CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.
Enforcement
- Recall Number
- Z-1396-2025
- Event ID
- 96340
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Calyxo, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 2, 2025
- Initiation Date
- February 19, 2025
- Classification Date
- March 21, 2025
- Address
- 4473 Willow Rd, N/A, Pleasanton, CA, 94588-8571, United States
Description
CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/treatment of urinary tract and kidney interior.
Aspiration system, for endoscopic examination/treatment of urinary tract and kidney interior, to have labeling update adding additional instructions, for patients with high viscosity fluid in the kidney, not to continue providing fluid inflow in the presence of unresolved slow or absent fluid outflow because this can create intrarenal pressure imbalance, and lead to excessive intrarenal pressure.
CVC127020-1 UDI-DI: 00860005357710, L00018 Rev C
US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV.
11,246