FDA Enforcement
Class II
Terminated
MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000 US) G46022 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.
Recall: Z-1377-2018
·
Reported April 25, 2018
Enforcement
- Recall Number
- Z-1377-2018
- Event ID
- 79711
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Cook Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 25, 2018
- Initiation Date
- January 12, 2018
- Classification Date
- April 17, 2018
- Termination Date
- May 8, 2020
- Address
- 750 N Daniels Way, Bloomington, IN, 47404-9120, United States
Description
MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000 US) G46022 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.
Reason
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
Code Info
All lots
Distribution
Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)
Quantity
483