FDA Enforcement Class II Ongoing

Microstream Instructions for Use and Part Number used with - Product Description, REF: Microstream Luer Intubated Adult Pediatric, PT00156355 - Microstream Luer Adult-Pediatric Intubated CO2 Sampling Line 2m/6.5ft Short-term use: Procedural/Emergency, MLAI

Recall: Z-1383-2025 · Reported March 26, 2025

Enforcement

Recall Number
Z-1383-2025
Event ID
96309
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Oridion Medical 1987 Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 26, 2025
Initiation Date
February 17, 2025
Classification Date
March 17, 2025
Address
Har Hotzvim Industrial Park Box 45025, Har Hotzvim Industrial Park, Jerusalem, N/A, N/A, Israel

Description

Microstream Instructions for Use and Part Number used with - Product Description, REF: Microstream Luer Intubated Adult Pediatric, PT00156355 - Microstream Luer Adult-Pediatric Intubated CO2 Sampling Line 2m/6.5ft Short-term use: Procedural/Emergency, MLAI

Reason

Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.

Code Info

Part Number and Revision - REF/UDI-DI(Lot): PT00156355 Rev A- MLAI/10884521762145(C210103390, C210103391, C210103392, C220100198, C220101566, C220210015, C220212812, C220430762, C220650161, C220975826, C221087553, C230208559, C230627706, C230733095, C240624686, D201159627, D201262180, D210206012)

Distribution

Worldwide - US Nationwide distribution including in the states of TX, OH, FL, CO, IL, CA, MD, PA, VA, WI, IN, MI, NY, GA, OR, SC, NC, PR, AZ, KY, AR, TN, MO, DE, MA, MS, UT, DC, NE, IA, WV, NJ, KS, LA, WA, HI, MN, AL, ME, RI, OK, MT, ID, CT, NH, AK, NV, SD, WY, ND, NM and the countries of Algeria, Andorra, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, Northern Ireland, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Virgin Islands, U.S.

Quantity

62877