Tibial Component. orthopedic prosthesis. Model Nos: Small, W 16-2817/02 Medium, W 16-2817/05 Large, W 16-2817/07
Enforcement
- Recall Number
- Z-1379-2022
- Event ID
- 90477
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- July 20, 2022
- Initiation Date
- May 24, 2022
- Classification Date
- July 12, 2022
- Address
- Oststr. 4-10, Norderstedt, N/A, Germany
Description
Tibial Component. orthopedic prosthesis. Model Nos: Small, W 16-2817/02 Medium, W 16-2817/05 Large, W 16-2817/07
There is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.
UDI-DI: 04026575359202 Small, W; 04026575359219 Medium, W; 04026575359226 Large, W; Serial/Lot Numbers: 201002/1683 211025/1543 210202/0275 210302/0308 210314/1872 201002/1687
Domestic distribution to AL, CA, FL, GA, IL, IN, KS, MI, MN, NV, NY, OH, TN, TX, WA. Foreign distribution to Algeria Argentina Australia Austria Belgium Bulgaria Canada China Colombia Denmark Ecuador Estonia Finland France Germany Great Britain Greece Hungary India Indonesia Israel Italy Latvia Libya Lithuania Mexico Netherlands Norway Pakistan Peru Philippines Poland Romania Saudi Arabia Slovakia Slovenia Spain Sudan Sweden Switzerland Thailand Turkey Ukraine United Arab Emirates Uruguy Vietnam
3785