INFX-8000V Interventional Angiography System. CAT-880B catheterization table. Cardiac and Vascular intervention with fluoroscopic and fluorographic images. For the USA and Canada the system is registered as lnfinix CF-i or lnfinix VF-i, for national regulations. INFX-8000V series includes a floor mounted multi-axis positioner.
Enforcement
- Recall Number
- Z-1394-2015
- Event ID
- 70748
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Toshiba American Medical Systems Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 15, 2015
- Initiation Date
- March 11, 2015
- Classification Date
- April 6, 2015
- Termination Date
- September 18, 2015
- Address
- 2441 Michelle Dr, P.O. Box 2068, Tustin, CA, 92780-7047, United States
Description
INFX-8000V Interventional Angiography System. CAT-880B catheterization table. Cardiac and Vascular intervention with fluoroscopic and fluorographic images. For the USA and Canada the system is registered as lnfinix CF-i or lnfinix VF-i, for national regulations. INFX-8000V series includes a floor mounted multi-axis positioner.
The table base rotational brake that holds the table top may not always be engaged when any force or side impact is applied to the table top.
Serial No. U7E1462009, ACA1472001, UEB1352004, UFD1442017, AAA1492001, ACA1492002, ULD1412014, AHA1482001, UPB1372006, ULD1392013, U6A1472001, ULD1452016, UPB1352005, ULD1452015, UDE1352017, A4B1472002, U4A14X2001, UPB1342004, 99B1342133, 99B1342134, UFD1452018, UFD13X2015, UFD13X2016, UFD1362014, UUC1452012, A6A1492002, UUC13Z2011, AAA14X2002, A4A13Z2001.
Nationwide Distribution
28 units