FDA Enforcement Class II Terminated

Ingenuity TF PET/CT, Model No. 882442 Product Usage: The device is an integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) system suitable for a wide range of diagnostic applications. The device utilizes the CT technology to obtain anatomic images of the human body and PET technology to obtain functional images of the human body. The clinical value of both technologies increases with the capability to fuse the CT and PET images using Philips fusion viewer Image Fusion software to create a composite image for diagnostic study and therapeutic planning. The system also provides tools for the quantification of results of the CT and PET images and provides the means for a simplified review of the CT, PET, and fused images. The integration of the anatomical data from CT with the metabolic data from PET gives clinicians the visual information necessary to define the severity, as well as the extent, of the disease.

Recall: Z-1392-2018 · Reported April 25, 2018

Enforcement

Recall Number
Z-1392-2018
Event ID
79696
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 25, 2018
Initiation Date
February 15, 2018
Classification Date
April 17, 2018
Termination Date
November 21, 2019
Address
595 Miner Rd, Cleveland, OH, 44143-2131, United States

Description

Ingenuity TF PET/CT, Model No. 882442 Product Usage: The device is an integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) system suitable for a wide range of diagnostic applications. The device utilizes the CT technology to obtain anatomic images of the human body and PET technology to obtain functional images of the human body. The clinical value of both technologies increases with the capability to fuse the CT and PET images using Philips fusion viewer Image Fusion software to create a composite image for diagnostic study and therapeutic planning. The system also provides tools for the quantification of results of the CT and PET images and provides the means for a simplified review of the CT, PET, and fused images. The integration of the anatomical data from CT with the metabolic data from PET gives clinicians the visual information necessary to define the severity, as well as the extent, of the disease.

Reason

Following a period of inactivity, the mass storage device may cause the acquisition console to become unresponsive. This may prevent an acquisition from proceeding.

Code Info

Serial No: 2090 2072 2077 2079 2086 2091 2093 2085 2076 2071 2078 2081

Distribution

Worldwide Distribution - US Nationwide in the states of: OH and OR, and Foreign distribution to Australia, China, Cuba, France, and South Africa.

Quantity

12