FDA Enforcement Class II Ongoing

Tibial Component Modular. orthopedic prosthesis. Model Nos: Small, W 16-2817/32 Medium, W 16-2817/35 Large, W 16-2817/37

Recall: Z-1380-2022 · Reported July 20, 2022

Enforcement

Recall Number
Z-1380-2022
Event ID
90477
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
July 20, 2022
Initiation Date
May 24, 2022
Classification Date
July 12, 2022
Address
Oststr. 4-10, Norderstedt, N/A, Germany

Description

Tibial Component Modular. orthopedic prosthesis. Model Nos: Small, W 16-2817/32 Medium, W 16-2817/35 Large, W 16-2817/37

Reason

There is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.

Code Info

UDI-DI: 04026575370870 Small, W; 04026575370887 Medium, W; 04026575370894 Large, W; Serial/Lot Numbers: 210308/0020 210308/0035 210222/0789 151116/4741

Distribution

Domestic distribution to AL, CA, FL, GA, IL, IN, KS, MI, MN, NV, NY, OH, TN, TX, WA. Foreign distribution to Algeria Argentina Australia Austria Belgium Bulgaria Canada China Colombia Denmark Ecuador Estonia Finland France Germany Great Britain Greece Hungary India Indonesia Israel Italy Latvia Libya Lithuania Mexico Netherlands Norway Pakistan Peru Philippines Poland Romania Saudi Arabia Slovakia Slovenia Spain Sudan Sweden Switzerland Thailand Turkey Ukraine United Arab Emirates Uruguy Vietnam

Quantity

3785