FDA Enforcement
Class II
Ongoing
Wright EVOLVE TRIAD Plate Cutter, REF 49510120
Recall: Z-1366-2022
·
Reported July 20, 2022
Enforcement
- Recall Number
- Z-1366-2022
- Event ID
- 90399
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Wright Medical Technology, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 20, 2022
- Initiation Date
- June 14, 2022
- Classification Date
- July 11, 2022
- Address
- 1023 Cherry Rd, Memphis, TN, 38117-5423, United States
Description
Wright EVOLVE TRIAD Plate Cutter, REF 49510120
Reason
The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use
Code Info
Lot codes: 955401, 1168853, 1349302, 1956936, 2486857
Distribution
Nationwide and PR, and Australia, Canada, Chile, Colombia, Hong Kong, Puerto Rico
Quantity
231