FDA Enforcement Class I Ongoing

CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.

Recall: Z-1396-2025 · Reported April 2, 2025

Enforcement

Recall Number
Z-1396-2025
Event ID
96340
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Calyxo, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 2, 2025
Initiation Date
February 19, 2025
Classification Date
March 21, 2025
Address
4473 Willow Rd, N/A, Pleasanton, CA, 94588-8571, United States

Description

CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.

Reason

Aspiration system, for endoscopic examination/treatment of urinary tract and kidney interior, to have labeling update adding additional instructions, for patients with high viscosity fluid in the kidney, not to continue providing fluid inflow in the presence of unresolved slow or absent fluid outflow because this can create intrarenal pressure imbalance, and lead to excessive intrarenal pressure.

Code Info

CVC127020-1 UDI-DI: 00860005357710, L00018 Rev C

Distribution

US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV.

Quantity

11,246