FDA Enforcement Class II Ongoing

SYRINGE,BULB,2OZ/60CC,GREEN, Part: 300055070, packaged inside Alcon Custom-Pak Surgical Procedure Packs, Products: 18018-06, 3865-44, 5675-76, 13554-11, 18018-06, 15190-16 , 15190-16, 15190-16, 15190-16, 15190-16, 13554-11, 5918-39, 15190-16, 15190-16, 9208-66, 9208-66, 5675-76, 5918-39, 9208-66, 15190-16, 9065-25, 5918-39, 16066-15, 16066-15

Recall: Z-1370-2024 · Reported April 3, 2024

Enforcement

Recall Number
Z-1370-2024
Event ID
94094
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Alcon Research, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 3, 2024
Initiation Date
February 21, 2024
Classification Date
March 22, 2024
Address
9965 Buffalo Speedway, N/A, Houston, TX, 77054-1309, United States

Description

SYRINGE,BULB,2OZ/60CC,GREEN, Part: 300055070, packaged inside Alcon Custom-Pak Surgical Procedure Packs, Products: 18018-06, 3865-44, 5675-76, 13554-11, 18018-06, 15190-16 , 15190-16, 15190-16, 15190-16, 15190-16, 13554-11, 5918-39, 15190-16, 15190-16, 9208-66, 9208-66, 5675-76, 5918-39, 9208-66, 15190-16, 9065-25, 5918-39, 16066-15, 16066-15

Reason

Medline Bulb irrigation syringes recalled due to potential for packaging sterile barrier to be breached, and these syringes were packed inside surgical procedure packs that were subsequently terminally sterilized, so after the packs are opened, syringes should be discarded at surgical case end, and not reshelved, because the use of non-sterile bulb irrigation syringes may increase infection risk.

Code Info

Product/Lot/Expiration/UDI-DI: 18018-06/15T707/02/28/2025/380655148440, 3865-44/15T2RC/02/28/2025/380655148150, 5675-76/15T70E/02/28/2025/380655148174, 13554-11/15V4RC/02/28/2025/380655143711, 18018-06/15V3E0/02/28/2025/380655148440, 15190-16/JZ5179517/06/30/2024/380655148303, 15190-16/JZ5179519/06/30/2024/380655148303, 15190-16/JZ5179521/06/30/2024/380655148303, 15190-16/JZ5179523/06/30/2024/380655148303, 15190-16/JZ5179529/06/30/2024/380655148303, 13554-11/15TAUD/01/31/2026/380655143711, 5918-39/15TE5F/02/28/2025/380655149140, 15190-16/15THVA/04/30/2025/380655148303, 15190-16/15TNXJ/04/30/2025/380655148303, 9208-66/15V8CU/04/30/2025/380651693326, 9208-66/15XVDD/05/31/2025/380651693326, 5675-76/JZ5180811/06/30/2024/380655148174, 5918-39/JZ5180812/06/30/2024/380655149140, 9208-66/JZ5180332/03/31/2025/380651693326, 15190-16/JZ5180856/05/31/2025/380655148303, 9065-25/JZ5187593/11/30/2025/380651682849, 5918-39/JZ5187260/06/30/2024/380655149140, 16066-15/JZ5190913/07/31/2025/380655149683, 16066-15/JZ5191891/07/31/2025/380655149683 Bulb Syringe lots: 96922110001 or 96923100001

Distribution

US Nationwide distribution in the states of OH, WV.

Quantity

886