FDA Enforcement Class III Terminated

Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-503 Product Usage: Used in facilities such as Central Sterile Supply Departments, next to the Operating Room, in dental clinics or hospital wards to clean, thermally or chemically disinfect and dry surgical goods.

Recall: Z-1391-2020 · Reported March 4, 2020

Enforcement

Recall Number
Z-1391-2020
Event ID
84885
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Maquet Cardiovascular Us Sales, Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 4, 2020
Initiation Date
January 16, 2020
Classification Date
February 27, 2020
Termination Date
October 23, 2023
Address
45 Barbour Pond Dr, N/A, Wayne, NJ, 07470-2094, United States

Description

Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-503 Product Usage: Used in facilities such as Central Sterile Supply Departments, next to the Operating Room, in dental clinics or hospital wards to clean, thermally or chemically disinfect and dry surgical goods.

Reason

Due to a missing water intake for Deionized water on the Getinge 46-Series Washer Disinfector. There were no valve, hose, or clamps for this option included with the device therefore the units could not be properly installed. There have been no adverse events reported resulting in serious illness or injuries.

Code Info

Serial Numbers: WAA090094, WAA090112, WAA090409, WAA090521, WAA090757, WAA090805, WAA091170, WAA091288, WAA091292, WAA091454, WAA091460 WAA091463, WAA091477. UDI Code 07340153700048

Distribution

U.S. Distribution: CA, CO, IA, MN, NY, SC, VA

Quantity

7 units