FDA Enforcement Class I Ongoing

Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, 2.1mm Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Kit for Leksell Stereotactic System. Each inner box of REF 911761 contains a Tyvek bag labeled REF 912465, Disposable Biopsy Needle and Aspirator Tube.

Recall: Z-1397-2024 · Reported April 24, 2024

Enforcement

Recall Number
Z-1397-2024
Event ID
94197
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Elekta Instrument AB
Voluntary / Mandated
Voluntary: Firm initiated
Report Date
April 24, 2024
Initiation Date
March 15, 2024
Classification Date
April 16, 2024
Address
Kungstensgatan 18, Stockholm, N/A, Sweden

Description

Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, 2.1mm Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Kit for Leksell Stereotactic System. Each inner box of REF 911761 contains a Tyvek bag labeled REF 912465, Disposable Biopsy Needle and Aspirator Tube.

Reason

Elekta has become aware that Disposable Biopsy Needles (911933) from one batch (837838839) can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles. The sterility of the biopsy needles has not been affected. This issue has been reported from one site.

Code Info

Lot number 837838839 appears on the labels for REF 911933, REF 911761, and REF 912465. UDI numbers: REF 911933 - (01)0 7340048 30034 6, (17)250401 (10) 837838839 (240) 911933; REF 911761 - (01)0 7340048 30800 7, (17)250401 (10) 837838839 (240) 911761; REF 912465 - (01)0 7340048 30801 4, (17)250401 (10) 837838839 (240) 912465.

Distribution

US and Australia, Austria, Belgium, France, Germany, Greece, Italy, Netherlands, Spain, Sweden, Switzerland

Quantity

601 needles