FDA Enforcement
Class II
Terminated
CHOL+HDL+GLU test strips, REF 2412
Recall: Z-1373-2019
·
Reported May 29, 2019
Enforcement
- Recall Number
- Z-1373-2019
- Event ID
- 82331
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Polymer Technology Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 29, 2019
- Initiation Date
- January 16, 2019
- Classification Date
- May 23, 2019
- Termination Date
- June 24, 2020
- Address
- 7736 Zionsville Rd, N/A, Indianapolis, IN, 46268-2175, United States
Description
CHOL+HDL+GLU test strips, REF 2412
Reason
Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed
Code Info
I808 I809 I811 I812 I814
Distribution
Domestic distribution nationwide.
Quantity
67643 total