Aftermarket Alaris Infusion Pump Module Model 8100 Front Bezel, installed by Pacific Medical Group (DBA Avante Health Solutions)
Enforcement
- Recall Number
- Z-1375-2021
- Event ID
- 87483
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Pacific Medical Group Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 28, 2021
- Initiation Date
- March 24, 2021
- Classification Date
- April 19, 2021
- Address
- 212 Avenida Fabricante, N/A, San Clemente, CA, 92672-7538, United States
Description
Aftermarket Alaris Infusion Pump Module Model 8100 Front Bezel, installed by Pacific Medical Group (DBA Avante Health Solutions)
Aftermarket front bezel components were installed during service/repair, not by the original manufacturer, within infusion pump modules. Over time, the posts on the front bezel component may crack or separate. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.
Serial Numbers of devices serviced/purchased 07/10/2020-02/18/2021: 4105026 4122780 9925856 9961244 12333241 12333419 12333802 12334211 12334212 12334213 12334214 12334319 12334403 12334424 12334452 12334652 12334713 12334833 12334896 12334950 12334951 12334957 12335116 12335128 12335147 12335159 12335163 12335212 12335232 12335260 12335304 12335380 12335427 12335481 12335513 12335522 12335570 12335596 12335606 12335625 12335642 12335673 12335722 12335754 12335790 12335792 12335883 12335904 12335921 12335926 12335967 12335970 12336005 12336011 12336156 12336165 12336273 12336275 12336279 12336280 12336293 12336305 12336395 12336402 12336403 12336430 12336435 12336461 12336630 12336668 12336694 12336971 12337823 12421700 12517224 12524554 12524783 12544522 12568207 12570110 12570217 12571055 12580490 12586058 12586274 12586333 12587144 12588342 12590617 12590762 12591783 12591900 12592059 12592353 12592442 12592545 12592622 12592624 12592771 12592840 12592879 12592944 12592968 12592979 12593139 12593231 12594623 12595031 12595247 12595693 12595881 12595943 12595992 12596068 12596243 12596303 12597046 12597178 12597462 12597465 12597556 12597699 12598072 12598088 12598112 12598125 12640247 12640292 12641907 12650197 12655100 12740634 12743978 12744142 12755303 12767321 12779073 12793502 12793519 12793565 12793610 12815356 12822657 12822783 12823029 12823031 12823099 12823177 12827621 12828337 12875779 12876033 12877579 12897320 12961435 12993617 13022654 13057056 13066426 13066442 13066455 13066498 13066510 13066511 13066525 13066560 13066570 13066618 13066637 13066674 13066680 13066683 13066696 13066713 13066715 13066756 13066798 13066808 13066858 13066877 13066887 13066899 13066965 13067206 13067213 13067257 13067270 13067271 13067282 13067283 13067340 13067343 13067423 13067428 13067451 13067453 13067456 13067482 13067484 13067510 13067530 13067559 13067568 13067569 13067570 13067571 13067649 13067669 13067715 13067747 13067751 13067774 13067796 13067797 13067808 13067813 13067827 13067879 13067883 13067906 13067913 13067928 13067948 13067954 13067956 13067957 13067967 13068037 13068038 13068048 13068050 13068052 13068295 13068362 13068384 13068386 13070783 13108584 13166329 13211801 13212320 13315884 13421569 13489942 13499587 13530717 13530791 13557812 13558687 13559309 13569015 13571437 13585486 13586852 13587307 13588968 13589012 13592584 13592826 13592978 13593436 13595698 13595718 13596145 13596439 13596492 13596561 13596703 13596833 13596896 13597393 13597456 13597596 13597604 13597651 13597698 13597921 13597922 13598257 13599210 13599280 13599312 13599388 13599392 13599519 13599542 13599972 13600058 13600936 13601022 13601030 13601031 13601239 13604404 13604690 13605007 13605028 13605044 13605145 13605153 13605223 13605288 13605327 13605341 13605348 13605349 13605495 13605514 13605611 13605612 13605627 13605640 13605648 13605660 13605687 13605690 13605692 13605768 13605776 13605779 13605780 13605785 13605803 13605805 13605828 13605829 13605856 13605867 13605891 13605893 13605915 13605923 13605955 13606085 13606119 13606128 13606143 13606199 13606220 13606222 13606223 13606232 13606275 13606352 13606402 13606446 13606462 13606527 13606534 13606566 13606572 13606582 13606585 13606590 13606600 13606608 13606615 13606626 13606636 13606644 13606649 13606693 13606715 13606737 13606741 13606743 13606762 13606765 13606771 13606787 13606817 13606821 13606857 13606862 13606864 13606873 13606877 13606880 13606884 13606886 13606887 13606894 13606895 13606912 13606913 13606933 13606938 13606940 13606949 13606953 13606962 13606963 13606964 13606992 13606993 13607004 13607013 13607015 13607016 13607022 13607024 13607028 13607033 13607041 13607047 13607050 13607054 13607061 13607063 13607064 13607072 13607073 13607081 13607082 13607084 13607085 13607086 13607089 13607090 13607097 13607098 13607103 13607110 13607114 13607116 13607122 13607123 13607136 13607162 13607196 13607203 13607213 13607438 13607453 13607467 13607521 13607525 13607526 13607573 13607617 13607619 13607621 13607622 13607626 13607627 13607628 13607642 13607643 13607717 13607726 13610054 13610888 13610991 13612073 13613850 13613949 13614806 13615336 13616177 13616266 13616312 13617896 13618080 13618110 13618117 13618350 13618386 13618387 13618402 13618404 13618405 13618421 13618423 13618424 13618439 13618444 13618458 13618493 13618497 13618512 13618515 13618526 13618577 13618589 13618603 13618620 13618635 13618679 13618681 13618688 13618693 13618694 13618696 13618703 13618704 13618708 13618711 13618729 13618730 13618736 13618760 13618776 13618799 13618831 13618834 13618843 13618860 13618876 13618897 13618899 13618916 13618918 13618919 13618920 13618922 13618929 13618933 13618939 13618940 13618942 13618943 13618945 13618947 13618948 13618949 13618955 13618960 13618966 13618968 13618969 13618971 13618972 13618973 13618975 13618977 13618980 13618981 13618983 13618988 13618989 13618990 13618994 13618997 13618998 13619006 13619007 13619009 13619013 13619016 13619019 13619020 13619039 13619040 13619041 13619044 13619059 13619063 13619066 13619069 13619072 13619075 13619083 13619097 13619122 13619132 13619209 13619321 13619411 13619413 13619420 13619426 13619427 13619429 13619433 13619437 13619438 13619446 13619447 13619450 13619454 13619457 13619458 13619462 13619463 13619465 13619466 13619499 13619502 13619519 13619521 13619522 13619524 13619539 13619543 13619544 13619558 13619561 13619586 13619610 13619626 13619666 13619679 13619701 13619703 13619723 13619724 13619725 13619727 13619729 13619741 13619749 13619750 13619759 13619760 13619762 13619764 13619771 13619790 13619792 13619803 13619804 13619827 13619828 13619838 13619839 13619841 13619888 13619892 13619893 13619903 13619911 13619915 13619923 13619924 13619932 13619935 13619938 13619943 13619965 13620025 13620038 13620039 13620043 13620051 13620053 13620057 13620059 13620063 13620069 13620070 13620078 13620086 13620093 13620094 13620096 13620097 13620099 13620103 13620104 13620107 13620110 13620112 13620116 13620117 13620119 13620122 13620124 13620125 13620132 13620133 13620150 13620155 13620156 13620157 13620170 13620179 13620188 13620205 13620242 13620243 13620257 13620317 13620318 13620320 13620324 13620326 13620345 13620346 13620347 13620348 13620356 13620362 13620365 13620373 13620376 13620413 13620417 13620442 13620518 13620520 13620521 13620522 13620535 13620536 13620553 13620562 13620566 13620571 13620579 13620581 13620586 13620588 13620591 13620612 13620625 13620633 13620638 13620654 13620665 13620670 13620674 13620698 13620702 13620705 13620719 13620741 13620752 13620754 13620760 13620768 13620769 13620784 13620785 13620787 13620802 13620807 13620810 13620814 13620821 13620823 13620829 13620844 13620858 13620877 13620890 13620898 13620900 13620946 13620949 13620957 13620963 13620964 13620995 13620996 13621030 13621050 13621070 13621109 13621111 13621114 13621131 13621133 13621135 13621136 13621138 13621149 13621155 13621156 13621158 13621164 13621169 13621171 13621174 13621181 13621192 13621195 13621196 13621198 13621204 13621206 13621224 13621226 13621229 13621230 13621231 13621233 13621238 13621244 13621285 13621287 13621301 13621304 13621310 13621313 13621316 13621318 13621320 13621346 13621349 13621352 13621368 13621372 13621377 13621381 13621458 13621480 13621492 13621563 13621576 13621587 13621593 13621610 13621616 13621663 13621703 13621730 13621786 13621823 13621845 13622002 13622022 13622031 13623101 13626135 13629763 13629965 13632088 13632093 13632948 13633449 13634333 13634968 13635259 13635520 13635548 13636223 13636319 13636606 13636668 13636819 13636843 13637450 13637710 13637733 13637823 13638090 13638091 13638345 13638451 13638808 13638896 13639105 13639315 13639684 13639787 13639832 13639889 13639895 13639965 13639980 13639981 13639999 13640034 13640064 13640069 13640097 13640123 13640126 13640128 13640189 13640223 13640234 13640327 13640454 13640651 13642377 13642387 13642393 13642410 13642435 13642442 13642447 13642673 13642992 13643019 13643225 13643463 13643473 13643474 13643479 13643486 13643489 13643490 13643491 13643499 13643500 13643501 13643507 13643508 13643509 13643520 13643528 13643531 13643536 13643544 13643552 13643583 13643586 13643594 13643610 13643611 13643616 13643637 13643763 13643768 13643783 13643836 13643843 13643860 13643861 13643875 13643878 13643897 13643899 13643936 13643937 13643938 13643940 13643952 13643953 13643964 13643965 13643966 13643968 13643969 13643970 13643987 13643988 13643990 13643998 13644018 13644030 13644031 13644044 13644045 13644110 13644119 13644138 13644145 13644146 13644152 13644153 13644166 13644191 13644196 13644198 13644217 13644222 13644224 13644237 13644238 13644244 13644247 13644248 13644251 13644252 13644253 13644254 13644262 13644265 13644275 13644285 13644294 13644296 13644297 13644298 13644299 13644301 13644327 13644331 13644465 13644479 13644487 13644496 13644597 13644600 13644930 13645074 13645110 13645125 13645126 13645128 13645146 13645154 13645155 13645158 13645163 13645174 13645187 13645189 13645197 13645200 13645204 13645230 13645232 13645233 13645252 13645273 13645316 13645323 13645324 13645329 13645331 13645333 13645334 13645340 13645344 13645351 13645355 13645358 13645370 13645372 13645386 13645402 13645404 13645410 13645411 13645414 13645429 13645432 13645433 13645436 13645438 13645439 13645445 13645448 13645449 13645452 13645453 13645454 13645464 13645466 13645469 13645479 13645480 13645481 13645484 13645485 13645487 13645488 13645489 13645491 13645496 13645500 13645501 13645503 13645511 13645520 13645522 13645523 13645524 13645528 13645535 13645536 13645537 13645541 13645543 13645549 13645550 13645551 13645553 13645555 13645560 13645561 13645567 13645570 13645578 13645584 13645585 13645586 13645587 13645589 13645591 13645593 13645596 13645606 13645607 13645609 13645612 13645615 13645617 13645618 13645620 13645622 13645624 13645625 13645629 13645637 13645638 13645641 13645642 13645645 13645646 13645648 13645649 13645650 13645651 13645654 13645661 13645662 13645663 13645666 13645672 13645675 13645679 13645685 13645690 13645691 13645704 13645709 13645710 13645717 13645724 13645736 13645785 13645786 13645805 13645898 13645929 13645930 13645943 13645968 13645974 13646010 13646013 13646019 13646056 13646065 13646068 13646081 13646190 13646215 13646234 13646337 13646341 13646352 13646353 13646381 13646388 13646389 13646411 13646456 13646497 13646589 13646601 13646610 13646655 13646984 13647053 13647134 13647147 13647173 13647178 13647436 13647526 13647539 13647553 13647649 13647799 13647810 13647971 13647981 13647992 13648019 13648053 13648069 13648082 13648090 13648133 13648143 13648204 13648205 13648207 13648228 13648273 13648277 13648284 13648287 13648319 13648373 13648689 13648765 13648778 13648799 13648800 13648833 13648837 13648848 13648864 13648875 13648884 13648886 13648922 13648929 13649007 13649038 13649052 13649056 13649135 13649159 13649167 13649176 13649199 13649204 13649205 13649210 13649211 13649218 13649222 13649225 13649239 13649241 13649244 13649246 13649251 13649253 13649254 13649255 13649256 13649260 13649262 13649267 13649268 13649275 13649278 13649282 13649283 13649284 13649291 13649292 13649298 13649303 13649305 13649306 13649313 13649318 13649320 13649325 13649327 13649329 13649330 13649332 13649335 13649336 13649338 13649347 13649348 13649356 13649360 13649364 13649368 13649373 13649377 13649398 13649401 13649411 13649413 13649416 13649417 13649420 13649422 13649423 13649424 13649425 13649427 13649428 13649429 13649437 13649438 13649439 13649442 13649444 13649445 13649447 13649448 13649453 13649454 13649455 13649459 13649463 13649465 13649468 13649469 13649473 13649476 13649477 13649478 13649482 13649483 13649486 13649488 13649489 13649490 13649492 13649497 13649499 13649500 13649502 13649506 13649510 13649513 13649519 13649528 13649554 13649580 13649582 13649586 13649587 13649618 13649619 13649786 13649961 13649992 13650003 13650064 13650065 13650067 13650090 13650121 13650124 13650230 13650558 13650755 13651068 13651102 13651686 13651832 13651991 13651999 13652011 13652545 13652576 13653241 13653594 13653607 13653611 13653624 13653710 13653810 13653847 13653851 13653861 13653950 13653961 13653964 13653968 13653976 13653990 13653997 13654204 13654243 13654264 13654278 13654280 13654283 13654290 13654291 13654297 13654340 13654347 13654362 13654364 13654396 13654402 13654403 13654416 13654418 13654419 13654421 13654448 13654449 13654459 13654461 13654464 13654473 13654524 13654541 13654551 13654552 13654582 13654598 13654607 13654612 13654633 13654647 13654648 13654715 13654776 13654787 13654793 13654795 13654796 13654806 13654810 13654814 13654818 13654830 13654845 13654848 13654849 13654855 13654859 13654860 13654861 13654886 13654895 13654897 13654898 13654899 13654901 13654906 13654907 13654908 13654910 13654915 13654940 13654946 13654955 13654956 13654957 13654959 13654961 13654964 13654966 13654971 13654979 13654992 13654996 13655008 13655015 13655022 13655023 13655024 13655027 13655029 13655031 13655033 13655036 13655037 13655038 13655040 13655041 13655042 13655043 13655048 13655055 13655057 13655058 13655063 13655089 13655094 13655099 13655101 13655102 13655107 13655111 13655112 13655116 13655128 13655129 13655131 13655137 13655138 13655144 13655146 13655147 13655179 13655185 13655190 13655191 13655195 13655196 13655200 13655205 13655207 13655208 13655210 13655212 13655214 13655215 13655216 13655218 13655219 13655220 13655229 13655230 13655236 13655238 13655240 13655247 13655251 13655252 13655255 13655256 13655259 13655260 13655263 13655264 13655265 13655267 13655274 13655277 13655278 13655285 13655293 13655295 13655296 13655297 13655299 13655300 13655304 13655305 13655306 13655308 13655309 13655319 13655325 13655329 13655330 13655332 13655334 13655336 13655337 13655339 13655340 13655341 13655342 13655344 13655345 13655346 13655349 13655350 13655352 13655353 13655358 13655361 13655365 13655366 13655367 13655368 13655370 13655376 13655378 13655379 13655383 13655385 13655386 13655388 13655390 13655391 13655392 13655394 13655396 13655429 13656085 13659411 13660743 13665243 13733358 13746337 13788287 13788833 13808047 13845635 13845726 13872906 13878075 13878653 13879137 13879149 13879198 13879211 13879241 13879272 13879333 13879539 13879749 13879958 13879973 13879976 13879977 13880034 13880039 13880047 13880158 13880182 13880183 13880185 13880190 13880193 13880199 13880208 13880236 13880248 13880251 13880307 13880462 13880490 13880608 13880622 13880623 13880672 13880684 13880711 13880712 13880713 13880720 13880731 13880737 13880744 13880795 13880808 13880810 13880811 13880812 13880813 13880814 13880815 13880816 13880818 13880824 13880825 13880898 13880900 13880902 13880908 13880945 13880947 13880956 13880958 13880961 13880968 13880969 13880990 13880991 13880995 13881000 13881001 13881014 13881051 13881066 13881076 13881080 13881153 13881156 13881157 13881171 13881183 13881185 13881192 13881195 13881204 13881205 13881223 13881239 13881240 13881241 13881242 13881254 13881261 13881263 13881287 13881290 13881298 13881311 13881314 13881315 13881316 13881329 13881334 13881336 13881337 13881344 13881345 13881346 13881368 13881373 13881374 13881375 13881376 13881377 13881384 13881386 13881389 13881390 13881391 13881397 13881399 13881401 13881414 13881417 13881422 13881425 13881427 13881435 13881439 13881445 13881502 13881504 13881505 13881506 13881526 13881528 13881529 13881575 13881578 13881581 13881583 13881584 13881585 13881586 13881609 13881611 13881615 13881616 13881620 13881627 13881629 13881630 13881637 13881640 13881642 13881653 13881656 13881658 13881659 13881661 13881685 13881686 13881692 13881693 13881694 13881695 13881696 13881708 13881710 13881711 13881714 13881716 13881729 13881760 13881761 13881762 13881769 13881772 13881773 13881775 13881777 13881781 13881784 13881793 13881794 13881839 13881840 13881842 13881846 13881858 13881860 13881863 13881865 13881866 13881868 13881869 13881875 13881876 13881878 13881881 13881883 13881889 13881891 13881892 13881909 13881910 13881911 13881912 13881913 13881914 13881922 13881923 13881925 13881949 13881950 13881953 13881964 13881965 13881966 13881980 13881983 13881984 13881988 13881990 13881992 13882001 13882003 13882004 13882006 13882007 13882019 13882021 13882022 13882036 13882037 13882039 13882040 13882041 13882042 13882043 13882057 13882059 13882060 13882063 13882074 13882075 13882077 13882079 13882080 13882082 13882090 13882092 13882093 13884387 13884429 13884430 13884447 13884753 13884770 13884783 13884870 13884894 13885080 13885135 13887449 13890945 13986573 14054503 14067692 14067826 14068954 14069590 14069593 14069679 14069943 14070424 14070425 14070430 14070468 14070501 14070990 14071002 14071035 14071121 14071161 14071230 14071389 14071396 14071420 14071489 14071497 14071519 14071569 14071570 14071602 14071685 14071746 14071749 14071768 14071781 14071808 14071813 14071822 14071961 14071979 14072000 14072046 14072325 14072396 14072401 14072441 14072488 14072647 14072655 14073516 14086875 14174858 14174873 14175004 14175016 14175026 14175042 14176226 14176305 14176307 14176320 14176362 14176622 14176624 14343279 14345979 14347710 14350979 14412449 14412585 14412590 14412737 14412895 14412948 14412960 14440259 14440262 14440263 14440266 14450726 14452000 14452232 14452395 14452800 14454424 14455305 14455508 14455950 14455981 14456021 14456028 14456032 14456042 14456048 14456083 14456085 14456087 14456112 14456113 14456115 14456117 14456120 14456124 14456125 14456127 14456138 14456144 14456156 14456157 14456170 14456194 14456198 14456203 14456207 14456208 14456209 14456214 14456217 14456225 14456227 14456228 14456232 14456234 14456236 14456279 14456283 14456307 14456315 14456319 14456365 14456370 14456375 14456417 14456421 14456423 14456425 14456433 14456434 14456442 14456447 14456452 14456454 14456466 14456470 14456471 14456476 14456483 14456485 14456488 14456490 14456492 14456499 14456501 14456508 14456517 14456527 14456530 14456532 14456534 14456540 14456543 14456546 14456547 14456562 14456563 14456564 14456566 14456569 14456570 14456572 14456574 14456576 14456583 14456590 14456591 14456599 14456613 14456614 14456617 14456625 14456628 14456644 14456652 14456655 14456660 14456665 14456669 14456673 14456686 14456687 14456691 14456693 14456696 14456699 14456720 14456730 14456756 14456763 14456769 14456779 14456784 14456785 14456786 14456790 14456792 14456793 14456797 14456802 14456804 14456805 14456807 14456832 14456836 14456841 14456843 14456854 14456858 14456881 14456891 14456892 14456893 14456896 14456902 14456903 14456961 14456962 14456974 14456979 14457060 14457063 14457170 14457181 14490547 14556891 14563443 14892974 14998379 15256256 15353003 15353574 15772092 15772541 15772665 15773405 15773569 15773737 15773964 15773984 15774024 15774037 15774082 15774265 15774287 15774292 65188509 136128906 136550641 14456571 1344180 1345078 1489010 2000153 2000856 4015036 4033181 4108676 4124626 4124892 4124987 4125058 4125131 9855973 9858058 9864621 9865213 9866687 9866688 9866816 9871193 9924841 12334212 12334713 12334896 12335212 12335304 12335596 12335708 12336395 12336668 12385657 12385967 12386195 12386200 12386214 12386292 12386697 12386701 12386729 12387053 12387054 12387058 12387089 12387130 12387196 12387227 12387482 12388237 12388565 12388863 12389239 12389294 12539993 12541261 12546951 12548420 12549062 12549080 12549095 12549778 12550196 12550743 12550773 12550918 12550964 12551584 12551739 12553107 12605247 12606928 12660756 12670159 12685932 12689307 12745045 12776625 12780687 12780728 12780773 12780779 12780791 12780889 12796216 12803071 12803102 12803110 12803113 12803316 12803627 12805140 12828292 12838128 12876033 12926449 12945095 13032146 13034250 13046796 13063625 13084497 13094801 13094995 13095132 13095195 13095238 13095835 13095932 13096042 13096141 13133106 13165199 13174991 13175209 13198809 13198814 13202502 13202642 13318260 13318363 13318366 13318482 13318556 13330805 13359904 13363948 13365137 13365178 13427021 13427072 13427209 13489546 13496266 13496859 13497767 13512978 13521109 13521678 13521697 13522479 13524005 13524798 13524913 13531252 13531491 13536612 13537919 13538037 13541451 13541563 13541723 13541834 13541973 13542342 13542357 13542438 13542635 13542810 13542818 13543010 13543018 13543035 13543078 13543080 13543819 13546875 13547381 13547566 13548020 13548093 13563382 13563454 13563625 13563740 13564140 13564204 13564232 13578173 13586025 13597602 13620957 13637300 13638534 13645609 13650558 13722500 13723007 13729882 13737440 13737531 13758414 13759282 13776072 13789278 13790518 13790685 13795844 13810134 13810160 13810626 13811935 13817076 13817564 13826694 13826755 13826757 13827054 13827111 13827293 13827365 13827379 13827505 13827552 13827554 13827662 13845761 13849399 13849937 13887260 13888010 13889694 13890305 13890480 13894407 13894459 13914962 13916360 13918412 13982181 13982263 13982553 13982596 13982647 14045740 14075807 14135841 14174749 14219444 14230172 14338921 14342642 14343836 14344820 14344841 14344885 14344916 14345391 14345625 14374735 14375504 14394316 14399001 14399586 14399630 14399635 14399687 14399702 14399903 14404453 14416718 14418610 14420132 14420735 14429395 14440521 14458327 14458436 14458506 14458846 14458876 14458883 14458938 14458943 14459048 14459084 14463486 14491397 14496475 14503360 14512903 14549416 14550181 14550269 14550335 14550574 14550806 14550853 14550970 14551188 14551193 14559593 14566192 14591575 14591715 14662804 14676210 14679706 14680992 14681018 14681019 14681024 14681158 14681189 14681624 14681696 14681729 14681822 14715546 14716510 14716774 14716972 14723525 14776780 14777067 14777344 14777793 14777920 14784096 14784160 14792838 14794194 14794676 14794826 14794857 14794983 14795014 14795133 14810137 14813577 14813586 14813717 14813812 14813815 14814515 14875505 14987133 14987427 15061985 15061988 15147713 15157129 15158015 15172499 15209394 15266838 15416783 15417698 15419034 15421615 15536898 113827537 13827537
Distribution US nationwide.
2452