FDA Enforcement Class II Ongoing

Titan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis

Recall: Z-1369-2023 · Reported April 19, 2023

Enforcement

Recall Number
Z-1369-2023
Event ID
91800
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Coloplast Manufacturing US, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 19, 2023
Initiation Date
March 1, 2023
Classification Date
April 12, 2023
Address
1601 W River Rd, N/A, Minneapolis, MN, 55411-3431, United States

Description

Titan Touch Scro Zero Ang 16cm, Catalog Number ES29162400; inflatable penile Prosthesis

Reason

A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.

Code Info

UDI/DI 05708932539173, Lot Numbers: 8840677

Distribution

Worldwide distribution - US Nationwide and the countries of Turkey, Italy, Germany, Belgium, France, United Kingdom, Spain, Portugal.

Quantity

1 unit