AURIGA 30 BRAZIL ZERO COST SYSTEM UPN: M0068S30BZ0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery.
Enforcement
- Recall Number
- Z-1396-2019
- Event ID
- 82691
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Boston Scientific Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 29, 2019
- Initiation Date
- April 8, 2019
- Classification Date
- May 22, 2019
- Termination Date
- May 10, 2023
- Address
- 300 Boston Scientific Way, N/A, Marlborough, MA, 01752-1291, United States
Description
AURIGA 30 BRAZIL ZERO COST SYSTEM UPN: M0068S30BZ0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery.
Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.
Serial Number: 21302966, 21303016, 21303021
Worldwide distribution to Australia, Belgium, Brazil, Chile, Colombia, Egypt, France, Germany, Great Britain, India, Indonesia, Iran, Italy, Lebanon, Malaysia, Mexico, Romania, Russia Fed., South Africa, South Korea, Spain, Taiwan, United Arab Emirates and Vietnam
3 (OUS)