FDA Enforcement Class II Terminated

Lipid Panel test strips, REF 1710, also private labeled under Henry Schein as REF 570-0414

Recall: Z-1371-2019 · Reported May 29, 2019

Enforcement

Recall Number
Z-1371-2019
Event ID
82331
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Polymer Technology Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 29, 2019
Initiation Date
January 16, 2019
Classification Date
May 23, 2019
Termination Date
June 24, 2020
Address
7736 Zionsville Rd, N/A, Indianapolis, IN, 46268-2175, United States

Description

Lipid Panel test strips, REF 1710, also private labeled under Henry Schein as REF 570-0414

Reason

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed.

Code Info

P832

Distribution

Domestic distribution nationwide.

Quantity

67643 total