EPIQ DIAGNOSTIC ULTRASOUND SYSTEM, Model EPIQ 5C, EPIC 5G, EPIQ 5W, EPIQ 7C, EPIC 7GC, and EPIQ 7W. Diagnostic Ultrasound System for ultrasound imaging in abdominal, cardiac adult, cardiac other (fetal), cardiac pediatric, cerebral vascular, cephalic (adult), cephalic (neonatal), fetal/obstetric, gynecological, intraoperative (vascular), intraoperative (cardiac), musculoskeletal (conventional), musculoskeletal (superficial), other: urology, pediatric, peripheral vessel, small organ (breast, thyroid, testicle), transesophageal (cardiac), trans rectal, transvaginal.
Enforcement
- Recall Number
- Z-1380-2016
- Event ID
- 73665
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Philips Ultrasound, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 20, 2016
- Initiation Date
- March 28, 2016
- Classification Date
- April 13, 2016
- Termination Date
- December 13, 2017
- Address
- 22100 Bothell Everett Hwy, N/A, Bothell, WA, 98021-8431, United States
Description
EPIQ DIAGNOSTIC ULTRASOUND SYSTEM, Model EPIQ 5C, EPIC 5G, EPIQ 5W, EPIQ 7C, EPIC 7GC, and EPIQ 7W. Diagnostic Ultrasound System for ultrasound imaging in abdominal, cardiac adult, cardiac other (fetal), cardiac pediatric, cerebral vascular, cephalic (adult), cephalic (neonatal), fetal/obstetric, gynecological, intraoperative (vascular), intraoperative (cardiac), musculoskeletal (conventional), musculoskeletal (superficial), other: urology, pediatric, peripheral vessel, small organ (breast, thyroid, testicle), transesophageal (cardiac), trans rectal, transvaginal.
The fasteners securing the control panel assembly to the base of the Philips EPIQ Ultrasound System may loosen over time, which could subsequently lead to the detachment of the entire assembly from the ultrasound system.
All Serial numbers
Worldwide Distribution: US (nationwide) including Washington, D.C., and countries of: Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bermuda, Bolivia, Brazil, Brunei Darussalam, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Malta, Martinique, Mayotte, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, State of, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿¿union, Romania, Russia , Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, and Viet Nam.
11,085 units total (4909 units in the US and 6176 units outside the US)