da Vinci S / Si Surgical System IS2000 / IS3000, Instrument Arm Drape, 20 Pack. Microtek Medical Equipment Drapes are to be used to cover a variety of surgical and non-surgical equipment in various settings throughout the clinical setting.
Enforcement
- Recall Number
- Z-1387-2015
- Event ID
- 70777
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Intuitive Surgical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 15, 2015
- Initiation Date
- March 16, 2015
- Classification Date
- April 6, 2015
- Termination Date
- December 3, 2015
- Address
- 1266 Kifer Rd Bldg 100, N/A, Sunnyvale, CA, 94086-5304, United States
Description
da Vinci S / Si Surgical System IS2000 / IS3000, Instrument Arm Drape, 20 Pack. Microtek Medical Equipment Drapes are to be used to cover a variety of surgical and non-surgical equipment in various settings throughout the clinical setting.
Cloudy/waxy appearance with potential to transfer waxy substance to the patient, and potential for drape to tear.
Product number 420015-03; da Vinci S / Si Instrument Arm Drape, 20 Pack. Lot Numbers for Cloudy/Waxy issue: ALL lot numbers Lot Numbers for Tear issue: D142175, D142535, D142895, D143255, D150135A, D150555, DA142695A, DA143465A, D142175A, D142535A, D142895A, D143255A, D150145, D150555A, DA142725, DA143515, D142185, D142545, D142905, D143285, D150145A, D150565, DA142745, DA150075, D142185A, D142545A, D142905A, D143285A, D150155, D150565A, DA142745A, DA150085, D142195, D142555, D142915, D143295, D150155A, D150575, DA142755, DA150085A, D142195A, D142555A, D142915A, D143295A, D150165, D150575A, DA142755A, DA150125, D142205, D142585, D142935, D143305, D150165A, D150625, DA142765, DA150125A, D142205A, D142585A, D142935A, D143305A, D150195, D150625A, DA142795, DA150135, D142215, D142595, D142945, D143315, D150195A, D150635, DA142795A, DA150135A, D142235, D142595A, D142945, D143315, D150205, D150635A, DA142825, DA150195, D142235A, D142605, D142955, D143325, D150205A, D150645, DA142835, DA150205, D142245, D142605A, D142955A, D143325A, D150215, D150645A, DA142835A, DA150205A, D142245A, D142615, D142965, D143335, D150215A, D150655, DA142885, DA150215, D142255, D142615A, D142965A, D143355, D150225, D150655A, DA142885A, DA150275, D142255A, D142625, D142975, D143355A, D150225A, D150685, DA142905, DA150275A, D142265, D142625A, D142975A, D143365, D150235, D150685A, DA142955, DA150285, D142265A, D142635, D142985, D143365A, D150235A, D150695, DA142965, DA150285A, D142275, D142655, D143005, D143375, D150275, D150695A, DA142965A, DA150295, D142275A, D142655A, D143005A, D143375A, D150275A, DA142185, DA142975, DA150335, D142305, D142665, D143015, D143385, D150285, DA142195, DA143005, DA150345, D142305A, D142665A, D143015A, D143385A, D150285A, DA142205, DA143025, DA150345A, D142315, D142675, D143025, D143395, D150295, DA142235, DA143025A, DA150365, D142315A, D142675A, D143025A, D143395A, D150295A, DA142235A, DA143035, DA150405, D142325, D142685, D143035, D143405, D150305, DA142245, DA143035A, DA150405A, D142325A, D142685A, D143035A, D143425, D150305A, DA142255, DA143085, DA150415, D142335, D142695, D143045, D143425A, D150335, DA142275, DA143095, DA150425, D142335A, D142695A, D143045A, D143435, D150335A, DA142305, DA143105, DA150435, D142345A, D142725, D143075, D143435A, D150345, DA142305A, DA143105A, DA150435A, D142375, D142725A, D143075A, D143445, D150345A, DA142315, DA143115, DA150475, D142375A, D142735, D143085, D143445A, D150355, DA142345, DA143155, DA150475A, D142385, D142735A, D143085A, D143455, D150355A, DA142345A, DA143155A, DA150495, D142385A, D142745, D143095, D143455A, D150375, DA142375, DA143165, DA150495A, D142395, D142745A, D143095A, D143465, D150405, DA142375A, DA143175, DA150515, D142395A, D142755, D143105, D143465A, D150405A, DA142385, DA143175A, DA150515A, D142405, D142755A, D143105A, D143495, D150415, DA142395, DA143185, DA150545, D142405A, D142765, D143115, D143495A, D150415A, DA142455, DA143215, DA150545A, D142415, D142765A, D143115A, D143505, D150425, DA142465, DA143235, DA150555, D142415A, D142795, D143155, D143505A, D150425A, DA142485, DA143235A, DA150555A, D142445, D142795A, D143155A, D143515, D150435, DA142515, DA143245, DA150635, D142445A, D142805, D143165, D143515A, D150435A, DA142515A, DA143295, DA150645, D142455, D142805A, D143165A, D150035, D150435A, DA142525, DA143295A, DA150645, D142455A, D142815, D143175, D150065, D150475, DA142535, DA143315, DA150645A, D142465, D142815A, D143175A, D150065A, D150475A, DA142535A, DA143315A, DA150655, D142465A, D142825, D143185, D150075, D150485, DA142585, DA143365, DA150685A, D142475, D142825A, D143185A, D150075A, D150485A, DA142585A, DA143365A, DA150685A, D142475A, D142835, D143215, D150085, D150495, DA142605, DA143385, DA150695, D142485, D142835A, D143215A, D150085A, D150495A, DA142615, DA143385A, DA150695A, D142485A, D142865, D143225, D150095, D150505, DA142615A, DA143395, D142515, D142875, D143225A, D150095A, D150505A, DA142625, DA143435, D142515A, D142875A, D143235, D150105, D150515, DA142655, DA143445, D142525, D142885, D143235A, D150125, D150515A, DA142665, DA143445A, D142525, D142885A, D143245, D150125A, D150545, DA142685, DA143455, D142525A, D142885A, D143245A, D150135, D150545A, DA142695, DA143465
Worldwide Distribution-Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China (including Hong Kong), Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, France, Germany, Greece, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Luxembourg, Malaysia, Mauritius, Mexico, Monaco, Netherlands, New Zealand, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, Venezuela, Vietnam.
442,475 total for all drape models