FDA Enforcement Class II Ongoing

MODULAR CATHCART FRACTURE HEAD HIP BALL, sizes 41mm to 60mm inclusive; metallic femoral head prosthesis, Product No. 136341000, 136342000, 136343000, 136344000, 136345000, 136346000, 136347000, 136348000, 136349000, 136350000, 136351000, 136352000, 136353000, 136354000, 136356000, 136358000, 136360000

Recall: Z-1375-2022 · Reported July 20, 2022

Enforcement

Recall Number
Z-1375-2022
Event ID
90476
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
DePuy Orthopaedics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 20, 2022
Initiation Date
May 18, 2022
Classification Date
July 12, 2022
Address
700 Orthopaedic Dr, Warsaw, IN, 46582-3994, United States

Description

MODULAR CATHCART FRACTURE HEAD HIP BALL, sizes 41mm to 60mm inclusive; metallic femoral head prosthesis, Product No. 136341000, 136342000, 136343000, 136344000, 136345000, 136346000, 136347000, 136348000, 136349000, 136350000, 136351000, 136352000, 136353000, 136354000, 136356000, 136358000, 136360000

Reason

The Cathcart Ball has an inherent +5mm offset which is not currently specified in its Instructions for Use or Surgical Technique Guide. The STG has been updated to reflect the clarification that the +5mm offset should be taken into consideration when using a Cathcart head in a construct for hemi-hip arthroplasty procedures.

Code Info

All lots until Instructions for Use (IFU) is revised

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Denmark, Finland, Germany, Greece, Guam, Hong Kong, Ireland, Netherlands, New Zealand, Norway, South Africa, Spain, Sweden, Switzerland, United Kingdom.

Quantity

3488