FDA Enforcement Class II Terminated

MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.

Recall: Z-1378-2018 · Reported April 25, 2018

Enforcement

Recall Number
Z-1378-2018
Event ID
79711
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cook Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 25, 2018
Initiation Date
January 12, 2018
Classification Date
April 17, 2018
Termination Date
May 8, 2020
Address
750 N Daniels Way, Bloomington, IN, 47404-9120, United States

Description

MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.

Reason

The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.

Code Info

All Lots

Distribution

Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)

Quantity

4,451