Medline Convenience Kits: 1) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069G; 2) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069I; 3) HEAD & NECK CDS-LF, Model Number: CDS983782C; 4) BAPTIST FLOYD BRONCHOSCOPY, Model Number: DYKE1456D; 5) BRONCHOSCOPY SETUP KIT, Model Number: DYKE1955; 6) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156; 7) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156A; 8) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156B; 9) KIT ENT LARYNGECTOMY, Model Number: DYKMBNDL174; 10) KIT ENT LARYNGECTOMY, Model Number: DYKMBNDL174A; 11) KIT ENT LARYNGECTOMY, Model Number: DYKMBNDL174B; 12) KIT PEDS HEAD AND NECK, Model Number: DYKMBNDL196; 13) KIT OMF LEFORTE, Model Number: DYKMBNDL1A; 14) KIT OMF LEFORTE, Model Number: DYKMBNDL1B; 15) KIT OMF LEFORTE, Model Number: DYKMBNDL1C; 16) KIT ENT ENDOSINUS GENERIC, Model Number: DYKMBNDL32; 17) KIT ENT ENDOSINUS GENERIC, Model Number: DYKMBNDL32A; 18) KIT ENT ENDOSINUS GENERIC, Model Number: DYKMBNDL32B; 19) KIT GEN SURG THYROIDECTOMY, Model Number: DYKMBNDL66A; 20) KIT GEN SURG THYROIDECTOMY, Model Number: DYKMBNDL66B; 21) KIT GEN SURG EXCISION THYROID, Model Number: DYKMBNDL88A; 22) KIT GEN SURG EXCISION THYROID, Model Number: DYKMBNDL88B; 23) KIT GEN SURG EXCISION THYROID, Model Number: DYKMBNDL88C; 24) KIT GEN SURG LAP VENTRA, Model Number: DYKMBNDL96C; 25) MINOR ENT PACK, Model Number: DYNJ01831I; 26) ENT BASIC PACK-LF, Model Number: DYNJ0387711F; 27) D AND C PACK-LF, Model Number: DYNJ0855090P; 28) CSS SINUS PACK, Model Number: DYNJ17005I; 29) CSS HEAD & NECK PACK, Model Number: DYNJ26780I; 30) NASALPLASTY #61-RF, Model Number: DYNJ27320V; 31) MASTOID PACK-WINTER PARK-LF, Model Number: DYNJ32060C; 32) T&A PACK, Model Number: DYNJ32809I; 33) HEAD AND NECK PACK, Model Number: DYNJ41169; 34) SIMPLE NECK DISSECTION, Model Number: DYNJ47481C; 35) ENT PACK, Model Number: DYNJ47882F; 36) ENT PACK, Model Number: DYNJ48401C; 37) ENT PACK, Model Number: DYNJ51934A; 38) PK,ORTHO-EXTREMITY-LOSROBLES, Model Number: DYNJ52916A; 39) HEAD & NECK PACK, Model Number: DYNJ54964A; 40) HEAD & NECK PACK, Model Number: DYNJ54964B; 41) ENT - MINOR ENT PACK-LF, Model Number: DYNJ56840B; 42) GLENNON HEAD AND NECK PACK-LF, Model Number: DYNJ57200C; 43) ENT PACK, Model Number: DYNJ57688F; 44) RF T AND A PACK, Model Number: DYNJ61393A; 45) T & A PACK, Model Number: DYNJ61935D; 46) ENT PACK, Model Number: DYNJ62093B; 47) ENT PACK, Model Number: DYNJ64024B; 48) OSCLJ HEAD AND NECK PACK, Model Number: DYNJ66509A; 49) HEAD AND NECK PACK, Model Number: DYNJ67208; 50) HEAD & NECK PACK, Model Number: DYNJ67328B; 51) PEDS HEAD AND NECK PACK, Model Number: DYNJ69746B; 52) NASAL PACK, Model Number: DYNJ81280A; 53) ENT-PLASTICS PACK, Model Number: DYNJ84258A; 54) ENT-PLASTICS PACK, Model Number: DYNJ84258B; 55) HEAD & NECK PACK, Model Number: DYNJ85078A; 56) ENT PACK, Model Number: DYNJ86194A; 57) EENT PACK, Model Number: DYNJ86235; 58) ENT PACK, Model Number: DYNJ87378; 59) ENT PACK, Model Number: DYNJ87378A; 60) ENT PACK, Model Number: DYNJ87378D; 61) ENT SPLIT PACK, Model Number: DYNJ87451; 62) ENT SINUS PACK, Model Number: DYNJ88498; 63) NASAL, Model Number: DYNJ902993F; 64) ENDO SINUS-MIDTOWN, Model Number: DYNJ905866; 65) T A, Model Number: DYNJ906292B; 66) THYROID, Model Number: DYNJ906841A; 67) THYROID - SMH AMB, Model Number: DYNJ907739B; 68) ENT, Model Number: DYNJ911623; 69) HEAD AND NECK PACK, Model Number: DYNJT5739; 70) RO MEDIASTINAL PACK-LF, Model Number: PHS393087007C
Enforcement
- Recall Number
- Z-1394-2026
- Event ID
- 98329
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 25, 2026
- Initiation Date
- January 7, 2026
- Classification Date
- February 17, 2026
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Medline Convenience Kits: 1) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069G; 2) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069I; 3) HEAD & NECK CDS-LF, Model Number: CDS983782C; 4) BAPTIST FLOYD BRONCHOSCOPY, Model Number: DYKE1456D; 5) BRONCHOSCOPY SETUP KIT, Model Number: DYKE1955; 6) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156; 7) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156A; 8) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156B; 9) KIT ENT LARYNGECTOMY, Model Number: DYKMBNDL174; 10) KIT ENT LARYNGECTOMY, Model Number: DYKMBNDL174A; 11) KIT ENT LARYNGECTOMY, Model Number: DYKMBNDL174B; 12) KIT PEDS HEAD AND NECK, Model Number: DYKMBNDL196; 13) KIT OMF LEFORTE, Model Number: DYKMBNDL1A; 14) KIT OMF LEFORTE, Model Number: DYKMBNDL1B; 15) KIT OMF LEFORTE, Model Number: DYKMBNDL1C; 16) KIT ENT ENDOSINUS GENERIC, Model Number: DYKMBNDL32; 17) KIT ENT ENDOSINUS GENERIC, Model Number: DYKMBNDL32A; 18) KIT ENT ENDOSINUS GENERIC, Model Number: DYKMBNDL32B; 19) KIT GEN SURG THYROIDECTOMY, Model Number: DYKMBNDL66A; 20) KIT GEN SURG THYROIDECTOMY, Model Number: DYKMBNDL66B; 21) KIT GEN SURG EXCISION THYROID, Model Number: DYKMBNDL88A; 22) KIT GEN SURG EXCISION THYROID, Model Number: DYKMBNDL88B; 23) KIT GEN SURG EXCISION THYROID, Model Number: DYKMBNDL88C; 24) KIT GEN SURG LAP VENTRA, Model Number: DYKMBNDL96C; 25) MINOR ENT PACK, Model Number: DYNJ01831I; 26) ENT BASIC PACK-LF, Model Number: DYNJ0387711F; 27) D AND C PACK-LF, Model Number: DYNJ0855090P; 28) CSS SINUS PACK, Model Number: DYNJ17005I; 29) CSS HEAD & NECK PACK, Model Number: DYNJ26780I; 30) NASALPLASTY #61-RF, Model Number: DYNJ27320V; 31) MASTOID PACK-WINTER PARK-LF, Model Number: DYNJ32060C; 32) T&A PACK, Model Number: DYNJ32809I; 33) HEAD AND NECK PACK, Model Number: DYNJ41169; 34) SIMPLE NECK DISSECTION, Model Number: DYNJ47481C; 35) ENT PACK, Model Number: DYNJ47882F; 36) ENT PACK, Model Number: DYNJ48401C; 37) ENT PACK, Model Number: DYNJ51934A; 38) PK,ORTHO-EXTREMITY-LOSROBLES, Model Number: DYNJ52916A; 39) HEAD & NECK PACK, Model Number: DYNJ54964A; 40) HEAD & NECK PACK, Model Number: DYNJ54964B; 41) ENT - MINOR ENT PACK-LF, Model Number: DYNJ56840B; 42) GLENNON HEAD AND NECK PACK-LF, Model Number: DYNJ57200C; 43) ENT PACK, Model Number: DYNJ57688F; 44) RF T AND A PACK, Model Number: DYNJ61393A; 45) T & A PACK, Model Number: DYNJ61935D; 46) ENT PACK, Model Number: DYNJ62093B; 47) ENT PACK, Model Number: DYNJ64024B; 48) OSCLJ HEAD AND NECK PACK, Model Number: DYNJ66509A; 49) HEAD AND NECK PACK, Model Number: DYNJ67208; 50) HEAD & NECK PACK, Model Number: DYNJ67328B; 51) PEDS HEAD AND NECK PACK, Model Number: DYNJ69746B; 52) NASAL PACK, Model Number: DYNJ81280A; 53) ENT-PLASTICS PACK, Model Number: DYNJ84258A; 54) ENT-PLASTICS PACK, Model Number: DYNJ84258B; 55) HEAD & NECK PACK, Model Number: DYNJ85078A; 56) ENT PACK, Model Number: DYNJ86194A; 57) EENT PACK, Model Number: DYNJ86235; 58) ENT PACK, Model Number: DYNJ87378; 59) ENT PACK, Model Number: DYNJ87378A; 60) ENT PACK, Model Number: DYNJ87378D; 61) ENT SPLIT PACK, Model Number: DYNJ87451; 62) ENT SINUS PACK, Model Number: DYNJ88498; 63) NASAL, Model Number: DYNJ902993F; 64) ENDO SINUS-MIDTOWN, Model Number: DYNJ905866; 65) T A, Model Number: DYNJ906292B; 66) THYROID, Model Number: DYNJ906841A; 67) THYROID - SMH AMB, Model Number: DYNJ907739B; 68) ENT, Model Number: DYNJ911623; 69) HEAD AND NECK PACK, Model Number: DYNJT5739; 70) RO MEDIASTINAL PACK-LF, Model Number: PHS393087007C
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
1) CDS983069G, UDI-DI: 10193489923926(each), 40193489923927(case), Lot Number: 21IBD399; 2) CDS983069G, UDI-DI: 10193489923926(each), 40193489923927(case), Lot Number: 21HBO921; 3) CDS983069I, UDI-DI: 10195327071868(each), 40195327071869(case), Lot Number: 22JBW315; 4) CDS983069I, UDI-DI: 10195327071868(each), 40195327071869(case), Lot Number: 22GBE731; 5) CDS983069I, UDI-DI: 10195327071868(each), 40195327071869(case), Lot Number: 22FBX804; 6) CDS983782C, UDI-DI: 10193489592047(each), 40193489592048(case), Lot Number: 21DKA756; 7) DYKE1456D, UDI-DI: 10195327639938(each), 40195327639939(case), Lot Number: 24CBK289; 8) DYKE1955, UDI-DI: 10195327539832(each), 40195327539833(case), Lot Number: 25HLA345; 9) DYKE1955, UDI-DI: 10195327539832(each), 40195327539833(case), Lot Number: 25FLA904; 10) DYKE1955, UDI-DI: 10195327539832(each), 40195327539833(case), Lot Number: 25DLB145; 11) DYKE1955, UDI-DI: 10195327539832(each), 40195327539833(case), Lot Number: 25CLB002; 12) DYKE1955, UDI-DI: 10195327539832(each), 40195327539833(case), Lot Number: 24DLA606; 13) DYKE1955, UDI-DI: 10195327539832(each), 40195327539833(case), Lot Number: 24ALA266; 14) DYKMBNDL156, UDI-DI: 10193489478242(each), 40193489478243(case), Lot Number: 21FBS478; 15) DYKMBNDL156, UDI-DI: 10193489478242(each), 40193489478243(case), Lot Number: 21FBO537; 16) DYKMBNDL156, UDI-DI: 10193489478242(each), 40193489478243(case), Lot Number: 21DBJ855; 17) DYKMBNDL156, UDI-DI: 10193489478242(each), 40193489478243(case), Lot Number: 21CBD340; 18) DYKMBNDL156, UDI-DI: 10193489478242(each), 40193489478243(case), Lot Number: 21BBG062; 19) DYKMBNDL156, UDI-DI: 10193489478242(each), 40193489478243(case), Lot Number: 21ABK323; 20) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 22GBH498; 21) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 22FBT038; 22) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 22EBS812; 23) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 22DBL428; 24) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 22CBX230; 25) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 22BBA231; 26) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 22ABG015; 27) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 21JBB014; 28) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 21JBA009; 29) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 21IBP624; 30) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 21HBF494; 31) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 21GBE737; 32) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 25GBX739; 33) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 25EBH975; 34) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 25DBQ008; 35) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 25BBM402; 36) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 25ABK769; 37) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 24KBJ157; 38) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 24HBU706; 39) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 24GBR935; 40) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 24GBD477; 41) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 24EBF125; 42) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 24DBV609; 43) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 24ABL113; 44) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23JBT911; 45) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23JBQ487; 46) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23HBV381; 47) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23HBT877; 48) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23FBG212; 49) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23EBS366; 50) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23DBM060; 51) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23ABQ109; 52) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 22KBH786; 53) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 22IBU647; 54) DYKMBNDL174, UDI-DI: 10193489478334(each), 40193489478335(case), Lot Number: 21FBQ554; 55) DYKMBNDL174, UDI-DI: 10193489478334(each), 40193489478335(case), Lot Number: 21BBP836; 56) DYKMBNDL174, UDI-DI: 10193489478334(each), 40193489478335(case), Lot Number: 21ABH824; 57) DYKMBNDL174A, UDI-DI: 10193489951882(each), 40193489951883(case), Lot Number: 22EBO543; 58) DYKMBNDL174A, UDI-DI: 10193489951882(each), 40193489951883(case), Lot Number: 22BBS429; 59) DYKMBNDL174A, UDI-DI: 10193489951882(each), 40193489951883(case), Lot Number: 21LBY299; 60) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 25LBM606; 61) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 24EBP260; 62) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 24DBB691; 63) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 24CBA389; 64) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 23JBN238; 65) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 23ABH675; 66) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 22LBI002; 67) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 22KBK464; 68) DYKMBNDL196, UDI-DI: 10193489479393(each), 40193489479394(case), Lot Number: 21CBT411; 69) DYKMBNDL196, UDI-DI: 10193489479393(each), 40193489479394(case), Lot Number: 21CBS041; 70) DYKMBNDL196, UDI-DI: 10193489479393(each), 40193489479394(case), Lot Number: 21CBM511; 71) DYKMBNDL196, UDI-DI: 10193489479393(each), 40193489479394(case), Lot Number: 21CBH559; 72) DYKMBNDL196, UDI-DI: 10193489479393(each), 40193489479394(case), Lot Number: 21CBF521; 73) DYKMBNDL196, UDI-DI: 10193489479393(each), 40193489479394(case), Lot Number: 21BBG851; 74) DYKMBNDL196, UDI-DI: 10193489479393(each), 40193489479394(case), Lot Number: 21ABP337; 75) DYKMBNDL1A, UDI-DI: 10195327245832(each), 40195327245833(case), Lot Number: 23BLA884; 76) DYKMBNDL1A, UDI-DI: 10195327245832(each), 40195327245833(case), Lot Number: 22KLA452; 77) DYKMBNDL1A, UDI-DI: 10195327245832(each), 40195327245833(case), Lot Number: 22JLB001; 78) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 25FLA924; 79) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 25FLA791; 80) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 25DLA251; 81) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 25CLA312; 82) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 25BLA837; 83) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 24JLA055; 84) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 24ILA560; 85) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 24HLB156; 86) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 24HLA924; 87) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 24GLA055; 88) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 24FLA349; 89) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 23LLA422; 90) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 23KLA620; 91) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 23JLA058; 92) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 23ILA756; 93) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 23HLA624; 94) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 23DLA495; 95) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 23CLA769; 96) DYKMBNDL1C, UDI-DI: 10198459448591(each), 40198459448592(case), Lot Number: 25ILA386; 97) DYKMBNDL1C, UDI-DI: 10198459448591(each), 40198459448592(case), Lot Number: 25ILA035; 98) DYKMBNDL1C, UDI-DI: 10198459448591(each), 40198459448592(case), Lot Number: 25HLA843; 99) DYKMBNDL1C, UDI-DI: 10198459448591(each), 40198459448592(case), Lot Number: 25HLA232; 100) DYKMBNDL1C, UDI-DI: 10198459448591(each), 40198459448592(case), Lot Number: 25HLA179; 101) DYKMBNDL32, UDI-DI: 10193489478723(each), 40193489478724(case), Lot Number: 21FBT798; 102) DYKMBNDL32, UDI-DI: 10193489478723(each), 40193489478724(case), Lot Number: 21EBM706; 103) DYKMBNDL32, UDI-DI: 10193489478723(each), 40193489478724(case), Lot Number: 21DBK216; 104) DYKMBNDL32, UDI-DI: 10193489478723(each), 40193489478724(case), Lot Number: 21CBS031; 105) DYKMBNDL32, UDI-DI: 10193489478723(each), 40193489478724(case), Lot Number: 21BBL111; 106) DYKMBNDL32, UDI-DI: 10193489478723(each), 40193489478724(case), Lot Number: 21ABJ985; 107) DYKMBNDL32, UDI-DI: 10193489478723(each), 40193489478724(case), Lot Number: 21ABJ862; 108) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 22HBP925; 109) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 22GBQ602; 110) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 22GBF522; 111) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 22FBE193; 112) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 22EBQ053; 113) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 22EBO633; 114) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 21JBJ743; 115) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 21IBD784; 116) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 24GBP188; 117) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 24GBC830; 118) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 24EBU213; 119) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 24DBU451; 120) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 24CBR483; 121) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 24ABK306; 122) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 24ABH039; 123) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23LBP166; 124) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23IBA611; 125) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23HBX217; 126) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23HBH678; 127) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23FBP266; 128) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23EBT500; 129) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23EBB247; 130) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23DBP706; 131) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23BBS356; 132) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 22JBW957; 133) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 22IBL167; 134) DYKMBNDL66A, UDI-DI: 10195327240349(each), 40195327240340(case), Lot Number: 24DBD514; 135) DYKMBNDL66A, UDI-DI: 10195327240349(each), 40195327240340(case), Lot Number: 24CBD159; 136) DYKMBNDL66A, UDI-DI: 10195327240349(each), 40195327240340(case), Lot Number: 24BBT350; 137) DYKMBNDL66A, UDI-DI: 10195327240349(each), 40195327240340(case), Lot Number: 23LBF184; 138) DYKMBNDL66A, UDI-DI: 10195327240349(each), 40195327240340(case), Lot Number: 22LBN128; 139) DYKMBNDL66A, UDI-DI: 10195327240349(each), 40195327240340(case), Lot Number: 22JBQ673; 140) DYKMBNDL66A, UDI-DI: 10195327240349(each), 40195327240340(case), Lot Number: 22JBG017; 141) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 25LBI754; 142) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 25JBF754; 143) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 25CBR258; 144) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 25ABH117; 145) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 24LBT153; 146) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 24KBT017; 147) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 24IBO509; 148) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 24HBH734; 149) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 24FBG015; 150) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 24DBR155; 151) DYKMBNDL88A, UDI-DI: 10195327240400(each), 40195327240401(case), Lot Number: 23CBN843; 152) DYKMBNDL88A, UDI-DI: 10195327240400(each), 40195327240401(case), Lot Number: 23CBJ801; 153) DYKMBNDL88A, UDI-DI: 10195327240400(each), 40195327240401(case), Lot Number: 23ABK217; 154) DYKMBNDL88A, UDI-DI: 10195327240400(each), 40195327240401(case), Lot Number: 22LBR829; 155) DYKMBNDL88A, UDI-DI: 10195327240400(each), 40195327240401(case), Lot Number: 22JBW066; 156) DYKMBNDL88A, UDI-DI: 10195327240400(each), 40195327240401(case), Lot Number: 22JBD610; 157) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 25FBT988; 158) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 25EBU506; 159) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 25EBK536; 160) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 25CBI877; 161) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 25ABS392; 162) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 25ABH118; 163) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24JBX567; 164) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24JBL602; 165) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24GBU190; 166) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24GBN944; 167) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24EBU598; 168) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24EBF126; 169) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24DBS395; 170) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24BBL266; 171) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24BBJ842; 172) DYKMBNDL88C, UDI-DI: 10198459584886(each), 40198459584887(case), Lot Number: 25LBP011; 173) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 25GBG516; 174) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 24KBA441; 175) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 24IBQ080; 176) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 24HBN645; 177) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 24HBJ497; 178) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 24GBH580; 179) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 24DBR908; 180) DYNJ01831I, UDI-DI: 10889942572502(each), 40889942572503(case), Lot Number: 21EMA909; 181) DYNJ0387711F, UDI-DI: 10198459005893(each), 40198459005894(case), Lot Number: 24HMA304; 182) DYNJ0855090P, UDI-DI: 10198459605826(each), 40198459605827(case), Lot Number: 25LMA083; 183) DYNJ17005I, UDI-DI: 10193489241860(each), 40193489241861(case), Lot Number: 23HBV662; 184) DYNJ26780I, UDI-DI: 10193489238822(each), 40193489238823(case), Lot Number: 23GBP197; 185) DYNJ27320V, UDI-DI: 10195327104719(each), 40195327104710(case), Lot Number: 22LBQ654; 186) DYNJ32060C, UDI-DI: 10884389637265(each), 40884389637266(case), Lot Number: 22LBS348; 187) DYNJ32809I, UDI-DI: 10193489448795(each), 40193489448796(case), Lot Number: 22IBP091; 188) DYNJ41169, UDI-DI: 10884389906699(each), 40884389906690(case), Lot Number: 22LBR515; 189) DYNJ47481C, UDI-DI: 10889942544028(each), 40889942544029(case), Lot Number: 21BBP023; 190) DYNJ47882F, UDI-DI: 10195327319472(each), 40195327319473(case), Lot Number: 23HME434; 191) DYNJ48401C, UDI-DI: 10198459175749(each), 40198459175740(case), Lot Number: 25ABD063; 192) DYNJ51934A, UDI-DI: 10193489584776(each), 40193489584777(case), Lot Number: 23HMG725; 193) DYNJ52916A, UDI-DI: 10195327311735(each), 40195327311736(case), Lot Number: 24GMK024; 194) DYNJ54964A, UDI-DI: 10198459033483(each), 40198459033484(case), Lot Number: 25DBB366; 195) DYNJ54964A, UDI-DI: 10198459033483(each), 40198459033484(case), Lot Number: 25BBM674; 196) DYNJ54964A, UDI-DI: 10198459033483(each), 40198459033484(case), Lot Number: 24LBN788; 197) DYNJ54964A, UDI-DI: 10198459033483(each), 40198459033484(case), Lot Number: 24IBV894; 198) DYNJ54964B, UDI-DI: 10198459428814(each), 40198459428815(case), Lot Number: 25KBH303; 199) DYNJ54964B, UDI-DI: 10198459428814(each), 40198459428815(case), Lot Number: 25HBR203; 200) DYNJ56840B, UDI-DI: 10193489416077(each), 40193489416078(case), Lot Number: 22GBJ777; 201) DYNJ56840B, UDI-DI: 10193489416077(each), 40193489416078(case), Lot Number: 21IBQ699; 202) DYNJ56840B, UDI-DI: 10193489416077(each), 40193489416078(case), Lot Number: 21IBQ696; 203) DYNJ57200C, UDI-DI: 10198459306754(each), 40198459306755(case), Lot Number: 25EMA242; 204) DYNJ57688F, UDI-DI: 10195327034610(each), 40195327034611(case), Lot Number: 22HMH769; 205) DYNJ61393A, UDI-DI: 10193489444179(each), 40193489444170(case), Lot Number: 23CMB523; 206) DYNJ61935D, UDI-DI: 10195327453336(each), 40195327453337(case), Lot Number: 24BBK459; 207) DYNJ61935D, UDI-DI: 10195327453336(each), 40195327453337(case), Lot Number: 24BBI869; 208) DYNJ61935D, UDI-DI: 10195327453336(each), 40195327453337(case), Lot Number: 23LBP950; 209) DYNJ61935D, UDI-DI: 10195327453336(each), 40195327453337(case), Lot Number: 23HBC470; 210) DYNJ62093B, UDI-DI: 10193489969351(each), 40193489969352(case), Lot Number: 23ABE410; 211) DYNJ64024B, UDI-DI: 10195327556617(each), 40195327556618(case), Lot Number: 24KBS794; 212) DYNJ64024B, UDI-DI: 10195327556617(each), 40195327556618(case), Lot Number: 24HBO119; 213) DYNJ64024B, UDI-DI: 10195327556617(each), 40195327556618(case), Lot Number: 24EBS530; 214) DYNJ64024B, UDI-DI: 10195327556617(each), 40195327556618(case), Lot Number: 24DBV691; 215) DYNJ64024B, UDI-DI: 10195327556617(each), 40195327556618(case), Lot Number: 24ABK239; 216) DYNJ66509A, UDI-DI: 10193489462708(each), 40193489462709(case), Lot Number: 24HMA316; 217) DYNJ67208, UDI-DI: 10193489391671(each), 40193489391672(case), Lot Number: 20XBA769; 218) DYNJ67328B, UDI-DI: 10195327271503(each), 40195327271504(case), Lot Number: 23HMD036; 219) DYNJ69746B, UDI-DI: 10195327475222(each), 40195327475223(case), Lot Number: 24JLA247; 220) DYNJ81280A, UDI-DI: 10195327491284(each), 40195327491285(case), Lot Number: 25LME467; 221) DYNJ84258A, UDI-DI: 10195327650742(each), 40195327650743(case), Lot Number: 24HBK264; 222) DYNJ84258A, UDI-DI: 10195327650742(each), 40195327650743(case), Lot Number: 24DBR594; 223) DYNJ84258A, UDI-DI: 10195327650742(each), 40195327650743(case), Lot Number: 24CBI987; 224) DYNJ84258A, UDI-DI: 10195327650742(each), 40195327650743(case), Lot Number: 24CBF775; 225) DYNJ84258B, UDI-DI: 10198459117985(each), 40198459117986(case), Lot Number: 25FBL492; 226) DYNJ84258B, UDI-DI: 10198459117985(each), 40198459117986(case), Lot Number: 25EBG049; 227) DYNJ84258B, UDI-DI: 10198459117985(each), 40198459117986(case), Lot Number: 24IBP146; 228) DYNJ84258B, UDI-DI: 10198459117985(each), 40198459117986(case), Lot Number: 24IBD271; 229) DYNJ85078A, UDI-DI: 10195327532543(each), 40195327532544(case), Lot Number: 24HMA736; 230) DYNJ86194A, UDI-DI: 10195327682293(each), 40195327682294(case), Lot Number: 24DBN706; 231) DYNJ86235, UDI-DI: 10195327606534(each), 40195327606535(case), Lot Number: 25LMD475; 232) DYNJ86235, UDI-DI: 10195327606534(each), 40195327606535(case), Lot Number: 24IBL260; 233) DYNJ86235, UDI-DI: 10195327606534(each), 40195327606535(case), Lot Number: 24HBV552; 234) DYNJ86235, UDI-DI: 10195327606534(each), 40195327606535(case), Lot Number: 24CBG395; 235) DYNJ87378, UDI-DI: 10198459034787(each), 40198459034788(case), Lot Number: 24HMA738; 236) DYNJ87378, UDI-DI: 10198459034787(each), 40198459034788(case), Lot Number: 24GMG790; 237) DYNJ87378A, UDI-DI: 10198459186394(each), 40198459186395(case), Lot Number: 24KMI285; 238) DYNJ87378D, UDI-DI: 10198459467257(each), 40198459467258(case), Lot Number: 25JMI380; 239) DYNJ87378D, UDI-DI: 10198459467257(each), 40198459467258(case), Lot Number: 25IMC112; 240) DYNJ87378D, UDI-DI: 10198459467257(each), 40198459467258(case), Lot Number: 25IMA065; 241) DYNJ87451, UDI-DI: 10198459040283(each), 40198459040284(case), Lot Number: 24HBH452; 242) DYNJ88498, UDI-DI: 10198459166181(each), 40198459166182(case), Lot Number: 24KMF610; 243) DYNJ88498, UDI-DI: 10198459166181(each), 40198459166182(case), Lot Number: 24KMC641; 244) DYNJ902993F, UDI-DI: 10193489490817(each), 40193489490818(case), Lot Number: 23IMD509; 245) DYNJ902993F, UDI-DI: 10193489490817(each), 40193489490818(case), Lot Number: 23IMA480; 246) DYNJ905866, UDI-DI: 10193489547184(each), 40193489547185(case), Lot Number: 23CBW319; 247) DYNJ906292B, UDI-DI: 10193489877588(each), 40193489877589(case), Lot Number: 24CBI703; 248) DYNJ906292B, UDI-DI: 10193489877588(each), 40193489877589(case), Lot Number: 24BBV224; 249) DYNJ906841A, UDI-DI: 10193489392197(each), 40193489392198(case), Lot Number: 21HBN905; 250) DYNJ906841A, UDI-DI: 10193489392197(each), 40193489392198(case), Lot Number: 21GBR282; 251) DYNJ906841A, UDI-DI: 10193489392197(each), 40193489392198(case), Lot Number: 21GBG669; 252) DYNJ907739B, UDI-DI: 10195327671778(each), 40195327671779(case), Lot Number: 25ABP522; 253) DYNJ907739B, UDI-DI: 10195327671778(each), 40195327671779(case), Lot Number: 25ABE814; 254) DYNJ911623, UDI-DI: 10198459441097(each), 40198459441098(case), Lot Number: 25HBH612; 255) DYNJT5739, UDI-DI: 10198459499661(each), 40198459499662(case), Lot Number: 25IMF120; 256) PHS393087007C, UDI-DI: 10193489222777(each), 40193489222778(case), Lot Number: 21ADA448
Worldwide distribution - US Nationwide.
9051 units