FDA Enforcement Class II Terminated

NUCLISENS¿ Lysis Buffer, IVD, REF 200292, 48X2 ML LYS Manufacturer bioMerieux SA 376, Chemin de l'Orme 69280 Marcy-l'Etoile - France - Product Usage: to be used for the release of total nucleic acid from biological specimens.

Recall: Z-1365-2020 · Reported March 4, 2020

Enforcement

Recall Number
Z-1365-2020
Event ID
84766
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
BIOMERIEUX - Grenoble
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 4, 2020
Initiation Date
January 10, 2020
Classification Date
February 26, 2020
Termination Date
April 14, 2022
Address
Biomerieux, Institut Christophe Merieux; 5 Rue Des Berges; Cs 30076, Grenoble Cedex 1, N/A, N/A, France

Description

NUCLISENS¿ Lysis Buffer, IVD, REF 200292, 48X2 ML LYS Manufacturer bioMerieux SA 376, Chemin de l'Orme 69280 Marcy-l'Etoile - France - Product Usage: to be used for the release of total nucleic acid from biological specimens.

Reason

Following a customer complaint about colored eluates, leading to invalid results bioMerieux has confirmed an issue with NUCLISENS Lysis Buffer, ref. 200292, lot 19040201. The issue is linked to drift of pH electrode dysfunction.

Code Info

Lot 19040201, Expiry 28-Feb-2021

Distribution

No US distribution. No Mexican consignees. International distribution to South Africa, Uruguay, Tanzania, Romania, Netherlands, Macao, Italy, India, Hong Kong, United Kingdom, France, Spain, Germany, China, Chile, Canada, Belgium, Australia, Argentina, Curacao.

Quantity

1024