ORTHOFIX, Connector System, LARGE SET SCREW, REF 79-2002

Recall: Z-1364-2018 · Reported April 18, 2018

Enforcement

Recall Number
Z-1364-2018
Event ID
79498
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Orthofix, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 18, 2018
Initiation Date
February 23, 2018
Classification Date
April 11, 2018
Termination Date
June 18, 2019
Address
3451 Plano Pkwy, Lewisville, TX, 75056-9453, United States

Description

ORTHOFIX, Connector System, LARGE SET SCREW, REF 79-2002

Reason

Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.

Code Info

UDI 18257200125364, Lot Numbers: O01 and O02

Distribution

US

Quantity

188 units