FDA Enforcement Class II Terminated

Trial Cutting Guide for Triathlon Total Knee System, Triathlon PKR System, Restoris MCK Knee System, prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer. Catalog Number: I-K3834CG00 Specialty Triathlon TS Trial Cutting Guide Set I-K3834CG1L Specialty Triathlon TS Trial Cutting Guide Size 1 Left I-K3834CG1R Specialty Triathlon TS Trial Cutting Guide Size 1 Right I-K3834CG2L Specialty Triathlon TS Trial Cutting Guide Size 2 Left I-K3834CG2R Specialty Triathlon TS Trial Cutting Guide Size 2 Right I-K3834CG3L Specialty Triathlon TS Trial Cutting Guide Size 3 Left I-K3834CG3R Specialty Triathlon TS Trial Cutting Guide Size 3 Right I-K3834CG4L Specialty Triathlon TS Trial Cutting Guide Size 4 Left I-K3834CG4R Specialty Triathlon TS Trial Cutting Guide Size 4 Right I-K3834CG5L Specialty Triathlon TS Trial Cutting Guide Size 5 Left I-K3834CG5R Specialty Triathlon TS Trial Cutting Guide Size 5 Right I-K3834CG6L Specialty Triathlon TS Trial Cutting Guide Size 6 Left I-K3834CG6R Specialty Triathlon TS Trial Cutting Guide Size 6 Right I-K3834CG7L Specialty Triathlon TS Trial Cutting Guide Size 7 Left I-K3834CG7R Specialty Triathlon TS Trial Cutting Guide Size 7 Right I-K3834CG8L Specialty Triathlon TS Trial Cutting Guide Size 8 Left I-K3834CG8R Specialty Triathlon TS Trial Cutting Guide Size 8 Right

Recall: Z-1379-2021 · Reported April 14, 2021

Enforcement

Recall Number
Z-1379-2021
Event ID
87517
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Howmedica Osteonics Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 14, 2021
Initiation Date
March 5, 2021
Classification Date
April 8, 2021
Termination Date
May 10, 2024
Address
325 Corporate Dr, N/A, Mahwah, NJ, 07430-2006, United States

Description

Trial Cutting Guide for Triathlon Total Knee System, Triathlon PKR System, Restoris MCK Knee System, prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer. Catalog Number: I-K3834CG00 Specialty Triathlon TS Trial Cutting Guide Set I-K3834CG1L Specialty Triathlon TS Trial Cutting Guide Size 1 Left I-K3834CG1R Specialty Triathlon TS Trial Cutting Guide Size 1 Right I-K3834CG2L Specialty Triathlon TS Trial Cutting Guide Size 2 Left I-K3834CG2R Specialty Triathlon TS Trial Cutting Guide Size 2 Right I-K3834CG3L Specialty Triathlon TS Trial Cutting Guide Size 3 Left I-K3834CG3R Specialty Triathlon TS Trial Cutting Guide Size 3 Right I-K3834CG4L Specialty Triathlon TS Trial Cutting Guide Size 4 Left I-K3834CG4R Specialty Triathlon TS Trial Cutting Guide Size 4 Right I-K3834CG5L Specialty Triathlon TS Trial Cutting Guide Size 5 Left I-K3834CG5R Specialty Triathlon TS Trial Cutting Guide Size 5 Right I-K3834CG6L Specialty Triathlon TS Trial Cutting Guide Size 6 Left I-K3834CG6R Specialty Triathlon TS Trial Cutting Guide Size 6 Right I-K3834CG7L Specialty Triathlon TS Trial Cutting Guide Size 7 Left I-K3834CG7R Specialty Triathlon TS Trial Cutting Guide Size 7 Right I-K3834CG8L Specialty Triathlon TS Trial Cutting Guide Size 8 Left I-K3834CG8R Specialty Triathlon TS Trial Cutting Guide Size 8 Right

Reason

The posterior condyle of the Specialty Triathlon TS Trial Cutting Guide(s), sizes 1 through 8, (Left and Right) has the potential to fracture at the augment cut slot. This can occur in the Trial Cutting Guide packaging during transport or during impaction of the instrument in surgery.

Code Info

All Lot numbers for each GTIN: GTIN: 07613327327595 GTIN: 07613327327533 GTIN: 07613327327748 GTIN: 07613327329209 GTIN: 07613327329193 GTIN: 07613327329100 GTIN: 07613327329087 GTIN: 07613327329063 GTIN: 07613327328899 GTIN: 07613327328868 GTIN: 07613327328707 GTIN: 07613327328691 GTIN: 07613327328592 GTIN: 07613327328530 GTIN: 07613327328639 GTIN: 07613327328424 GTIN: 07613327328288

Distribution

Distribution in US to: CA,, GA, NC, NJ, NV, OK, TX, and to Canada

Quantity

113 components