 21BBH637; 23) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 21ABJ202; 24) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 20LBY951; 25) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 23IBB851; 26) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 23HBG507; 27) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 23HBB877; 28) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 23GBR888; 29) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 22JBM366; 30) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25LBM160; 31) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25KBR870; 32) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25GBA895; 33) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25EBM448; 34) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25DBS731; 35) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25DBK034; 36) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25BBN216; 37) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25BBA340; 38) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 24IBB931; 39) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 24FBL546; 40) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 24FBD539; 41) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 23LBI220; 42) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 22FBC879; 43) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 22EBO665; 44) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 22BBN306; 45) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 21DBC569; 46) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 21ABX025; 47) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 21ABI990; 48) DYKMBNDL182A, UDI-DI: 10195327633806(each), 40195327633807(case), Lot Number: 24BBN507; 49) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 25FBO935; 50) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 25ABL533; 51) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 24KBS891; 52) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 24IBQ469; 53) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 24IBL678; 54) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 24DBR913; 55) DYKMBNDL183, UDI-DI: 10193489478495(each), 40193489478496(case), Lot Number: 21JBJ252; 56) DYKMBNDL183, UDI-DI: 10193489478495(each), 40193489478496(case), Lot Number: 21CBG305; 57) DYKMBNDL183, UDI-DI: 10193489478495(each), 40193489478496(case), Lot Number: 21ABI992; 58) DYKMBNDL183A, UDI-DI: 10195327246594(each), 40195327246595(case), Lot Number: 23BBI487; 59) DYKMBNDL183B, UDI-DI: 10195327633813(each), 40195327633814(case), Lot Number: 24GBC829; 60) DYKMBNDL183B, UDI-DI: 10195327633813(each), 40195327633814(case), Lot Number: 24DBT970; 61) DYKMBNDL183B, UDI-DI: 10195327633813(each), 40195327633814(case), Lot Number: 24DBJ591; 62) DYKMBNDL34A, UDI-DI: 10193489872033(each), 40193489872034(case), Lot Number: 21DBQ512; 63) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 22IBC434; 64) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 22GBM861; 65) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 22BBU074; 66) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 21LBJ742; 67) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 21GBR352; 68) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 21CBD335; 69) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 21ABJ002; 70) DYKMBNDL39A, UDI-DI: 10195327241308(each), 40195327241309(case), Lot Number: 23ABH677; 71) DYKMBNDL39C, UDI-DI: 10198459112461(each), 40198459112462(case), Lot Number: 25GBW751; 72) DYKMBNDL39C, UDI-DI: 10198459112461(each), 40198459112462(case), Lot Number: 24JBO020; 73) DYKMBNDL46, UDI-DI: 10193489478761(each), 40193489478762(case), Lot Number: 21GBR353; 74) DYKMBNDL46, UDI-DI: 10193489478761(each), 40193489478762(case), Lot Number: 21BBM839; 75) DYKMBNDL46, UDI-DI: 10193489478761(each), 40193489478762(case), Lot Number: 21ABL308; 76) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21CBG791; 77) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21BBH629; 78) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21ABO607; 79) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21ABM089; 80) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21ABJ589; 81) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 20LBY967; 82) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 22JBC409; 83) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 22IBR235; 84) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 22HBQ100; 85) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 22HBN457; 86) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21JBS176; 87) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21IBV702; 88) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21HBO272; 89) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21HBM983; 90) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21HBK340; 91) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21GBR324; 92) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21DBC571; 93) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21ABL345; 94) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21ABJ013; 95) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 23LBA275; 96) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 23JBJ695; 97) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 23GBL960; 98) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 23FBP246; 99) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22KBO876; 100) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22JBQ852; 101) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22JBJ150; 102) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22JBC626; 103) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22JBC618; 104) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 25KBP763; 105) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 25GBL984; 106) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 25FBH462; 107) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 24KBT018; 108) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 24HBM099; 109) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 24GBF481; 110) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 24EBR715; 111) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22EBS816; 112) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22CBX537; 113) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22BBL848; 114) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22ABL251; 115) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22ABG018; 116) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21LBY278; 117) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21LBO572; 118) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21KBY757; 119) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21JBJ751; 120) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21IBV643; 121) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21IBB097; 122) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21HBK310; 123) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21FBR398; 124) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21FBF907; 125) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21DBC549; 126) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21BBK531; 127) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21ABX017; 128) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21ABJ216; 129) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 20XBB257; 130) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 23LBE774; 131) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 23JBN862; 132) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 23GBP831; 133) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 23GBO451; 134) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 22KBO921; 135) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 22JBQ856; 136) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 22JBJ143; 137) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 22JBJ019; 138) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25FBP527; 139) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25EBM007; 140) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25EBE325; 141) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25CBG915; 142) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25BBU089; 143) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25BBO685; 144) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25BBE752; 145) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24LBL035; 146) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24HBD157; 147) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24EBU597; 148) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24EBE953; 149) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24CBR114; 150) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24CBF892; 151) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22OBH455; 152) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22IBC435; 153) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22HBO907; 154) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22FBT013; 155) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22FBF307; 156) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21LBH705; 157) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21LBD628; 158) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21JBH111; 159) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21IBT011; 160) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21GBR338; 161) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21GBF538; 162) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21EBH614; 163) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21CBM835; 164) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21ABU822; 165) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21ABM088; 166) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25LBQ326; 167) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25LBP864; 168) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25HBG553; 169) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25GBF486; 170) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25FBB154; 171) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25DBO560; 172) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 24LBS883; 173) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 24LBI135; 174) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 24JBE336; 175) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 24IBE175; 176) DYNJ0785310G, UDI-DI: 10193489268966(each), 40193489268967(case), Lot Number: 23KMH075; 177) DYNJ0854650J, UDI-DI: 10193489832426(each), 40193489832427(case), Lot Number: 24IMF274; 178) DYNJ36469I, UDI-DI: 10195327513931(each), 40195327513932(case), Lot Number: 24BBU341; 179) DYNJ38988J, UDI-DI: 10193489315233(each), 40193489315234(case), Lot Number: 21CDB475; 180) DYNJ38988J, UDI-DI: 10193489315233(each), 40193489315234(case), Lot Number: 21BDB779; 181) DYNJ38988J, UDI-DI: 10193489315233(each), 40193489315234(case), Lot Number: 21ADA533; 182) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24KMD041; 183) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24IME867; 184) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24GMA471; 185) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24DMJ046; 186) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24CMD088; 187) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24AMI035; 188) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23LMB124; 189) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23KMA468; 190) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23IMG016; 191) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23HMA900; 192) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23AME665; 193) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 22IMF713; 194) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 22HMH034; 195) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 22GMF684; 196) DYNJ44852K, UDI-DI: 10195327127589(each), 40195327127580(case), Lot Number: 22CMI118; 197) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 25EME929; 198) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 25CME690; 199) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 24KME600; 200) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 24GMI186; 201) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 24CMC758; 202) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 24AMB315; 203) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23HME689; 204) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23FMH424; 205) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23DMD096; 206) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23CMF343; 207) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23AMC530; 208) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23AMC374; 209) DYNJ50957G, UDI-DI: 10193489448863(each), 40193489448864(case), Lot Number: 23GBV065; 210) DYNJ55358G, UDI-DI: 10195327173500(each), 40195327173501(case), Lot Number: 23FBN323; 211) DYNJ56737A, UDI-DI: 10193489395594(each), 40193489395595(case), Lot Number: 23CME388; 212) DYNJ60210A, UDI-DI: 10193489985108(each), 40193489985109(case), Lot Number: 21KBP457; 213) DYNJ64927F, UDI-DI: 10195327143091(each), 40195327143092(case), Lot Number: 23FBV335; 214) DYNJ68844A, UDI-DI: 10195327216504(each), 40195327216505(case), Lot Number: 23HMB858; 215) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 24BBL537; 216) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23KBN973; 217) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23JBU455; 218) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23HBY228; 219) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23CBG820; 220) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23BBP453; 221) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23BBD661; 222) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 21LBP376; 223) DYNJ83788, UDI-DI: 10195327305833(each), 40195327305834(case), Lot Number: 23CBO212; 224) DYNJ83884B, UDI-DI: 10198459234620(each), 40198459234621(case), Lot Number: 25GBO120; 225) DYNJ83950A, UDI-DI: 10198459018510(each), 40198459018511(case), Lot Number: 25ABF915; 226) DYNJ83950A, UDI-DI: 10198459018510(each), 40198459018511(case), Lot Number: 24JBV233; 227) DYNJ83950A, UDI-DI: 10198459018510(each), 40198459018511(case), Lot Number: 24FBL440; 228) DYNJ83950A, UDI-DI: 10198459018510(each), 40198459018511(case), Lot Number: 24EBR736; 229) DYNJ83950B, UDI-DI: 10198459257827(each), 40198459257828(case), Lot Number: 25DBT195; 230) DYNJ83950B, UDI-DI: 10198459257827(each), 40198459257828(case), Lot Number: 25CBF253; 231) DYNJ83950B, UDI-DI: 10198459257827(each), 40198459257828(case), Lot Number: 25ABT952; 232) DYNJ86165, UDI-DI: 10195327599324(each), 40195327599325(case), Lot Number: 25ABJ747; 233) DYNJ86816, UDI-DI: 10195327679903(each), 40195327679904(case), Lot Number: 24FMB609; 234) DYNJ87444, UDI-DI: 10198459040122(each), 40198459040123(case), Lot Number: 24GBQ988; 235) DYNJ902005I, UDI-DI: 10195327347710(each), 40195327347711(case), Lot Number: 24DLA844; 236) DYNJ902558G, UDI-DI: 10193489728859(each), 40193489728850(case), Lot Number: 23HMG989; 237) DYNJ903555D, UDI-DI: 10195327084615(each), 40195327084616(case), Lot Number: 22HBF672; 238) DYNJ903555D, UDI-DI: 10195327084615(each), 40195327084616(case), Lot Number: 22GBT713; 239) DYNJ903555F, UDI-DI: 10195327224585(each), 40195327224586(case), Lot Number: 23GBP519; 240) DYNJ903555F, UDI-DI: 10195327224585(each), 40195327224586(case), Lot Number: 23GBP518; 241) DYNJ903555F, UDI-DI: 10195327224585(each), 40195327224586(case), Lot Number: 23GBG876; 242) DYNJ903555F, UDI-DI: 10195327224585(each), 40195327224586(case), Lot Number: 23FBN475; 243) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 23DBS608; 244) DYNJ907847A, UDI-DI: 10193489937589(each), 40193489937580(case), Lot Number: 23IMF989; 245) DYNJ907874A, UDI-DI: 10195327241247(each), 40195327241248(case), Lot Number: 23FLA249; 246) DYNJ910462, UDI-DI: 10195327695637(each), 40195327695638(case), Lot Number: 24JMD608; 247) DYNJ910462, UDI-DI: 10195327695637(each), 40195327695638(case), Lot Number: 24IMF741; 248) DYNJ910462, UDI-DI: 10195327695637(each), 40195327695638(case), Lot Number: 24GME383; 249) DYNJT3283, UDI-DI: 10198459323645(each), 40198459323646(case), Lot Number: 25JBI499; 250) DYNJT3417, UDI-DI: 10198459324888(each), 40198459324889(case), Lot Number: 25EBN203; 251) PHS806973G, UDI-DI: 10193489926996(each), 40193489926997(case), Lot Number: 23EBQ702

Distribution

Worldwide distribution - US Nationwide.

Quantity

4757 units