FDA Enforcement Class II Terminated

BD BBL Martin-Lewis Agar 20 Ea - Product Usage: use for the isolation of pathogenic Neisseria from specimens containing mixed flora of bacteria and fungi. Catalog Number: 221557

Recall: Z-1378-2021 · Reported April 14, 2021

Enforcement

Recall Number
Z-1378-2021
Event ID
87546
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Becton Dickinson & Co.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 14, 2021
Initiation Date
March 17, 2021
Classification Date
April 8, 2021
Termination Date
August 5, 2022
Address
7 Loveton Cir, N/A, Sparks, MD, 21152-9212, United States

Description

BD BBL Martin-Lewis Agar 20 Ea - Product Usage: use for the isolation of pathogenic Neisseria from specimens containing mixed flora of bacteria and fungi. Catalog Number: 221557

Reason

Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reaction is partial inhibition after 18-48 hours incubation) and reduce the identification of Neisseria contributing to a delay in time to definitive diagnosis or a delay in proper treatment

Code Info

Lot Number: 0317558, Exp. Date 2021-MAR-11 UDI: (01)30382902215579 (10)0317558(17)031121 (30)0020

Distribution

Worldwide distribution - US Nationwide distribution and the countries of Canada, Korea.

Quantity

33660 EA