FDA Enforcement Class II Terminated

BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.

Recall: Z-1370-2015 · Reported April 8, 2015

Enforcement

Recall Number
Z-1370-2015
Event ID
70678
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
BD Biosciences, Systems & Reagents
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 8, 2015
Initiation Date
March 6, 2015
Classification Date
April 2, 2015
Termination Date
April 4, 2016
Address
2350 Qume Dr, N/A, San Jose, CA, 95131-1812, United States

Description

BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.

Reason

The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.

Code Info

Serial Nos. E34297300656, E34297300657, E34297300659, E34297300660, E34297300661, E34297300666, E34297300667, E34297300668, E34297300669, E34297300670, E34297300671, E34297300673, E34297300675, E34297300676, E34297300677, E34297300678 E34297502563, E34297502564, E34297502565, E34297502566, E34297502567, E34297502568, E34297502569, E34297502570, E34297502571, E34297502572, E34297502573, E34297502574, E34297502575, E34297502576, E34297502577, E34297502578, E34297502579, E34297502580, E34297502581, E34297502583, E34297502584, E34297502585, E34297502586, E34297502587, E34297502588, E34297502589, E34297502590, E34297502591, E34297502592, E34297502593, E34297502594, E34297502595, E34297502596, E34297502597, E34297502599, E34297502600, E34297502601, E34297502602, E34297502603, E34297502604, E34297502605, E34297502606, E34297502607, E34297502608, E34297502612, E34297502613, E34297502619, E34297502620, E34297502621, E34297502629, E34297600499, E34297600501, E34297600502, E34297600503, E34297600507, E34297600508.

Distribution

Worldwide Distribution -US including TX, NJ, and OK and Internationally to Hong Kong, Japan, Singopore, and Belguim

Quantity

72 devices