ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003
Recall: Z-1363-2018
·
Reported April 18, 2018
Enforcement
- Recall Number
- Z-1363-2018
- Event ID
- 79498
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Orthofix, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 18, 2018
- Initiation Date
- February 23, 2018
- Classification Date
- April 11, 2018
- Termination Date
- June 18, 2019
- Address
- 3451 Plano Pkwy, Lewisville, TX, 75056-9453, United States
Description
ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003
Reason
Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.
Code Info
UDI 18257200125371, Lot Numbers: O02, O03 and O04
Distribution
US
Quantity
397 units