Cobas¿ 8000 modular analyzer series software version 03-01, 02, 03 and 04, impacting the cobas e 602 module only. Product Usage: Fully automated immunoassay analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids. This analyzer is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes and photometric analysis modules
Enforcement
- Recall Number
- Z-1368-2013
- Event ID
- 64850
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Diagnostics Operations, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- N/A
- Report Date
- May 29, 2013
- Initiation Date
- April 4, 2013
- Classification Date
- May 23, 2013
- Termination Date
- February 24, 2015
- Address
- 9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States
Description
Cobas¿ 8000 modular analyzer series software version 03-01, 02, 03 and 04, impacting the cobas e 602 module only. Product Usage: Fully automated immunoassay analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids. This analyzer is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes and photometric analysis modules
When using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead of Diluent Universal. This may cause incorrect results under specific circumstances if all of the following conditions are met simultaneously: More than a total of 2000 reagent kits or diluents have been registered on the cobas e 602 modules within one system core, and two different diluent
Part Number 05990378001 Serial Numbers 1017-06, 1021-03, 1021-05, 1142-09, 1145-10, 1140-01, 1125-07, 1128-02, 1146-01, 1146-02, 1146-03, 1171-10, 1169-06, 1169-08, 1169-09, 1172-09, 1156-01, 1156-02, 1156-09, 1160-04, 1160-06, 1016-05, 1017-01, 1139-08, 1139-10, 1140-07, 1140-08, 1140-10, 1156-06, 1156-07, 1158-09, 1160-01, 1160-05, 1173-01, 1139-06, 1139-07, 2374-11, 1127-08, 1138-08, 1138-10, 1139-01, 1156-08, 1156-10, 1173-10, 1156-05, 1158-10, 1139-09, 1131-08, 1132-01, 1132-02, 1171-07, 1171-08, 1171-09, 1147-05, 1149-01, 1124-05, 1124-06, 1132-03, 1132-04, 1147-05, 1149-01, 1128-05, 1128-06, 1139-02, 1139-03, 1139-04, 1139-05, 1146-05, 1146-06, 1147-04, 1156-04, 1132-05, 1132-10, 1137-02, 1147-05, 1149-01, 1128-05, 1128-06, 1149-07, 1149-08, and 1167-02.
USA Nationwide Distribution
72 units