FDA Enforcement Class II Ongoing

Panther Fusion Tube Tray, REF PRD-04000, PN: FA-15004, Cont. 18 Tube Trays - Product Usage: an ancillary consumable to the Panther Fusion System that is used to contain the PCR and RT-PCR steps of the assay processing.

Recall: Z-1369-2021 · Reported April 14, 2021

Enforcement

Recall Number
Z-1369-2021
Event ID
87557
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Hologic, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 14, 2021
Initiation Date
February 26, 2021
Classification Date
April 7, 2021
Address
10210 Genetic Center Dr, N/A, San Diego, CA, 92121-4362, United States

Description

Panther Fusion Tube Tray, REF PRD-04000, PN: FA-15004, Cont. 18 Tube Trays - Product Usage: an ancillary consumable to the Panther Fusion System that is used to contain the PCR and RT-PCR steps of the assay processing.

Reason

There is a potential for tube trays to leak which could potentially result in false positive results, false negative results or invalid patient results.

Code Info

Part No. (PN) FAB-15004 (Box Lot Number)/Catalog No PRD-0400 (Box/Tray Lot Number) 285762 / 647288, 285765 / 647288, 286025 / 648958, 286026 / 648958, 286028 / 648959, 286852 / 650655, 287410 / 650656, 289297 / 648959, 290057 / 647288, 290068/654833, 290409 / 654833, 291425 / 657545

Distribution

Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CT, DC, FL, GA, HI, IL, IN, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NY, OH, OR, PA, TX, UT, VA, VT, WA, WI, WV and the countries of Austria (AT), Belgium (BE), France (FR), Germany (DE), Great Britain (GB), Italy (IT) Netherlands (NL), Spain (ES), Switzerland (CH),, Canada (CA), Hong Kong (HK), Australia (AU), New Zealand (NZ).

Quantity

1355 kits