FDA Enforcement Class II Terminated

Human Fetal Hemoglobin (HBF-1) FITC Conjugate

Recall: Z-1388-2018 · Reported April 25, 2018

Enforcement

Recall Number
Z-1388-2018
Event ID
79603
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Life Technologies Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 25, 2018
Initiation Date
February 7, 2018
Classification Date
April 17, 2018
Termination Date
May 8, 2019
Address
7335 Executive Way, Frederick, MD, 21704-8354, United States

Description

Human Fetal Hemoglobin (HBF-1) FITC Conjugate

Reason

Three lots of Human Fetal Hemoglobin FITC Conjugate are displaying high background and unusual bright staining in the controls and patient samples which could potentially lead to erroneous results (i.e. false positives)

Code Info

Model # MHFH01 UDI (01)10190302005579(17)180430 (10)1626670D(240)MHFH01 Lot# 1626670D Model # MHFH01 UDI (01)10190302005579(17)190830 (10)1873068B(240)MHFH01 Lot# 1873068B Model # MHFH01 UDI (01)10190302005579(17)191030 (10)1915224B(240)MHFH01 Lot# 1915224B

Distribution

MI, CA, IL, TX, SC, NY, MN, DE, NJ, WA , OR, IN, MA, DE, FL, PA