FDA Enforcement
Class II
Ongoing
Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.
Recall: Z-1370-2026
·
Reported February 25, 2026
Enforcement
- Recall Number
- Z-1370-2026
- Event ID
- 98315
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medica Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 25, 2026
- Initiation Date
- December 31, 2025
- Classification Date
- February 13, 2026
- Address
- 5 Oak Park Dr, Bedford, MA, 01730-1430, United States
Description
Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.
Reason
A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.
Code Info
Model No. 7303, UDI-DI: 00840095607575, Lot No. 25XXX.
Distribution
International distribution to the countries of Russia, Bulgaria, Kazakhstan, Ukraine, Pakistan, Belarus.
Quantity
1,578 units