FDA Enforcement Class II Terminated

Cold Storage Solution, 2L, labeled as Waters IGL Cold Storage Solution, Catalog No. CSS/2000/US; and, Belzer UW Cold Storage Solution, Catalog No. BUW_002. Intended for the hypothermic flushing and storage of specific organs in preparation for storage, transportation, and transplantation into a recipient.

Recall: Z-1387-2018 · Reported April 25, 2018

Enforcement

Recall Number
Z-1387-2018
Event ID
79378
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Preservation Solutions, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 25, 2018
Initiation Date
February 26, 2018
Classification Date
April 16, 2018
Termination Date
September 9, 2022
Address
1099 Proctor Dr, Elkhorn, WI, 53121-2027, United States

Description

Cold Storage Solution, 2L, labeled as Waters IGL Cold Storage Solution, Catalog No. CSS/2000/US; and, Belzer UW Cold Storage Solution, Catalog No. BUW_002. Intended for the hypothermic flushing and storage of specific organs in preparation for storage, transportation, and transplantation into a recipient.

Reason

Bags of solution may have a small leak at the base of the injection port where the port is welded into the body of the bag..

Code Info

UDI 00810805000163, and UDI 00810805000071.

Distribution

Worldwide Distribution -- USA and Europe.

Quantity

31,104 units