FDA Enforcement Class II Ongoing

Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS¿ ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS¿ 5600/XT 7600 Integrated Systems. Product Code: 6199960

Recall: Z-1367-2022 · Reported July 20, 2022

Enforcement

Recall Number
Z-1367-2022
Event ID
90408
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ortho-Clinical Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 20, 2022
Initiation Date
May 26, 2022
Classification Date
July 12, 2022
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101, United States

Description

Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS¿ ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS¿ 5600/XT 7600 Integrated Systems. Product Code: 6199960

Reason

On-Analyzer Stability (OAS) time, supported for 6 weeks; not 8 weeks as is currently stated in the Instructions For Use (IFU).

Code Info

UDI-DI :10758750034413. All expired, in-date and future released lots

Distribution

Worldwide distribution - US Nationwide and the countries of Colombia, India, Singapore.

Quantity

20858