AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)

Recall: Z-1364-2022 · Reported July 20, 2022

Enforcement

Recall Number
Z-1364-2022
Event ID
90411
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Fresenius Medical Care Holdings, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
July 20, 2022
Initiation Date
August 3, 2021
Classification Date
July 11, 2022
Termination Date
April 10, 2024
Address
920 Winter St Bld 950, Waltham, MA, 02451-1521, United States

Description

AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)

Reason

AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation

Code Info

UDI-DI: (01)04251285200227, (01)04251285200289, (01)0425128520; Serial Numbers: 7BPS1621 7BDS0783 7BFS0517 7BPS1611 7BDS0776 7BFS0502

Distribution

US Nationwide distribution in the states of NC and WV.

Quantity

2 units