FDA Enforcement
Class II
Terminated
Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412
Recall: Z-1372-2019
·
Reported May 29, 2019
Enforcement
- Recall Number
- Z-1372-2019
- Event ID
- 82331
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Polymer Technology Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 29, 2019
- Initiation Date
- January 16, 2019
- Classification Date
- May 23, 2019
- Termination Date
- June 24, 2020
- Address
- 7736 Zionsville Rd, N/A, Indianapolis, IN, 46268-2175, United States
Description
Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412
Reason
Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being di P832 Chemistry (Diagnostic) CHH 2 Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412
Code Info
Q802 Q811 Q813 Q816 Q820 Q823 Q831 Q903
Distribution
Domestic distribution nationwide.
Quantity
67643 total